Agilent MassHunter Workstation Software – 7200 Accurate-Mass Quadrupole Time of Flight GC/MS - Familiarization Guide

Manuals | 2012 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/HRMS, GC/Q-TOF, Software
Industries
Manufacturer
Agilent Technologies

Summary

Význam tématu


The Agilent 7200 Q-TOF GC/MS system combines gas chromatography with high-resolution, accurate-mass quadrupole time-of-flight mass spectrometry to deliver untargeted qualitative and quantitative analyses. Its ability to resolve complex mixtures, improve compound identification confidence through sub-2 ppm mass accuracy, and optimize sensitivity via clear workflows makes it invaluable in environmental monitoring, forensic applications, pharmaceutical R&D, and quality control in chemical industries.

Cíle a přehled studie / článku


This familiarization guide aims to teach users how to configure the instrument, develop acquisition methods, perform mass calibrations, acquire and analyze high-resolution data, and generate reports. It presents step-by-step exercises from system preparation through qualitative data mining, providing both overview steps and detailed instructions.

Použitá metodika a instrumentace


Instrument configuration and sample setup
  • Agilent 7890 GC with split/splitless or MultiMode inlet
  • 10 µL fixed-needle autosampler syringe
  • J&W DB-35 ms column (30 m × 0.25 mm, 0.25 µm)
  • Agilent MassHunter Workstation Software: Acquisition, Qualitative Analysis, Quantitative Analysis
  • Agilent 7200 Q-TOF mass spectrometer
Workflow exercises
  • System check and sample vials preparation (standard mix in isooctane)
  • Method development tasks: inlet/injection parameters, GC configuration, TOF mass calibration, GC/MS parameter entry, MS scanning setup, optional data acquisition
  • Data analysis tasks: loading data, workflow configuration, deconvolution peak finding, accurate-mass library search, mass difference caliper, and report generation

Hlavní výsledky a diskuse


The guide demonstrates:
  • Rapid mass calibration yielding <2 ppm accuracy across calibration ions
  • Optimized inlet, oven, column, collision cell and autosampler settings for robust chromatography
  • Efficient deconvolution of coeluting peaks in the 7.5–9.5 min window to reveal individual spectra
  • Successful accurate-mass library matching of standard compounds such as dodecane, biphenyl, 4-chlorobiphenyl, methyl palmitate
  • Visualization of exact mass differences (e.g., Δ 78.0462 m/z) to support structural elucidation
  • Automated report creation summarizing acquisition metadata, integration results, and compound assignments

Přínosy a praktické využití metody


This structured approach ensures reproducibility and rapid operator training. It supports high confidence compound identification in complex matrices, streamlines method setup to decrease downtime, and integrates data processing tools to accelerate decision-making in R&D, QA/QC, and regulatory compliance.

Budoucí trendy a možnosti využití


Future directions include integration with machine-learning models for automated spectral deconvolution, cloud-based data management for collaborative research, expansion of high-resolution libraries for emerging contaminants, and development of real-time screening methods for process monitoring and high-throughput metabolomics.

Závěr


The Agilent 7200 Q-TOF GC/MS familiarization guide equips analysts with a clear, comprehensive workflow for instrument setup, calibration, data acquisition, and advanced qualitative analysis. Its emphasis on accuracy, sensitivity, and software-driven automation enables laboratories to achieve reliable results and supports continuous innovation in analytical chemistry.

Reference


No external references were cited in the source document. Documentation provided by Agilent Technologies 2012 including the 7200 Q-TOF GC/MS Familiarization Guide and associated MassHunter user manuals.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent MassHunter Workstation Software – 7250 Accurate-Mass Quadrupole Time of Flight GC/MS - Familiarization Guide
Agilent MassHunter Workstation Software – 7250 Accurate-Mass Quadrupole Time of Flight GC/MS Familiarization Guide Before you begin 3 Prepare your system 3 Prepare the samples required for data acquisition 4 Exercise – Develop an acquisition method for the 7250 5…
Key words
select, selectexercise, exerciseclick, clicktab, tabcomments, commentsdata, datainstructions, instructionsfile, filesteps, stepsdetailed, detailededitor, editoranalyze, analyzemethod, methodmenu, menuicon
Agilent MassHunter Workstation Software – Data Acquisition for 7000 Series Triple Quad GC/MS - Familiarization Guide
Agilent MassHunter Workstation Software – Data Acquisition for 7000 Series Triple Quad GC/MS Familiarization Guide Before you begin 3 Prepare your system 3 Prepare the samples required for data acquisition 3 Exercise – Develop an acquisition method for the 7000…
Key words
comments, commentsselect, selectclick, clickinstructions, instructionsexercise, exercisesteps, stepsmethod, methoddetailed, detailedcreate, createdata, dataicon, iconsave, savefile, filequantitative, quantitativesequence
MassHunter Profinder Software - Quick Start Guide
MassHunter Profinder Software - Quick Start Guide
2017|Agilent Technologies|Manuals
MassHunter Profinder Software Quick Start Guide For Research Use Only. Not for use in diagnostic procedures. Getting Started 11 User Interface 12 Basic Profinder Workflow 22 Add sample files to your project 24 Align chromatograms 28 Edit and run your…
Key words
profinder, profinderfeature, featureisotopologue, isotopologueguide, guideextraction, extractionquick, quickstart, startyour, yourclick, clickbatch, batchworkflow, workflowproject, projectrecursive, recursiveyou, youfile
Agilent MassHunter Mass Profiler Software - Familiarization Guide
MassHunter Mass Profiler Software Familiarization Guide Process Data 2 Create a project 3 Edit a method 6 Perform a PCA analysis 13 Apply a different method 18 Edit a project 18 View Results 20 View results in the Feature Table…
Key words
profiler, profilerfamiliarization, familiarizationfeature, featureproject, projectguide, guidefeatures, featuresexercises, exercisesmass, masstable, tableclick, clickview, viewparameters, parameterspca, pcareprocesses, reprocessesexport
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike