Organic Solvents
Applications | | GL SciencesInstrumentation
Accurate measurement of volatile organic solvents is essential for environmental monitoring, pharmaceutical quality control, food safety and industrial process validation. Gas chromatography with inert surface treatments minimizes analyte adsorption and tailing, improving sensitivity and reproducibility when quantifying low-level residual solvents.
This study evaluates the performance of an InertCap® 1301 capillary column in separating 35 common organic solvents. The goals are to demonstrate baseline resolution, inertness toward active compounds, and robust analytical conditions suitable for routine GC/FID applications.
A mixed standard containing 35 analytes ranging from methanol and acetone to xylene isomers and glycol ethers was prepared in equal proportions. Key chromatographic parameters:
The InertCap 1301 column achieved baseline separation of all 35 analytes, including challenging pairs such as isomeric xylenes and glycol ether derivatives. Peak shapes for polar compounds (alcohols, ketones, esters) exhibited minimal tailing, indicating effective deactivation of active sites. Retention times were highly reproducible, with relative standard deviations below 0.5 %. The inert surface coating prevented adsorption of reactive analytes, resulting in improved sensitivity, especially at low concentration levels.
Using an inert GC column offers several advantages for routine solvent analysis:
These features make the method suitable for residual solvent testing in pharmaceuticals, VOC emission monitoring in environmental laboratories, and quality assurance in food and packaging industries.
Advances in column surface chemistry and ultra-inert deactivation will further improve analysis of highly active compounds. Integration with mass spectrometric detection and automated sample handling will enable high-throughput screening. Development of greener carrier gases and reduced energy consumption in temperature programming will align with sustainable laboratory practices.
The InertCap 1301 column demonstrates superior inertness and resolution for a broad range of organic solvents under GC/FID conditions. Its reliable performance and ease of use support diverse analytical tasks in environmental, pharmaceutical and industrial settings.
GL Sciences Inc. InertSearch for GC™ InertCap® Applications – Organic Solvents, Data No. GA138-0644
GC, GC columns, Consumables
IndustriesManufacturerGL Sciences
Summary
Significance of Organic Solvent Analysis
Accurate measurement of volatile organic solvents is essential for environmental monitoring, pharmaceutical quality control, food safety and industrial process validation. Gas chromatography with inert surface treatments minimizes analyte adsorption and tailing, improving sensitivity and reproducibility when quantifying low-level residual solvents.
Objectives and Overview
This study evaluates the performance of an InertCap® 1301 capillary column in separating 35 common organic solvents. The goals are to demonstrate baseline resolution, inertness toward active compounds, and robust analytical conditions suitable for routine GC/FID applications.
Methodology and Experimental Conditions
A mixed standard containing 35 analytes ranging from methanol and acetone to xylene isomers and glycol ethers was prepared in equal proportions. Key chromatographic parameters:
- Instrumentation: GC/FID system
- Column: InertCap 1301, 0.25 mm I.D. × 60 m, film thickness 1.0 μm (Cat. No. 1010-60165)
- Carrier gas: Helium at 160 kPa
- Injection: Split mode, 100 mL/min, injector temperature 200 ℃, sample volume 0.2 μL
- Oven program: 40 ℃ hold for 5 min, ramp 5 ℃/min to 200 ℃
- Detection: Flame ionization detector, temperature 200 ℃, sensitivity range 10¹
Used Instrumentation
- Gas chromatograph with split/splitless injector and FID
- InertCap® 1301 fused-silica capillary column
- High-purity helium supply
- Precision temperature controller
Main Results and Discussion
The InertCap 1301 column achieved baseline separation of all 35 analytes, including challenging pairs such as isomeric xylenes and glycol ether derivatives. Peak shapes for polar compounds (alcohols, ketones, esters) exhibited minimal tailing, indicating effective deactivation of active sites. Retention times were highly reproducible, with relative standard deviations below 0.5 %. The inert surface coating prevented adsorption of reactive analytes, resulting in improved sensitivity, especially at low concentration levels.
Benefits and Practical Applications
Using an inert GC column offers several advantages for routine solvent analysis:
- Enhanced peak symmetry and reduced tailing for polar and reactive species
- Lower detection limits supporting trace-level quantification
- Wider analytical window to resolve complex mixtures
- Robust performance over repeated thermal cycles and sample injections
These features make the method suitable for residual solvent testing in pharmaceuticals, VOC emission monitoring in environmental laboratories, and quality assurance in food and packaging industries.
Future Trends and Applications
Advances in column surface chemistry and ultra-inert deactivation will further improve analysis of highly active compounds. Integration with mass spectrometric detection and automated sample handling will enable high-throughput screening. Development of greener carrier gases and reduced energy consumption in temperature programming will align with sustainable laboratory practices.
Conclusion
The InertCap 1301 column demonstrates superior inertness and resolution for a broad range of organic solvents under GC/FID conditions. Its reliable performance and ease of use support diverse analytical tasks in environmental, pharmaceutical and industrial settings.
Reference
GL Sciences Inc. InertSearch for GC™ InertCap® Applications – Organic Solvents, Data No. GA138-0644
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Residual solvents in pharmaceuticals
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications R id l solvents Residual l t in i pharmaceuticals h ti l Data No. GA123-0850 180000 160000 140000 66 64 56 120000 63 58+59 100000 Acetone 80000 60000 40000 20000 0 0 アセトン…
Key words
alcohol, alcoholbutyl, butylmethyl, methylether, etherketone, ketoneethyl, ethylacetate, acetatepropyl, propylisopropyl, isopropylcellosolve, cellosolveacetic, aceticキシレン, キシレンester, esterisobutyl, isobutylacid
Organic solvents
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organic solvents Data No. GA140-851 14+15 200000 11 7 13 6 100000 2 16 8 10 4 12 9 1 3 17 5 0 0 2 Conditions Methyl alcohol 4 6 Time (min) 8…
Key words
xylene, xylenebutyl, butylキシレン, キシレンalcohol, alcoholvol, volcyclohexane, cyclohexaneエチレングリ, エチレングリチルアセテ, チルアセテブチルベンゼン, ブチルベンゼンルモノブチルエ, ルモノブチルエsolvents, solventsモノブチルグリコールエーテル, モノブチルグリコールエーテルイソブタノール, イソブタノールメチルイソブチルケトン, メチルイソブチルケトンester
Organic solvents
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organic solvents Data No. GA023-0850 14000 1 12000 10000 9 13 12 14 15 17 8000 6000 3 6 4 4000 11 2 78 10 166 5 19 18 2000 0 10 Acetic acid…
Key words
キシレン, キシレンacetate, acetatebutyl, butylalcohol, alcoholアニソール, アニソールテトラヒドロフラン, テトラヒドロフランヘプタン, ヘプタンethyl, ethylエチルベンゼン, エチルベンゼンアセトン, アセトンエタノール, エタノールanisole, anisoleacetic, aceticxylene, xyleneheptane
Organic solvents
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organic Solvents Data No. GA202-0410 40000 22 12+13 GC GA202 TC-BOND Q Methanol メタノール Methanol Methyl alcohol メタノール メチルアルコール 67-56-1 GC GA202 TC-BOND Q Ethanol エタノール Ethanol Ethyl alcohol エタノール エチルアルコール 64-17-5 GC GA202…
Key words
ヘキサン, ヘキサンuvolt, uvoltinertsearch, inertsearchinertcap, inertcapsolvebts, solvebtsブタノール, ブタノールenviroment, enviromentエチルエーテル, エチルエーテルジエチルエーテル, ジエチルエーテルメチルプロピルアルコール, メチルプロピルアルコールglycol, glycoltert, tertイソプロピルアセテート, イソプロピルアセテートテトラヒドロフラン, テトラヒドロフランポーラスポリマー