Extraction and GC/MS Analysis of Melamine From Dog Food Using A Zebron™ ZB-5MSPLUS ™ GC Column

Applications | 2015 | PhenomenexInstrumentation
GC/MSD, GC columns, Consumables
Industries
Food & Agriculture
Manufacturer
Phenomenex

Summary

Significance of the Topic


Pet food adulteration with melamine poses serious health risks to companion animals and can lead to regulatory recalls and liability concerns. Accurate and reliable methods for detecting melamine and its analogues in complex matrices like dog food are essential to ensure product safety, maintain consumer confidence, and comply with regulatory standards.

Objectives and Overview


This study aimed to develop a streamlined extraction and gas chromatography‐mass spectrometry (GC/MS) method to quantify melamine, ammelide, ammeline, cyanuric acid, and benzoguanamine in dog food. A Zebron™ ZB-5MSPLUS™ GC column was evaluated for its inertness and resolution performance to deliver clear separation of target analytes from matrix interferences.

Methodology


The sample preparation process involved:
  1. Weighing 0.5 g of homogenized dog food and extracting with 10 mL of diethanolamine/water/acetonitrile (1:4:5, v/v/v).
  2. Sonication for 30 minutes followed by centrifugation at 5000 rpm for 10 minutes.
  3. Transferring 100 µL of the supernatant to an autosampler vial and evaporating to dryness under nitrogen.
  4. Reconstituting the residue in 100 µL of acetonitrile/pyridine (1:1, v/v) and derivatizing with 100 µL of BSTFA containing 1 % TMCS at 70 °C for 45 minutes.

Used Instrumentation


  • GC Column: Zebron ZB-5MSPLUS™, 30 m × 0.25 mm I.D., 0.25 µm film thickness
  • Injector: Splitless injection (1 µL) at 280 °C
  • Carrier Gas: Helium at 1 mL/min (constant flow)
  • Oven Program: 75 °C (1 min), ramp to 320 °C @ 15 °C/min, hold 4 min
  • Mass Spectrometer: Electron impact ionization, detector temperature 320 °C

Main Results and Discussion


The method achieved baseline separation of all five analytes with sharp, Gaussian peaks and minimal matrix interference. Recovery at a spiked level of 100 µg/g was within acceptable limits, demonstrating method accuracy. Total ion chromatograms (TIC) and extracted ion chromatograms (EIC) confirmed selective detection of melamine and related compounds in the complex dog food matrix.

Benefits and Practical Applications


This protocol offers:
  • High sensitivity and selectivity for melamine and analogues
  • Efficient sample cleanup and minimal matrix effects
  • Reproducible performance suitable for routine QA/QC laboratories
  • Compliance with regulatory requirements for pet food safety

Future Trends and Applications


Advancements may include automated high-throughput sample preparation, coupling with tandem MS for enhanced sensitivity, miniaturized extraction techniques to reduce solvent use, and extension of the approach to other food matrices and contaminants.

Conclusion


The developed GC/MS method using the Zebron ZB-5MSPLUS™ column provides a robust, reproducible, and sensitive approach for quantifying melamine and related analytes in dog food, supporting industry needs for safety monitoring and regulatory compliance.

References


  • Matt Trass and Tim Anderson, Phenomenex Application Note 23245, 2015.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Zebron ZB-5MSPLUS™
Zebron ZB-5MSPLUS™
2015|Phenomenex|Brochures and specifications
™ ZB-5MSPLUS ™ • Improved responses • Better peak shapes • Same 5ms selectivity The Next Generation of Inertness GUARANTEED alternative to: • Agilent® DB®-5ms Ultra Inert • Agilent HP®-5ms Ultra Inert • Restek® Rxi®-5Sil MS • Supelco® SLB®-5ms…
Key words
phenomenex, phenomenexzebron, zebroninertness, inertnessmin, minyour, yourultra, ultrainert, inertpeak, peaknext, nextgeneration, generationtest, testbetter, bettersurpass, surpassprice, pricephases
Extraction and GC/MS Analysis of Cannabinoids from Brownies using roQ™ QuEChERS Extraction Salts and a Zebron™ ZB-5MSPLUS ™ GC Column
.25 Extraction and GC/MS Analysis of Cannabinoids from Brownies using roQ™ QuEChERS Extraction Salts and a Zebron™ ZB-5MSPLUS™ GC Column Matt Trass and Tim Anderson Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA Overview Extraction Procedure This application note…
Key words
phenomenex, phenomenexbrownies, browniestrass, trassroq, roqanderson, andersonextraction, extractionmatt, mattterms, termspuerto, puertorico, ricousa, usatim, timluxembourg, luxembourgcorporate, corporatemexico
Ethanolamine Analysis by GC/MS using a Zebron™ ZB-5MSPLUS ™ GC Column
APPLICATION Ethanolamine Analysis by GC/MS using a Zebron™ ZB-5MSPLUS™ GC Column Matt Trass and Tim Anderson Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA Conditions Overview This application note describes the GC/MS analysis of ethanolamines. Testing of ethanolamine mixtures,…
Key words
phenomenex, phenomenexfilm, filmawesome, awesometrass, trassanderson, andersonmatt, mattethanolamines, ethanolaminesterms, termstriethanolamine, triethanolaminepuerto, puertorico, ricousa, usatim, timluxembourg, luxembourgthick
Drugs of Abuse Analysis by GC/MS using a Zebron™ ZB-5MSPLUS ™ GC Column
APPLICATION Drugs of Abuse Analysis by GC/MS using a Zebron™ ZB-5MSPLUS™ GC Column Matt Trass and Tim Anderson Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA Overview Conditions This application note describes the GC/MS analysis of 23 drugs of…
Key words
phenomenex, phenomenextrass, trassanderson, andersonmatt, mattterms, termspuerto, puertorico, ricousa, usatim, timluxembourg, luxembourgconditions, conditionscorporate, corporatemexico, mexicoviewed, viewedireland
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike