GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Phthalate Esters on Equity-1701

Applications | 2003 | MerckInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Merck

Summary

Significance of the Topic


Phthalate esters are widely studied as potential endocrine disruptors and are classified as priority pollutants by regulatory agencies such as the US EPA. Accurate analysis of these compounds is critical for environmental monitoring, quality control in manufacturing, and health risk assessment.

Objectives and Study Overview


The primary objective of the study was to achieve baseline separation of 17 phthalate esters listed in EPA Methods 8061A, 506, and 606 using a single Equity-1701 GC column. This work serves to validate the method as a confirmatory tool and to demonstrate its compatibility with sensitive detectors.

Methodology and Instrumentation


The separation was performed on a gas chromatograph equipped with a low-bleed Equity-1701 capillary column. Experimental conditions included
  • Column dimensions 30 m x 0.25 mm ID, film thickness 0.25 mu m
  • Oven program 40 degC (1 min) to 280 degC at 10 degC/min
  • Injection splitless, 1 uL sample volume, inlet at 250 degC
  • Detector flame ionization detector at 325 degC, helium carrier gas at constant flow 1.3 mL/sec
  • Sample load 50 ng on column of a custom phthalate ester mix

Main Results and Discussion


The method achieved excellent peak shapes and resolution for all 17 target esters, including structural isomers. The low bleed profile of the Equity-1701 phase at high temperatures ensures minimal background and extends compatibility to more sensitive detectors such as electron capture detectors. Chromatographic data confirmed baseline separation across a 14 to 32 minute retention window.

Benefits and Practical Applications


  • Capable of serving as a confirmatory column for EPA phthalate analysis methods
  • High temperature stability and low bleed enable use with trace detectors
  • Robust separation suitable for routine environmental and industrial QA/QC analyses

Future Trends and Potential Uses


Advancements may include coupling the method with mass spectrometric detection for enhanced specificity, automating sample preparation workflows, and adapting the approach for rapid screening in emerging industrial sectors. Exploration of narrower and shorter capillaries for faster throughput and integration with multidimensional GC could further extend its applicability.

Conclusion


The Equity-1701 column provides reliable, high resolution separation of a comprehensive panel of phthalate esters under standard EPA method conditions. Its low bleed performance and thermal stability support its role as both a primary and confirmatory analytical tool in environmental and industrial laboratories.

References


Application Report 17, Sigma-Aldrich Co, 2003. EPA Methods 8061A, 506, and 606

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Phthalate Esters on Equity-5
Phthalate Esters on Equity-5
2003|Merck|Applications
Application Report 16 Author: K. Stenerson Phthalate Esters on Equity-5 Raw Data File Name: C:\data\1\GS100App\frot0055.D (HPCS53) Phthalate esters are frequently listed as suspected endocrine disrupters, and the US EPA has included several on their list of priority pollutants. In this…
Key words
phthalate, phthalatebenzyl, benzylphthtlate, phthtlatediamyl, diamylnonyl, nonyldicyclohexyl, dicyclohexyldiisobutyl, diisobutyldibutyl, dibutylbenzoate, benzoateids, idsoctyl, octylhexyl, hexyldiethyl, diethylbleed, bleedshapes
Equity™ Capillary GC Columns
Equity™ Capillary GC Columns
2003|Merck|Brochures and specifications
SUPELCO EQUITY CAPILLARY GC COLUMNS Equity™ Capillary GC Columns The performance you demand... the service you deserve... from the company you trust. Updated with: ● NEW Equity-1701 columns ● New Applications 1 sigma-aldrich.com/supelco SUPELCO EQUITY CAPILLARY GC COLUMNS The performance…
Key words
equity, equitysupelco, supelcocapillary, capillarymin, mincolumns, columnscapseal, capsealexcellent, excellentbullet, bulletbleed, bleedresponse, responseshapes, shapesacetate, acetatemethyl, methylendosulfan, endosulfanbhc
GC-MS Analysis of Phthalates: Comparison of GC Stationary Phase Performance
GC-MS Analysis of Phthalates: Comparison of GC Stationary Phase Performance By Dan Li, Rebecca Stevens, and Chris English Abstract Phthalates are ubiquitous in the environment and have attracted attention due to their potential adverse impact on human health. For this…
Key words
neat, neatphthalate, phthalateisomer, isomerrxi, rxitechnical, technicalisophthalate, isophthalatebutyl, butylxlb, xlbrtxclpesticides, rtxclpesticidesoctyl, octyltic, ticmix, mixdecyl, decylpriority, priorityepa
Fast GC - Increase GC Speed Without Sacrificing Resolution
Fast GC - Increase GC Speed Without Sacrificing Resolution
2015|Merck|Brochures and specifications
Fast GC Increase GC Speed Without Sacrificing Resolution Increase Productivity Utilize Resources More Effectively Achieve Quicker Turn-Around Times Analyze More Samples Fast GC | Increase GC Speed Without Sacrificing Resolution Overview Shorter analysis times allow increased sample throughput, which translates…
Key words
sacrificing, sacrificingfast, fastmin, mintrifluoromethylsulfonyl, trifluoromethylsulfonylincrease, increasebenzo, benzoimide, imidephthalate, phthalateresolution, resolutionfluoranthene, fluoranthenespeed, speedlimonene, limonenebhc, bhcbis, bisneral
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike