EPA and EU Phthalates on Rxi®-XLB
Applications | | RestekInstrumentation
Phthalate esters are common plasticizers with potential health impacts, requiring accurate monitoring in environmental and industrial samples. Efficient separation and detection of multiple phthalates enhance laboratory throughput and compliance with regulatory methods.
This study evaluates the performance of an Rxi®-XLB capillary column coupled to GC-MS for the rapid separation and quantitation of 20 EPA and EU priority phthalates according to Method 8061A. The goal is to achieve full resolution within a short analysis time.
Phthalate standards were prepared at 50 μg/mL (80 μg/mL benzyl benzoate as internal standard) in methylene chloride. A 1 μL split injection (20:1) was performed on a 30 m × 0.25 mm ID, 0.25 μm Rxi-XLB column. The oven program ran from 200 °C to 330 °C at 30 °C/min with helium at constant linear velocity (66.7 cm/s). The transfer line, source, and injector were maintained at 300 °C, 280 °C, and 280 °C, respectively. Full-scan MS detection (59–400 amu) used electron ionization at 70 eV.
The total ion chromatogram showed baseline resolution of all 20 phthalates in under 5.3 minutes. Early eluting dimethyl and diethyl phthalates were separated at 1.3–1.6 minutes, while high-boiling compounds such as diisononyl and diisodecyl phthalates eluted at 4.5–5.2 minutes. Peak shapes were symmetrical with minimal tailing. Consistent retention times and reproducible peak areas demonstrated the robustness of the method.
The Rxi®-XLB column combined with GC-MS provides a fast, reliable, and regulatory-compliant method for separating and detecting a broad range of phthalates. This approach supports efficient laboratory workflows and sensitive phthalate monitoring.
GC/MSD, GC/SQ, GC columns, Consumables
IndustriesEnvironmental
ManufacturerShimadzu, Restek
Summary
Importance of the Topic
Phthalate esters are common plasticizers with potential health impacts, requiring accurate monitoring in environmental and industrial samples. Efficient separation and detection of multiple phthalates enhance laboratory throughput and compliance with regulatory methods.
Objectives and Study Overview
This study evaluates the performance of an Rxi®-XLB capillary column coupled to GC-MS for the rapid separation and quantitation of 20 EPA and EU priority phthalates according to Method 8061A. The goal is to achieve full resolution within a short analysis time.
Methodology
Phthalate standards were prepared at 50 μg/mL (80 μg/mL benzyl benzoate as internal standard) in methylene chloride. A 1 μL split injection (20:1) was performed on a 30 m × 0.25 mm ID, 0.25 μm Rxi-XLB column. The oven program ran from 200 °C to 330 °C at 30 °C/min with helium at constant linear velocity (66.7 cm/s). The transfer line, source, and injector were maintained at 300 °C, 280 °C, and 280 °C, respectively. Full-scan MS detection (59–400 amu) used electron ionization at 70 eV.
Instrumentation Used
- Shimadzu 2010 GC coupled with QP2010+ quadrupole MS
- Rxi®-XLB column, 30 m × 0.25 mm ID, 0.25 μm film
- Premium 3.5 mm liner with wool for split injections
Main Results and Discussion
The total ion chromatogram showed baseline resolution of all 20 phthalates in under 5.3 minutes. Early eluting dimethyl and diethyl phthalates were separated at 1.3–1.6 minutes, while high-boiling compounds such as diisononyl and diisodecyl phthalates eluted at 4.5–5.2 minutes. Peak shapes were symmetrical with minimal tailing. Consistent retention times and reproducible peak areas demonstrated the robustness of the method.
Benefits and Practical Applications
- Rapid analysis increases sample throughput in QA/QC and environmental labs
- Compliance with EPA and EU phthalate monitoring requirements
- Low detection limits and high reproducibility for trace-level screening
Future Trends and Opportunities
- Integration with automated sample preparation for high-volume testing
- Coupling to high-resolution MS for nontargeted phthalate screening
- Miniaturized GC-MS systems for field-deployable analysis
Conclusion
The Rxi®-XLB column combined with GC-MS provides a fast, reliable, and regulatory-compliant method for separating and detecting a broad range of phthalates. This approach supports efficient laboratory workflows and sensitive phthalate monitoring.
References
- Restek Corporation. EPA and EU Phthalates on Rxi®-XLB. Application Note.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Phthalates on Rxi®-XLB (Extended List)
|Shimadzu|Applications
Phthalates on Rxi®-XLB (Extended List) Peaks 1. Dimethyl phthalate 2. Dimethyl isophthalate 3. Diethyl phthalate 4. Benzyl benzoate 5. Diisobutyl phthalate 6. Di-n-butyl phthalate 7. Bis(2-methoxyethyl) phthalate 8. Bis[4-methyl-2-pentyl] phthalate isomer 1 9. Bis[4-methyl-2-pentyl] phthalate isomer 2 10. Bis(2-ethoxyethyl) phthalate…
Key words
rxi, rxixlb, xlbconsultation, consultationscan, scanpeaks, peaksamu, amuphthalates, phthalatesacknowledgement, acknowledgementphthalate, phthalatemethylene, methylenethank, thankscans, scansvent, ventmin, minextended
EPA and EU Phthalates on Rxi®-35Sil MS
|Shimadzu|Applications
EPA and EU Phthalates on Rxi®-35Sil MS Peaks 1. Dimethyl phthalate 2. Diethyl phthalate 3. Benzyl benzoate 4. Diisobutyl phthalate 5. Di-n-butyl phthalate 6. Bis(2-methoxyethyl) phthalate 7. Bis[4-methyl-2-pentyl] phthalate isomer 1 8. Bis[4-methyl-2-pentyl] phthalate isomer 2 9. Bis(2-ethoxyethyl) phthalate 10.…
Key words
consultation, consultationscan, scanpeaks, peaksamu, amuacknowledgement, acknowledgementthank, thankscans, scansvent, ventmin, mintic, ticdelay, delaysec, seclike, likegroup, groupshimadzu
EPA and EU Phthalates on Rxi®5ms
|Shimadzu|Applications
EPA and EU Phthalates on Rxi®5ms Peaks 1. Dimethyl phthalate 2. Diethyl phthalate 3. Benzyl benzoate 4. Diisobutyl phthalate 5. Di-n-butyl phthalate 6. Bis(2-methoxyethyl) phthalate 7. Bis[4-methyl-2-pentyl] phthalate isomer 1 8. Bis[4-methyl-2-pentyl] phthalate isomer 2 9. Bis(2-ethoxyethyl) phthalate 10. Di-n-pentyl…
Key words
consultation, consultationscan, scanpeaks, peaksamu, amuacknowledgement, acknowledgementthank, thankscans, scansvent, ventmin, mintic, ticdelay, delaysec, seclike, likegroup, groupshimadzu
EPA and EU Phthalates on Rtx®-CLPesticides2
|Shimadzu|Applications
EPA and EU Phthalates on Rtx®-CLPesticides2 Peaks 1. Dimethyl phthalate 2. Diethyl phthalate 3. Benzyl benzoate 4. Diisobutyl phthalate 5. Di-n-butyl phthalate 6. Bis(2-methoxyethyl) phthalate 7. Bis[4-methyl-2-pentyl] phthalate isomer 1 8. Bis[4-methyl-2-pentyl] phthalate isomer 2 9. Bis(2-ethoxyethyl) phthalate 10. Di-n-pentyl…
Key words
scan, scanconsultation, consultationpeaks, peaksamu, amuacknowledgement, acknowledgementscans, scansthank, thankvent, ventmin, mintic, ticdelay, delaysec, seclike, likegroup, groupshimadzu