Fast GC PAH Analysis - ZB-PAH-EU and ZB-PAH-SeleCT
Brochures and specifications | 2020 | PhenomenexInstrumentation
Polycyclic aromatic hydrocarbons (PAHs) are persistent environmental pollutants regulated by agencies worldwide. Fast, reliable separation and quantitation of PAHs in environmental, food, and industrial samples are essential for public health monitoring and quality assurance. The development of specialized GC columns enhances throughput and accuracy, reducing analysis time while maintaining resolution of critical isomeric pairs.
This work evaluates two novel capillary GC columns—Zebron ZB-PAH-EU and ZB-PAH-SeleCT—for the analysis of the EU-15+1 and EPA-610 PAH panels. Key goals include:
Analyses were performed on a GC-MS system using helium carrier gas under constant pressure. Detailed conditions:
ZB-PAH-EU achieved up to 70 % faster analysis while fully resolving the EU-15+1 and EPA-610 PAH suites within ~50 min. Baseline separation was confirmed for all critical isomer pairs including benzo[b,j,k]fluoranthene and benzo[c]fluorene.
ZB-PAH-SeleCT exhibited enhanced selectivity, delivering complete resolution of chrysene and triphenylene and minimizing coelution interferences. Resolution of benzo[b] versus benzo[k] versus benzo[j] fluoranthene was markedly improved, ensuring accurate quantitation.
The elevated thermal stability of both phases (up to 360 °C) supports robust operation and extended column lifetimes, even under demanding temperature programs.
Advances in stationary phase chemistry will continue to enhance selectivity and thermal resilience. Integration with ultra-fast GC systems and high-resolution mass spectrometry will further shorten analysis times. Machine-learning algorithms for automated peak deconvolution and identification are emerging, enabling high-throughput PAH screening in complex matrices.
Field-portable GC-MS platforms equipped with optimized PAH columns may soon allow on-site environmental assessments.
The Zebron ZB-PAH-EU and ZB-PAH-SeleCT GC columns offer a powerful combination of speed, resolution, and thermal stability for routine PAH analysis. Their performance supports regulatory compliance and enhances laboratory productivity in environmental, food, and industrial applications.
No formal literature references were provided in the source document.
GC columns, Consumables
IndustriesEnvironmental, Food & Agriculture, Materials Testing
ManufacturerPhenomenex
Summary
Importance of Rapid and High-Resolution PAH Analysis
Polycyclic aromatic hydrocarbons (PAHs) are persistent environmental pollutants regulated by agencies worldwide. Fast, reliable separation and quantitation of PAHs in environmental, food, and industrial samples are essential for public health monitoring and quality assurance. The development of specialized GC columns enhances throughput and accuracy, reducing analysis time while maintaining resolution of critical isomeric pairs.
Objectives and Study Overview
This work evaluates two novel capillary GC columns—Zebron ZB-PAH-EU and ZB-PAH-SeleCT—for the analysis of the EU-15+1 and EPA-610 PAH panels. Key goals include:
- Assessing reduction of analysis time compared with conventional PAH methods
- Demonstrating baseline resolution of regulated PAHs, including challenging isomeric pairs
- Evaluating column thermal stability and selectivity improvements
Methodology and Instrumentation
Analyses were performed on a GC-MS system using helium carrier gas under constant pressure. Detailed conditions:
- Columns:
- ZB-PAH-EU: 30 m × 0.25 mm ID × 0.20 µm df, temperature limit up to 360 °C
- ZB-PAH-SeleCT: 40 m × 0.18 mm ID × 0.14 µm df, temperature limit up to 340 °C
- Injection: Split ratio 5:1 (ZB-PAH-EU) or 30:1 (ZB-PAH-SeleCT), injection port 320–330 °C, 1 µL sample
- Oven program:
- Start 45 °C (0.8 min), ramp to 200 °C at 45 °C/min, then to 226 °C at 3 °C/min, final to 320 °C at 10 °C/min (ZB-PAH-EU)
- Start 45 °C (0.8 min), ramp to 200 °C at 45 °C/min, to 265 °C at 3 °C/min, to 270 °C at 1 °C/min, final to 320 °C at 10 °C/min (ZB-PAH-SeleCT)
- Detector: Mass selective detector (MSD) in full-scan (50–500 m/z) or SIM mode with target ions for PAH quantitation
Main Results and Discussion
ZB-PAH-EU achieved up to 70 % faster analysis while fully resolving the EU-15+1 and EPA-610 PAH suites within ~50 min. Baseline separation was confirmed for all critical isomer pairs including benzo[b,j,k]fluoranthene and benzo[c]fluorene.
ZB-PAH-SeleCT exhibited enhanced selectivity, delivering complete resolution of chrysene and triphenylene and minimizing coelution interferences. Resolution of benzo[b] versus benzo[k] versus benzo[j] fluoranthene was markedly improved, ensuring accurate quantitation.
The elevated thermal stability of both phases (up to 360 °C) supports robust operation and extended column lifetimes, even under demanding temperature programs.
Benefits and Practical Applications
- Increased sample throughput via shorter run times
- Accurate quantitation and reliable regulatory compliance for environmental and food monitoring
- Reduced false positives through superior isomer separation
- Extended column lifetime under elevated temperatures
Future Trends and Possibilities
Advances in stationary phase chemistry will continue to enhance selectivity and thermal resilience. Integration with ultra-fast GC systems and high-resolution mass spectrometry will further shorten analysis times. Machine-learning algorithms for automated peak deconvolution and identification are emerging, enabling high-throughput PAH screening in complex matrices.
Field-portable GC-MS platforms equipped with optimized PAH columns may soon allow on-site environmental assessments.
Conclusion
The Zebron ZB-PAH-EU and ZB-PAH-SeleCT GC columns offer a powerful combination of speed, resolution, and thermal stability for routine PAH analysis. Their performance supports regulatory compliance and enhances laboratory productivity in environmental, food, and industrial applications.
Reference
No formal literature references were provided in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast GC PAH Analysis by ZB-PAH-EU and ZB-PAH-SeleCT
2020|Phenomenex|Brochures and specifications
Unleash the Power of Fast GC PAH Analysis ZB-PAH-EU • Up to 70 % Faster PAH Analysis • Elevated Temperature Stability (340/360 °C) ZB-PAH-SeleCT • Enhanced Resolution for Chrysene and Triphenylene (PAH Interferences) • Increased Benzo[b,k] fluoranthene Separation www.phenomenex.com/GCPAH ZB-PAH-EU…
Key words
zebron, zebronpah, pahbenzo, benzophenomenex, phenomenexfluoranthene, fluoranthenefilter, filtertaper, taperindicator, indicatorliner, linerclick, clickgas, gaschrysene, chryseneelectronic, electronicpahs, pahsyour
Optimal Separation of PAH Compounds including Chrysene & Triphenylene using Zebron™ ZB-PAH-SeleCT GC column
2020|Shimadzu|Applications
TN-2094 APPLICATIONS Ramkumar Dhandapani, Ph.D. Product Manager - Gas Chromatography TN-2094 Optimal Separation of PAH Compounds including Chrysene & Triphenylene using Zebron™ ZB-PAH-SeleCT GC column He has a PhD in Analytical Chemistry and over 15 years experience in chromatographic method…
Key words
zebron, zebrontriphenylene, triphenylenepah, pahchrysene, chryseneselect, selectreproducing, reproducinglove, lovetrouble, troubletaper, tapermeter, meterelabaji, elabajirola, rolaresolution, resolutionpahs, pahsdhandapani
Rapid Analysis of 24 Regulated Polycyclic Aromatic Hydrocarbons (PAHs) on a Zebron™ ZB-PAH-EU GC Column
2020|Phenomenex|Applications
TN-2092 APPLICATIONS Ramkumar Dhandapani, Ph.D. Product Manager - Gas Chromatography Rapid Analysis of 24 Regulated Polycyclic Aromatic Hydrocarbons (PAHs) on a Zebron™ ZB-PAH-EU GC Column Ramkumar Dhandapani, Rola El abaji, Zachary Woodward and Zeshan Aqeel Phenomenex, Inc., 411 Madrid Ave.,…
Key words
benzo, benzozebron, zebronfluoranthene, fluoranthenepah, pahfluorene, fluoreneanthracene, anthracenepyrene, pyreneabundance, abundancereproducing, reproducinglove, lovetaper, taperrelative, relativeliner, linertrouble, troublebenz
Fast and Accurate GC-MS Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) in Rubber and Plastic using a Zebron™ ZB-PAH-EU GC Column
2020|Phenomenex|Applications
TN-2091 APPLICATIONS Ramkumar Dhandapani, Ph.D. Product Manager - Gas Chromatography Fast and Accurate GC-MS Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) in Rubber and Plastic using a Zebron™ ZB-PAH-EU GC Column He has a PhD in Analytical Chemistry and over 15…
Key words
zebron, zebronbenzo, benzopah, pahfluoranthene, fluoranthenepyrene, pyrenerubber, rubberplastic, plasticreproducing, reproducinganthracene, anthracenelove, lovetrouble, troubleliner, linerbenz, benztaper, tapermeter