Validation of Volatile Organic Compound by USEPA Method 8260C
Applications | 2011 | Teledyne LABSInstrumentation
Volatile organic compounds (VOCs) are widely monitored in environmental matrices due to their toxicity, persistence, and regulatory importance. Reliable quantification in water and soil supports risk assessment and remediation efforts.
This study validates USEPA Method 8260C for VOC analysis in water and soil by combining automated purge-and-trap sample preparation with GC/MS detection. Calibration ranges of 0.5–200 ppb (water) and 1–200 ppb (soil) were established alongside method detection limits (MDLs).
The automated Atomx purge-and-trap system paired with Agilent 7890/5975 GC/MS reliably quantifies VOCs in water and soil in full compliance with USEPA Method 8260C. Calibration linearity, low MDLs, and enhanced throughput demonstrate a robust solution for environmental laboratories.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, Teledyne LABS
Summary
Significance of the Topic
Volatile organic compounds (VOCs) are widely monitored in environmental matrices due to their toxicity, persistence, and regulatory importance. Reliable quantification in water and soil supports risk assessment and remediation efforts.
Objectives and Overview
This study validates USEPA Method 8260C for VOC analysis in water and soil by combining automated purge-and-trap sample preparation with GC/MS detection. Calibration ranges of 0.5–200 ppb (water) and 1–200 ppb (soil) were established alongside method detection limits (MDLs).
Methodology and Instrumentation
- Sample Preparation: Atomx automated system integrating purge-and-trap concentrator and multi-matrix autosampler processed 5 mL water or 5 g soil vials, with programmed internal standard addition and wash cycles.
- Gas Chromatography: Agilent 7890A with Restek RTX-624 column (20 m × 0.18 mm × 1 µm), helium carrier gas; oven ramp from 40 °C to 200 °C in ~15 min.
- Mass Spectrometry: Agilent 5975C inert XL MSD, source at 230 °C, quadrupole at 150 °C, scan range m/z 35–270 at ~5.8 scans/sec.
- Calibration: Serial dilutions in deionized water (0.5–200 ppb) and soil extract (1–200 ppb) with 25 ppb internal standard added automatically in 5 µL increments.
- MDL Determination: Seven replicates at low spike levels (0.5 ppb water, 1.0 ppb soil) to calculate MDLs per EPA guidelines.
Main Results and Discussion
- Calibration Performance: All target analytes exhibited linear response and relative standard deviations generally below 15%, in accordance with Method 8260C criteria.
- Method Detection Limits: MDLs were typically under 0.3 ppb for both matrices, with several compounds achieving sub-0.1 ppb sensitivity.
- Chromatographic Analysis: Total ion chromatograms at 20 ppb demonstrated clear separation of VOCs in both water and soil extracts, confirming method robustness.
- Automation Impact: The Atomx system enabled continuous sample processing with minimal manual intervention, improving reproducibility and throughput.
Benefits and Practical Applications
- High Throughput: An 80-position autosampler allows sequential analysis of multiple matrices without user intervention.
- Operational Efficiency: Automated standard and surrogate addition plus wash steps reduce preparation time and error.
- Matrix Versatility: System accommodates water, soil, and methanol extracts, supporting varied environmental monitoring needs.
- Regulatory Compliance: Fully meets USEPA Method 8260C performance requirements for QA/QC applications.
Future Trends and Opportunities
- Enhanced Digital Integration: Real-time data processing and remote monitoring to accelerate decision-making.
- Expanded Matrix Scope: Adaptation for air sampling and complex waste analyses using similar purge-and-trap workflows.
- Advanced Detection: Coupling with high-resolution mass spectrometry for broader VOC screening and identification.
- Sustainability: Development of energy-efficient instrumentation and reduced solvent strategies to minimize laboratory footprint.
Conclusion
The automated Atomx purge-and-trap system paired with Agilent 7890/5975 GC/MS reliably quantifies VOCs in water and soil in full compliance with USEPA Method 8260C. Calibration linearity, low MDLs, and enhanced throughput demonstrate a robust solution for environmental laboratories.
References
- USEPA Method 8260C: Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry, Revision 3, August 2006.
- USEPA Method 5030: Purge-And-Trap for Aqueous Samples, Revision 3, May 2003.
- USEPA Method 5035: Closed-System Purge-And-Trap and Extractions for Volatile Organics in Soil and Waste Samples, Revision 1, July 2002.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS
2015|PerkinElmer|Applications
Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS Application Note By Tyler Trent Abstract In order to determine the concentration of volatile organic compounds (VOCs) in water and soil matrices 1 2 3 the…
Key words
temp, tempbake, bakepurge, purgerinse, rinsecondensate, condensatesweep, sweeptime, timedesorb, desorbsoil, soilflow, flowmdl, mdlwater, waterprepurge, prepurgeneedle, needlespike
Volatile Organic Compounds in Every Day Food
2011|Agilent Technologies|Applications
Volatile Organic Compounds in Every Day Food Application Note Abstract Volatile Organic Compounds (VOCs) have been analyzed in drinking water and pharmaceuticals extensively in recent years. Now, VOCs in food products have became a major interest among researchers. Utilizing an…
Key words
temp, tempppb, ppbacetone, acetoneacetate, acetatepurge, purgedesorbs, desorbsbaby, babycheese, cheeseisopropyltoluene, isopropyltolueneconcentration, concentrationcompound, compoundbake, bakefood, fooddisulfide, disulfidecondensate
USEPA Method 8260C using the Tekmar Atomx Automated VOC Sample Prep System and a Thermo Focus/DSQ™ II GCMS
2010|Thermo Fisher Scientific|Applications
USEPA Method 8260C using the Tekmar Atomx Automated VOC Sample Prep System and a Thermo Focus/DSQ™ II GCMS Application Note By Roger Bardsley Abstract The United States Environmental Protection Agency (USEPA) developed Method 8260 to determine the concentration of Volatile…
Key words
temp, temppurge, purgebake, bakerinse, rinsedesorb, desorbcondensate, condensatesweep, sweepdsq, dsqsoil, soiltime, timewater, waterflow, flowprepurge, prepurgeneedle, needlemdl
Analysis of Volatile Organic Compounds in Different Beverages
2011|Agilent Technologies|Applications
Analysis of Volatile Organic Compounds in Different Beverages Abstract Volatile Organic Compounds (VOCs) have been a major concern of the U.S. Environmental Protection Agency (USEPA) leading strict regulations of VOCs present in drinking water. This study will focus on the…
Key words
mdl, mdlsoda, sodatemp, tempbottle, bottlepurge, purgedesorbs, desorbsbake, bakecondensate, condensatediet, dietpreheat, preheatcal, calbromodichloromethane, bromodichloromethanebrand, brandflow, flowtime