Determination of diacetyl (butanedione) & pentanedione in beer by HS-GC

Applications | 2012 | Agilent TechnologiesInstrumentation
GC, HeadSpace
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Beer fermentation naturally produces diacetyl and 2,3-pentanedione, which can impart undesirable off-flavors such as rancidity or buttery notes. Accurate monitoring of these compounds is vital for ensuring product quality, consumer satisfaction, and compliance with flavor standards.

Objectives and Overview of Study


This application explores a headspace gas chromatography method coupled with an electron capture detector to quantify diacetyl and 2,3-pentanedione in beer. The primary goals are to establish a sensitive, reproducible, and linear analytical protocol across concentrations relevant to brewing quality control.

Methodology


Sample preparation involves transferring 5 mL of beer or standard solution into 20 mL headspace vials. Standards are prepared in 4% ethanol. Headspace extraction is performed at 50 °C for 30 minutes with a 100 °C transfer loop, followed by injection of 1 mL of headspace gas into the GC.
Chromatographic separation uses a DB-5 capillary column (60 m × 0.53 mm × 5 µm) under a temperature program from 45 °C (2 min) to 150 °C at 10 °C/min. Carrier gas is high-purity nitrogen at 10 mL/min, and detection is achieved by a µECD at 150 °C.

Used Instrumentation


  • Agilent 7697A Headspace Autosampler with 111-position tray
  • Agilent 7890A Gas Chromatograph equipped with µECD detector
  • High-purity nitrogen as carrier and make-up gas

Main Results and Discussion


  • Calibration curves for diacetyl and 2,3-pentanedione exhibit excellent linearity from 0.25 to 500 µg/L.
  • Repeatability (n=6) yielded RSD values of 2.4% for diacetyl and 1.2% for 2,3-pentanedione peak areas.
  • Retention time stability was exceptional with RSD <0.01%.
  • Analysis of commercial beer samples revealed diacetyl concentrations ranging from 7.85 to 40.80 µg/L and 2,3-pentanedione from 6.41 to 19.82 µg/L, reflecting differences in fermentation and processing conditions.

Benefits and Practical Applications


This headspace GC–µECD method provides high sensitivity, precision, and robust performance, making it ideal for routine brewery quality control and research into fermentation dynamics. It supports consistent flavor profiling and regulatory compliance.

Future Trends and Potential Applications


  • Automation of headspace sampling for real-time fermentation monitoring.
  • Extension to other volatile by-products for comprehensive flavor fingerprinting.
  • Coupling with mass spectrometry to enhance compound identification.
  • Integration into sensory studies to optimize beer flavor and reduce off-flavors.

Conclusion


The described HS-GC–µECD approach offers a reliable, linear, and precise solution for quantifying diacetyl and 2,3-pentanedione in beer. It underpins effective quality assurance and process optimization in brewing operations.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Flavor Compounds in Beer using the Integrated Agilent 8697 Headspace Sampler with the Agilent 8890 GC System
Application Note Food Testing and Agriculture Analysis of Flavor Compounds in Beer using the Integrated Agilent 8697 Headspace Sampler with the Agilent 8890 GC System Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd., Shanghai 200131 P. R. China Abstract This…
Key words
butanedione, butanedionebeer, beerethyl, ethylecd, ecdpentanedione, pentanedioneflavor, flavorcompounds, compoundsacetate, acetateisoamylol, isoamylolformate, formateisobutyraldehyde, isobutyraldehydestock, stockconcentration, concentrationcaprylate, caprylateused
Analysis of Diacetyl and 2,3-pentanedione in Beer Using Nexis™ GC-2030
Application News Analysis of Diacetyl and 2,3-pentanedione in Beer Using Nexis™ GC-2030 „ Chromatogram and Calibration Curves of Standard Samples The chromatogram and calibration curves of standard samples are shown in Fig. 1 and Fig. 2, respectively. A. Hashimoto, Y.…
Key words
vdk, vdkdiacetyl, diacetylfermentation, fermentationbottling, bottlingbeer, beermain, mainchromatogram, chromatogramquantitative, quantitativepressurization, pressurizationyeasts, yeastsbefore, beforesamples, samplesratio, ratiovial, vialafter
Analysis of Diacetyl and 2,3-Pentanedione in Beer Using Brevis GC-2050 (Carrier Gas: N2)
Gas Chromatograph HS-20 NX/Brevis™ GC-2050 Application News Analysis of Diacetyl and 2,3-Pentanedione in Beer Using Brevis™ GC-2050 (Carrier Gas: N2) Akara Hashimoto1, Suzuki Arisa1, Shiro Yamada2 1 Shimadzu Corporation, 2 Far Yeast Brewing Company User Benefits  Quantitation of diacetyl…
Key words
diacetyl, diacetylvdk, vdkfermentation, fermentationafter, aftersecondary, secondarybeer, beeraftersecondary, aftersecondaryvdks, vdksbefore, beforeratio, ratiohopping, hoppingdry, drystandard, standardtreatment, treatmentnews
Simultaneous Analysis of Vicinal Diketones, Volatile Compounds and Ethanol Content in Beer with a Single Headspace Injection (Shimadzu Beer 3-in-1 Analyzer)
No. SSI-GC-2006 Gas Chromatography No. GC-2008 Simultaneous Analysis of Vicinal Diketones, Volatile Compounds and Ethanol Content in Beer with a Single Headspace Injection (Shimadzu Beer 3-in-1 Analyzer) ■ Introduction The chemistry of beer brewing is quite complex. Numerous compounds are…
Key words
vdk, vdkvoc, vocacetate, acetateethanol, ethanolabv, abvisoamyl, isoamylamyl, amylalcohol, alcoholiso, isoacetaldehyde, acetaldehydestabilwax, stabilwaxethyl, ethylbeer, beerratio, ratiovdks
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike