Hydrocarbons, C1 – C6 - Analysis of natural gas
Applications | 2011 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Importance of the topic
The composition of natural gas is critical for energy efficiency, safety and regulatory compliance. A reliable method to quantify hydrocarbons from methane (C1) to hexane (C6) ensures accurate quality control and supports fuel valuation.Objectives and study overview
This application note evaluates the performance of an Agilent CP-SilicaPLOT column for direct injection gas chromatography analysis of natural gas C1–C6. The study demonstrates peak separation, resolution and operational robustness under typical energy and fuels laboratory conditions.Methodology and instrumentation
- Technique: Wide-bore GC with direct injection (sample size 2 µL).
- Column: Agilent CP-SilicaPLOT, 0.53 mm × 30 m, df = 6 µm.
- Carrier gas: Helium at 20 kPa (0.2 bar).
- Oven program: 50 °C (5 min) → 225 °C at 5 °C/min.
- Injector temperature: 225 °C; Detector: FID at 250 °C.
Main results and discussion
The method delivered narrow methane peaks and baseline separation of C1–C6 hydrocarbons. Branched paraffins eluted before their linear analogues, confirming reliable isomer resolution. The CP-SilicaPLOT column exhibited high mechanical stability, enabling high flow rates, valve switching and backflush techniques without loss of performance.Benefits and practical applications
- Direct injection reduces sample preparation time and potential errors.
- High separation factor provides precise quantification of light hydrocarbons.
- Robust column withstands demanding operational modes, supporting routine analysis.
- Applicable for QC/QA in energy, petrochemical and environmental laboratories.
Future trends and opportunities
Integration with mass spectrometry or advanced detectors can enhance sensitivity and identification capabilities. Development of portable GC systems using PLOT technology offers on-site monitoring potential. Expanded use of backflush and valve switching will allow analysis of heavier components in complex gas mixtures.Conclusion
The Agilent CP-SilicaPLOT column combined with direct injection GC-FID provides a fast, reliable and high-resolution approach for C1–C6 natural gas analysis. Its robustness and performance make it a valuable tool for diverse industrial and research applications.References
- Agilent Technologies, Inc. Application Note A01354, 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Hydrocarbons, C1 – C4 - Analysis of impurities in propylene
2011|Agilent Technologies|Applications
Hydrocarbons, C1 – C4 Analysis of impurities in propylene Application Note Energy & Fuels Authors Introduction Agilent Technologies, Inc. The separation of impurities in propylene can be done very well on an Agilent CP-SilicaPLOT column. The 0.53 mm id column…
Key words
propylene, propyleneimpurities, impuritiesfuels, fuelsdone, doneprinted, printedinjector, injectorusa, usaenergy, energyauthors, authorssize, sizevery, verytechnique, techniquecarrier, carrieridentification, identificationwell
Solutions for light hydrocarbons and gasses: PLOT columns
2019|Agilent Technologies|Presentations
Solutions for light hydrocarbons and gasses: PLOT columns 1 April 17, 2019 Separation Superhero Seminar Tour Solutions for Light Hydrocarbons and Gasses: PLOT Columns Allen K. Vickers Agilent Technologies Folsom Technology Center California, USA April 18, 2019 2 Table of…
Key words
superhero, superherotour, tourseminar, seminarseparation, separationplot, plotporous, porousoxygenates, oxygenatesparticle, particlecolumns, columnscolumn, columnalcohol, alcoholalumina, aluminahydrocarbons, hydrocarbonsgas, gasether
PLOT Columns Separation Solutions for Light Hydrocarbons & Gases
2021|Agilent Technologies|Presentations
PLOT Columns Separation Solutions for Light Hydrocarbons & Gases Porous Layer Open Tubulars Johan Kuipers Training & Development Sept 2021 1 Webinar Agilent PLOT Columns DE44454.2362731481 Table of Boiling Point Fractions Carbon # C1 C2 C3 C4 C5 C6 C7…
Key words
plot, plotwebinar, webinarcolumns, columnsagilent, agilentporous, porousoxygenates, oxygenatesalumina, aluminaalcohol, alcoholpolymer, polymerparticle, particlehydrocarbons, hydrocarbonspolar, polarether, etherselectivity, selectivitycolumn
Halogenated hydrocarbons - Analysis of chlorinated volatiles and C1-C6 hydrocarbons
2011|Agilent Technologies|Applications
Halogenated hydrocarbons Analysis of chlorinated volatiles and C1-C6 hydrocarbons Application Note Environmental Authors Introduction Agilent Technologies, Inc. Gas chromatography with an Agilent CP-SilicaPLOT column separates nine chlorinated volatiles and hydrocarbons in 14 minutes. Chloromethane (methyl chlorlde) and vinylchloride are separated…
Key words
vinylchloride, vinylchloridebutane, butanechloromethane, chloromethanechlorlde, chlorldesilicaplot, silicaplotchloride, chloridemethyl, methylchloroethane, chloroethaneethane, ethanehalogenated, halogenatedpropane, propanepentane, pentanecourtesy, courtesyseparates, separateschlorinated