Low Level Determination of Polynuclear Aromatic Hydrocarbons in Environmental Samples Using GC/MS

Applications | 1999 | MerckInstrumentation
GC/MSD
Industries
Environmental
Manufacturer
Merck

Summary

Significance of the Topic


The accurate determination of polynuclear aromatic hydrocarbons (PAHs) at trace levels is critical for environmental monitoring and regulatory compliance. PAHs are common contaminants in soil and groundwater, often originating from petroleum spills and combustion processes. Conventional HPLC methods (EPA SW-846 Method 8310) rely heavily on retention times, which can lead to misidentification when complex matrices are present. Adopting an optimized GC/MS approach enhances both sensitivity and specificity, supporting reliable data for risk assessment and remediation.

Aims and Study Overview


This study aimed to develop and validate a GC/MS procedure capable of detecting PAHs at sub-microgram per milliliter levels in environmental samples. The investigators sought to combine the selectivity of mass spectrometry with the sensitivity required for regulatory reporting limits. Key objectives included optimizing sample introduction, reducing chromatographic run times, and demonstrating method performance in soil extracts.

Methodology


• Sample introduction employed a pressure-pulse splitless injection. Initial injector head pressure was set to 20 psi for one minute, maximizing analyte transfer onto the column.
• Following the pulse, column flow was maintained at 1.0 mL/min under constant flow programming.
• Oven temperature program: hold at 50 °C for 3 minutes; ramp to 200 °C at 10 °C/min; then to 290 °C at 20 °C/min. Total run time: 30 minutes.
• Calibration covered 0.25–30 µg/mL. Response factors demonstrated linearity (RSD < 10%). Method detection limits in soil extracts ranged from 0.04 to 0.14 µg/mL (MDL calculation, n=9).

Instrumentation Used


  • Gas chromatograph equipped for pressure programming and splitless injection.
  • PTE-5 capillary column (30 m × 0.32 mm ID, 0.25 µm film).
  • Single-quadrupole mass spectrometer operating in full-scan mode (e.g., HP 5972).

Main Results and Discussion


The optimized GC/MS method resolved key PAH isomers including benzo(b)fluoranthene vs. benzo(k)fluoranthene and benzo(a)anthracene vs. chrysene. Full-scan mass spectra provided unequivocal compound identification and library search capability, overcoming coelution challenges in petroleum-contaminated matrices. Calibration data showed consistent response factors across six concentration levels with mean RSD values below 8.5%. Soil spike experiments yielded MDLs well below 0.1 µg/mL for most PAHs.

Benefits and Practical Applications


  • Enhanced selectivity through unique mass spectral ions improves confidence in compound identification.
  • Lower reporting limits (typically < 0.25 µg/mL) facilitate compliance with stringent environmental regulations.
  • Shorter analysis times (30 minutes) increase laboratory throughput.
  • Applicability to various extraction techniques (Soxhlet, sonication, liquid-liquid).

Future Trends and Applications


Ongoing advances in GC/MS technology, such as faster column materials, improved ionization sources, and high-resolution analyzers, will further lower detection limits and accelerate analyses. Integration with automated sample preparation and data processing platforms promises to enhance reproducibility. Expanding the approach to emerging contaminants and coupling with tandem MS detection offers broader applicability in environmental forensics and public health studies.

Conclusion


The optimized GC/MS method provides a robust, sensitive, and selective alternative to traditional HPLC for low-level PAH determination in environmental samples. Pressure-pulse injection and pressure-programmed chromatography reduce analyte loss and analysis time, while full-scan mass spectrometry ensures reliable identification in complex matrices.

References


  • Cleghorn N., Loj A., Zaworski S. Low Level Determination of Polynuclear Aromatic Hydrocarbons in Environmental Samples Using GC/MS. The Supelco Reporter. Vol. 15, No. 4; 1996.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
GC Analyses of Polynuclear Aromatic Hydrocarbons
Application Note 108 GC Analyses of Polynuclear Aromatic Hydrocarbons Polynuclear aromatic hydrocarbons (PAHs) can be separated by packed column or capillary column gas chromatography. An SPB-5 bonded phase fused silica capillary column separates the 16 PAHs considered priority pollutants in…
Key words
benzo, benzofluoranthene, fluoranthenepyrene, pyreneanthracene, anthracenepolynuclear, polynuclearphenanthrene, phenanthrenecustom, customperylene, perylenecapillary, capillaryfluorene, fluorenechrysene, chrysenepahs, pahsaromatic, aromaticpacked, packedhydrocarbons
Solid Phase Microextraction of Semivolatile Compounds
Application Note 6 Solid Phase Microextraction of Semivolatile Compounds An SPME fiber coated with 7µm of polydimethylsiloxane is ideal for extracting semivolatile compounds from water. The SPME technology eliminates the use of solvents in extracting water samples for organic compound…
Key words
benzo, benzobutylphthalate, butylphthalatemanual, manualsampling, samplinganthracene, anthracenefluoranthene, fluoranthenespme, spmediethylphthalate, diethylphthalatedimethylphthalate, dimethylphthalatepyrene, pyrenesemivolatile, semivolatiledibenz, dibenzmicroextraction, microextractionoctylphthalate, octylphthalateghi
USEPA Procedures for Wastewater Analyses by Packed Column GC and HPLC
Bulletin 775E USEPA Procedures for Wastewater Analyses by Packed Column GC and HPLC This bulletin outlines the United States Environmental Protection Agency’s procedures for analyzing 113 organic priority pollutants in wastewater. The chromatographic column or column packing listed by the…
Key words
confirmational, confirmationalqualifications, qualificationscolumn, columnbenzo, benzopollutants, pollutantssupelco, supelcobhc, bhcfluoranthene, fluoranthenemin, mindetector, detectorconcentration, concentrationfid, fidanthracene, anthracenesample, sampleendosulfan
Analysis of multiple matrices with a single calibration curve for polycyclic aromatic hydrocarbons (PAHs) with the ISQ 7610 GC-MS system following EPA Method 8270E
Application note | 000455 Environmental analysis Analysis of multiple matrices with a single calibration curve for polycyclic aromatic hydrocarbons (PAHs) with the ISQ 7610 GC-MS system following EPA Method 8270E Authors Goal Chiara Calaprice , Brian Pike , The aim…
Key words
benzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenethra, thranaphthalene, naphthalenedibenzofuran, dibenzofuranacenaphthene, acenaphthenechrysene, chrysenepahs, pahsphenanthrene, phenanthrenebenz, benzaph, aphacenaphthylene, acenaphthylenefluorene