Orthogonal Selectivity for Separation of Light and Heavy Petroleum Hydrocarbons by 2D GC using Zebron™ ZB-1HT Inferno™ and ZB-35HT Inferno GC Columns

Applications | 2019 | PhenomenexInstrumentation
GCxGC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, Phenomenex

Summary

Importance of the Topic


Complex petroleum hydrocarbon mixtures present significant challenges for conventional one-dimensional gas chromatography due to coelutions and limited peak capacity. Two-dimensional GC offers enhanced peak capacity and orthogonal selectivity to fully resolve light and heavy hydrocarbon fractions, enabling detailed compositional analysis essential for refining, quality control, and research.

Objectives and Overview of the Study


This work aimed to evaluate the performance of a two-dimensional GC method using a Zebron ZB-1HT Inferno column in the first dimension and a Zebron ZB-35HT Inferno column in the second dimension. The study focused on separating a gas oil petroleum fraction and a heavy petroleum fraction, mapping individual aliphatic and aromatic species up to C42, and demonstrating the orthogonal selectivity afforded by these columns.

Instrumentation Used


  • Gas Chromatograph: Agilent 7890A system modified with a gas-phase microfluidics interface
  • First Dimension Column: Zebron ZB-1HT Inferno (100% dimethylpolysiloxane, 15 m × 0.10 mm × 0.10 μm)
  • Second Dimension Column: Zebron ZB-35HT Inferno (35% phenyl, 65% dimethylpolysiloxane, 5 m × 0.25 mm × 0.18 μm)
  • Carrier Gas: Hydrogen (1.0 mL/min first dimension; 2.4 mL/min second dimension)
  • Injection: Split 20:1 at 250 °C, 1 μL sample, ZebronPLUS Z-Liner
  • Detector: Flame Ionization (FID) at 360 °C, 100 Hz acquisition
  • Oven Programs: Optimized temperature ramps up to 400–430 °C for comprehensive elution

Main Results and Discussion


The two-dimensional analysis successfully resolved complex hydrocarbon mixtures up to C42. The first dimension provided boiling-point separation of linear and branched alkanes, while the second dimension exploited pi-pi interactions to separate aromatic species. Key compound classes separated include:
  • Linear alkanes (n-paraffins and iso-paraffins)
  • Linear alkenes (olefins)
  • Cyclic alkanes
  • Aromatics: one-ring derivatives (benzene and tetralin), two-ring (naphthalene derivatives), di-benzene, three-ring, four-ring, and five-ring compounds
The Zebron Inferno columns feature Engineered Cross-linking™ bonding and a high-temperature-stable polyimide coating, delivering extremely low bleed and extended lifetime at temperatures up to 430 °C.

Benefits and Practical Applications


  • Comprehensive resolution of light and heavy hydrocarbon fractions with minimal coelution
  • High-temperature capability enabling full elution of refractory compounds
  • Orthogonal selectivity for precise distinction of aliphatic and aromatic species
  • Enhanced sensitivity and dynamic range using hydrogen carrier gas and FID detection
  • Valuable for refining process optimization, quality control, environmental monitoring, and research applications

Future Trends and Opportunities


  • Integration of rapid-scanning detection technologies to capture ultra-narrow second-dimension peaks
  • Application of automated data analysis tools and chemometric methods for matrix complexity
  • Expansion of two-dimensional GC to other complex organic systems, such as biofuels and environmental pollutants
  • Development of advanced microfluidic interfaces to reduce dead volume and improve modulation

Conclusion


The combination of Zebron ZB-1HT Inferno and ZB-35HT Inferno columns in a two-dimensional GC setup provides robust high-temperature operation and true orthogonal selectivity. This approach delivers detailed molecular mapping of complex petroleum fractions, facilitating improved process control and analytical insight.

Acknowledgement: The authors thank IRCELYON for collaboration and assistance in method development.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Zebron Inferno™ Rugged GC Performance to 430 ℃!
Zebron Inferno™ Rugged GC Performance to 430 ℃!
2015|Phenomenex|Brochures and specifications
™ GC Columns Zebron Inferno ™ Rugged GC Performance to 430 ℃! 4 Versatile Selectivities • ZB-1HT • ZB-5HT • ZB-35HT • ZB-XLB-HT www.phenomenex.com/InfernoGC Introducing the Zebron™ Experience Get More Than Just A Column When you choose Zebron, you get…
Key words
phenomenex, phenomenexinferno, infernozebron, zebronxlb, xlbmin, mincolumn, columnbleed, bleedapplications, applicationsordering, orderingferrule, ferrulegraphite, graphitecolumns, columnshigh, highinlet, inletdistillation
GC×GC-VUV and GC×GC-FID Analysis of Gasoline, Middle Distillates and Crude Oil Distillation Cuts using Zebron™ ZB-35HT and ZB-1plus GC Columns
TN-2089 APPLICATIONS GC×GC-VUV and GC×GC-FID Analysis of Gasoline, Middle Distillates and Crude Oil Distillation Cuts using Zebron™ ZB-35HT and ZB-1plus GC Columns Max Jennerwein2, Markus Eschner2, Thomas Wilharm2, Ramkumar Dhandapani1, and Tim Nelson1 1 Phenomenex, Inc., 411 Madrid Ave., Torrance,…
Key words
reproducing, reproducingvuv, vuvlove, lovemodulation, modulationdimensional, dimensionaltrouble, troublemodulator, modulatoroil, oilinsight, insightapplications, applicationsolefins, olefinshelp, helptim, timhaving, havingheating
Group-Type Analysis of Petroleum Products by GC × GC using Zebron® ZB-1HT Inferno™ and ZB-50 Complementary Selectivity GC Columns
TN-2086 APPLICATIONS Ramkumar Dhandapani, Ph.D. He has a PhD in Analytical Chemistry and over 15 years experience in chromatographic method development and troubleshooting. Ramkumar loves to write poems, read Shakespeare, and attend Shakespeare plays. Group-Type Analysis of Petroleum Products by…
Key words
secs, secschromspace, chromspacehthale, hthalenes, nesboiling, boilinglandmark, landmarknap, napphase, phasereproducing, reproducingreversed, reversedlove, lovehomologous, homologouspoint, pointdimension, dimensionzebron
GC Method Development Special - Optimize GC Column Selection and Save!
GC Method Development Special Optimize GC Column Selection and Save! Column selection tips and guidelines inside Introducing Phase Selection the Zebron Tips Experience ™ Selectivity Has the Biggest Impact on Resolution Resolution between two analytes is mainly determined by the…
Key words
price, pricelimits, limitsphenomenex, phenomenexxlb, xlbproprietary, proprietaryzebron, zebronexclusive, exclusivephase, phasewax, waxsemivolatiles, semivolatilesinferno, infernowaxplus, waxplusbioethanol, bioethanolstabilwax, stabilwaxffap
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike