GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Applications | 2011 | Agilent TechnologiesInstrumentation
GC, HeadSpace
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Headspace gas chromatography is widely used in pharmaceutical analysis to detect and quantify residual solvents at trace levels. Precise control of vial and sample loop pressures can improve analytical sensitivity and reproducibility, which is vital for meeting stringent regulatory standards such as USP <467>.

Objectives and Study Overview


This application note evaluates different vial pressurization and loop fill schemes on the Agilent 7697A Headspace Sampler using aqueous solutions of USP <467> residual solvents (Class 1, 2A and 2B) at their limit concentrations. Two pressurization modes (“Flow to Pressure” and “To Pressure”) are compared for their impact on signal response and repeatability, especially for Class 2A solvents.

Methodology


Samples of residual solvents in water were prepared at USP limit concentrations and transferred into 20 mL headspace vials. Key parameters included:
  • Oven equilibration: 85 °C for 40 min
  • Vial pressurization gas: helium
  • Pressurization modes: flow-limited (50 mL/min) vs. fast fill (200 mL/min)
  • Loop fill rate: 20 psi/min up to set pressures
  • Sample loop volume: 1 mL
  • Final vial pressures tested: up to 15 psi for Class 1/2B and 20 psi for Class 2A

Instrumentation


The study employed an Agilent 7697A Headspace Sampler interfaced to an Agilent 7890A GC with FID detection. Key components:
  • Pressure Control Module (PCM) for forward and backpressure regulation
  • 0.45 mm i.d. deactivated fused silica headspace transfer line
  • Split/splitless inlet at 150 °C, helium carrier gas
  • VF-624 ms column (30 m × 0.25 mm × 1.4 μm)

Main Results and Discussion


Increasing the sample loop pressure above ambient produced a nearly linear increase in peak area for most solvents, consistent with the ideal gas law. Solvents with low headspace partition coefficients (k) exhibited the steepest slopes. Exceptions included 1,4-dioxane (non-ideal behavior) and carbon tetrachloride (minimal area enhancement). Comparisons of “Flow to Pressure” vs. “To Pressure” modes for Class 2A solvents revealed:
  • Similar repeatability (RSD < 2%) for most analytes
  • Significantly improved RSD (from ~5% to ~1.5%) for low-k solvents (cyclohexane, methylcyclohexane) using the “To Pressure” mode

Practical Benefits and Applications


Operating the headspace sampler under elevated loop pressures enhances method sensitivity by increasing the moles of analyte introduced into the GC. Improved repeatability simplifies regulatory compliance, lowers detection limits, and supports robust QA/QC in pharmaceutical and other industries requiring trace volatile analysis.

Future Trends and Applications


As regulatory limits for trace impurities tighten, optimized headspace pressurization schemes will become increasingly important. Future developments may integrate dynamic pressure programming, automated method selection based on solvent properties, and coupling with mass spectrometric detection to further extend sensitivity and selectivity.

Conclusion


The Agilent 7697A headspace sampler’s advanced pressure control enables precise vial pressurization and loop filling. Raising the loop pressure above ambient systematically boosts sensitivity and enhances repeatability. The “To Pressure” mode, in particular, offers superior performance for challenging analytes with low headspace partitioning.

References


1. Firor RL. Analysis of USP <467> Residual Solvents with Improved Repeatability using the Agilent 7697A Headspace Sampler. Agilent Publication 5990-7625EN, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467>
System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467> Technical Overview Introduction The Agilent 7697A Headspace Sampler introduced onboard electronic pneumatics control and several instrument control enhancements to improve performance over earlier models…
Key words
vial, vialloop, looppressure, pressurefill, fillpressurization, pressurizationprocedure, procedurecontrol, controlsettings, settingsflow, flowshaking, shakingmin, minresidual, residualamount, amountdefault, defaultmode
USP<467> residual solvents
USP<467> residual solvents
2018|Agilent Technologies|Applications
Application Note Pharmaceuticals USP<467> residual solvents Applying the Agilent 5977A MSD with the Agilent 7697A headspace sampler and Agilent 7890B GC Authors Roger L. Firor and Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808-1610 USA Abstract The…
Key words
vial, vialheadspace, headspacesolvents, solventsresidual, residualloop, loopacquisition, acquisitiontetralin, tetralintetrahyrofuran, tetrahyrofurannitromethane, nitromethanesim, simmethylcyclohexane, methylcyclohexanecounts, countspressure, pressurexylene, xyleneusp
Agilent 7697A GC Headspace Sampler - Data Sheet
Agilent 7697A GC Headspace Sampler - Data Sheet
2019|Agilent Technologies|Brochures and specifications
Data Sheet Agilent 7697A GC Headspace Sampler Overview The Agilent 7697A headspace sampler is part of Agilent’s line of gas chromatography products. Based on the architecture of the market-leading GC and 7693A ALS line of products, the 7697A has been…
Key words
vial, vialsettable, settableheadspace, headspaceincrements, incrementspressurization, pressurizationcontrol, controlsampling, samplingsample, samplepressure, pressureuser, useroptional, optionalparameters, parametersepc, epcmenu, menuvia
Residual Solvents Analysis Using an Agilent Intuvo 9000 GC with 8697 Headspace Sampler
Application Note Pharmaceutical Small Molecules Residual Solvents Analysis Using an Agilent Intuvo 9000 GC with 8697 Headspace Sampler Author Ian Eisele Agilent Technologies, Inc. Abstract In accordance with United States Pharmacopeia (USP) <467>, residual solvent analysis was performed on an…
Key words
loop, loopwax, waxsolvents, solventsresidual, residualcpme, cpmevial, vialmethylisobutylketone, methylisobutylketonecyclopentyl, cyclopentylprocedure, procedureheadspace, headspaceinlet, inlettime, timechip, chipfill, fillxylene
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike