GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

High Temperature Static Headspace Analysis of Polymers using Comprehensive GCXGC with a Mass Selective Detector

Posters | 2012 | Agilent TechnologiesInstrumentation
GCxGC, GC/MSD
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies

Summary

Importance of Topic


High-temperature static headspace analysis provides critical insights into thermal stability and degradation pathways of polymers at elevated temperatures. Capturing volatile and semi-volatile compounds formed at 300°C supports materials research, quality control, and failure analysis in polymer science.

Objectives and Study Overview


This work evaluates pulsed flow modulation comprehensive two-dimensional gas chromatography (GCxGC) with simultaneous mass selective detection (MSD) and flame ionization detection (FID) for headspace analysis of polymers heated to 300°C. The goals include demonstrating high-temperature equilibration, optimizing flow splits, and generating characteristic fingerprints for various polymer types.

Methodology and Instrumentation Used


  • Static headspace sampler operated at 300°C for 30 minutes in 10 mL vials containing 1.7 g of polymer; septa lined with Kapton to prevent bleed.
  • Pulsed flow modulation GCxGC setup: first-dimension column DB-5ms (10 m × 0.15 mm × 0.15 µm); second-dimension DB-Wax (10 m × 0.25 mm × 0.10 µm).
  • Hydrogen carrier gas maintained constant flow and pressure across oven temperatures from 40°C to 260°C.
  • Detector configuration via Capillary Flow Technology splitter: ~2–4 mL/min to MSD, ~17–25 mL/min to FID.
  • MSD parameters: scan range 30–275 amu, scan rate 12 500 amu/s, ~26 scans/s to resolve 120–150 ms GCxGC peaks.

Main Results and Discussion


  • GCxGC contour plots revealed unique emission patterns for linear low-density polyethylene, styrene-based copolymers (styrene–methyl methacrylate, styrene–butadiene), PET, and low-density polyethylene.
  • One-dimensional GC temperature ramp experiments (85°C to 300°C) showed increased diversity of volatiles at higher temperatures.
  • Kapton-lined septa exhibited negligible siloxane bleed, confirming suitability for prolonged 300°C operation.
  • Comprehensive GCxGC effectively separated minor additives such as BHT from polymer degradation fragments, demonstrating high analytical resolution.

Benefits and Practical Applications


  • Rapid, non-destructive fingerprinting of polymer composition and degradation products without pyrolysis.
  • Quality control monitoring of thermal stability and additive distribution in plastic materials.
  • Support for failure analysis and research into polymer breakdown mechanisms.

Future Trends and Opportunities


Advancements in GCxGC instrumentation, including high-resolution mass spectrometry and automated data processing, will enhance polymer profiling. Integration with machine learning for fingerprint recognition and real-time inline headspace monitoring on production lines represents promising directions.

Conclusion


High-temperature static headspace GCxGC with combined MSD and FID detection provides a powerful approach for characterizing polymer emissions at 300°C. The method delivers detailed compositional fingerprints, complements traditional pyrolysis techniques, and supports diverse applications in polymer quality control and research.

References


  • Shin T, Hajime O, Chuichi W. Pyrolysis-GC/MS Data Book of Synthetic Polymers. Oxford, UK: Elsevier; 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Multi-detector Configurations and Applications of Differential Flow Modulation GCXGC
Multi-detector Configurations and Applications of Differential Flow Modulation GCXGC (K.29) Roger L Firor, Agilent Technologies, Inc., Wilmington, DE, USA Multi-Detector Configurations using CFT Technology Introduction ECD-FID-MSD System Performance Pesticide Mix 10ppm: ECD MSD-FID-ECD System Pneumatics The enhanced separation power of…
Key words
mod, modecd, ecdmsd, msdfid, fidmodulated, modulatedrestrictor, restrictorbhc, bhcflow, flowscans, scanspneumatics, pneumaticssecond, secondmodulator, modulatormmi, mmisplitter, splittermodulation
Reliable Analysis of the Light Ends of Petroleum Fractions and Crude Oil Using Capillary Flow Backflush Technology
Reliable Analysis of the Light Ends of Petroleum Fractions and Crude Oil Using Capillary Flow Backflush Technology Roger L. Firor Agilent Technologies Chemical Analysis Group Wilmington, DE USA He IN EPC1 P2 Oven Pre-Column Purged Union Det Column 2 Gulf…
Key words
backflush, backflushpre, precolumn, columnaux, auxunion, unioncrudes, crudespurged, purgedcft, cftanalytical, analyticalinlet, inletcrude, crudeflow, flowmultimode, multimodeheavy, heavymmi
Hydrocarbon Class Analysis of Conventional and Synthetic Aviation Turbine Fuels by ASTM D8396
Application Note Energy and Chemicals Hydrocarbon Class Analysis of Conventional and Synthetic Aviation Turbine Fuels by ASTM D8396 Flow-modulated GCxGC-FID using the Agilent Reverse Flow Modulator and 8890 GC Author Abstract Scott Hoy Agilent Technologies, Inc. This application note demonstrates…
Key words
naphthenes, naphthenesparaffins, paraffinsdimension, dimensionrfm, rfmretention, retentiontotal, totalisoparaffins, isoparaffinsgcxgc, gcxgchydrogen, hydrogensec, secmodulation, modulationtemplate, templatecyclohexane, cyclohexanefuel, fuelmolecular
Stop Flow Modulated GCxGC
Stop Flow Modulated GCxGC
2020|Agilent Technologies|Technical notes
Stop-Flow Modulated GCxGC Product Note Cost-effective comprehensive GCxGC Author Tommy Saunders Activated Research Company 7561 Corporate Way Eden Prairie, MN 55344 [email protected] Abstract A simple, cost-effective technique for flow modulated comprehensive two-dimensional gas chromatography (GCxGC) has been applied on an…
Key words
accumulation, accumulationmodulation, modulationdimension, dimensionflow, flowbiodiesel, biodieselmodulator, modulatorstop, stopdimensional, dimensionalcomprehensive, comprehensivecapillary, capillaryconnected, connectedcontrolled, controlledactivated, activatedalphatics, alphaticssecond
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike