Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Applications | 2011 | Agilent TechnologiesInstrumentation
GC, GCxGC, GC/MSD, GC/SQ
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Chiral analysis of natural oils and botanical extracts is critical for quality assurance, origin determination, and detection of adulteration in high-value products such as fragrances, flavors, and nutraceuticals. The enantiomeric composition of terpenoid compounds like linalool or citronellol influences sensory properties, bioactivity, and regulatory compliance. Rapid, reliable multidimensional gas chromatographic methods that preserve column integrity and maximize throughput are increasingly demanded by analytical laboratories across industries.

Study Objectives and Overview


This application note demonstrates a two-oven multidimensional GC approach using a conventional air-bath oven coupled to a low thermal mass (LTM) Series II module and a CFT Deans switch. The goals are to:
  • Achieve baseline separation of linalool enantiomers and other chiral analytes in botanical matrices
  • Maintain optimal column conditions by limiting cyclodextrin column temperatures below 150 °C
  • Compare FID/MSD and dual-FID detection setups for routine and confirmatory analyses

Methodology and Instrumentation


Independent temperature control enables optimized programming of each column stage. Key elements include:
  • Primary separation: Agilent 7890A GC with 30 m × 0.25 mm × 0.25 µm DB-5ms column in constant-flow mode (1.3 mL/min)
  • Heart-cut valve: CFT Deans switch to selectively transfer target fractions
  • Secondary separation: LTM Series II 5″ direct-heated module fitted with 30 m × 0.25 mm × 0.25 µm CycloDex B chiral column (constant flow 2.4–2.5 mL/min)
  • Detectors:
    • FID/MSD configuration (Agilent 5975C MSD) for structural confirmation
    • Dual-FID arrangement for high-throughput screening

Split/splitless injection at 280 °C, split ratios of 200:1–400:1, and tailored oven programs (primary: 70 °C to 280 °C @ 10 °C/min; LTM: 70 °C to 130 °C @ 2 °C/min) balance resolution, analysis time, and column lifetime.

Main Results and Discussion


Initial single-column runs on DB-5ms showed co-elution of linalool enantiomers in geranium extract. Heart-cut to the CycloDex B column yielded near-baseline separation of R and S linalool isomers within a narrow time window (8.25–8.38 min), confirmed by MSD spectra. Similar enantiomeric ratios were observed in natural extracts of balm leaf, cypress, jasmine, and orange oil, with orange oil exhibiting near-R-pure linalool. Multi-cut approaches permitted simultaneous analysis of citronellol, nerol, and related terpenoids in a single injection. The dual-FID setup reduced method cycle time by nearly 30% with minimal loss of chiral resolution, suitable for routine QC.

Benefits and Practical Applications


  • Enhanced flexibility: Independent oven control allows optimal temperature gradients for chiral and achiral columns without thermal degradation.
  • Improved throughput: Heart-cutting concentrates instrument time on analytes of interest, reducing overall cycle time.
  • Robust quality control: Enantiomeric profiling distinguishes synthetic adulterants from natural sources.
  • Modularity: Simple column and detector swaps adapt the platform to diverse analytical challenges beyond chiral separations.

Future Trends and Applications


Emerging developments include integration of enhanced data-driven method translation tools to automate parameter optimization, coupling LTM modules with fast quadrupole MS for real-time chiral screening, and expansion into other stereoisomeric analyses (e.g., amino acids, pharmaceuticals). Advances in polymeric chiral phases may further lower bleed and extend column lifetimes under fast-ramp LTM conditions. There is also potential to incorporate automated heart-cut scheduling software for unattended, high-capacity laboratories.

Conclusion


The LTM Series II GC system with Deans switch heart-cutting offers a powerful, flexible platform for enantiomeric analysis of natural oils and extracts. By decoupling oven programs and employing chiral CycloDex columns at moderated temperatures, analysts gain improved resolution, faster cycle times, and enhanced column longevity. Both FID/MSD and dual-FID configurations serve complementary roles from in-depth structural confirmation to routine QC screening.

References


  1. Agilent Technologies. Method Translation Software. www.chem.agilent.com.
  2. David F., Klee M. Independent Column Temperature Control using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds. Application Note 5990-3428EN.
  3. Agilent Technologies. Deans Switch Calculator, Agilent Capillary Flow Technology Accessories CD.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds
Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds Application Note Food and Flavors Author Abstract Frank David An improved chiral separation of the enantiomers of linalool using a multidimensional Research Institute for…
Key words
linalool, linaloolltm, ltmenantiomers, enantiomersheart, heartchiral, chiralcyclodextrin, cyclodextrindeans, deanstemperature, temperaturemultidimensional, multidimensionallavender, lavenderperfume, perfumeresponse, responsecut, cutcolumn, columnswitch
Analysis of Suspected Flavor and Fragrance Allergens in Perfumes Using Two-Dimensional GC with Independent Column Temperature Control Using an LTM Oven Module
Analysis of Suspected Flavor and Fragrance Allergens in Perfumes Using Two-Dimensional GC with Independent Column Temperature Control Using an LTM Oven Module Application Note Food and Flavors Authors Abstract Frank David Several different analytical methods based on GC/MS are used…
Key words
ltm, ltmallergens, allergensfragrance, fragrancecinnamaldehyde, cinnamaldehydebenzyl, benzyldeans, deansalcohol, alcoholheartcutting, heartcuttingoven, ovenflavor, flavorallergen, allergendimensional, dimensionalswitch, switchcosmetic, cosmeticsolutes
Simple Heart Cutting with Deans Switch and Backflushing with Switch and Backflushing with Capillary Flow Technolgy
Simple Heart Cutting with Deans Switch and Backflushing with Capillary Flow Technolgy Simon Jones Applications pp Engineer g GC or GC/MS Analysis in Complex Matrices • In complex sample matrices, there are often too many overlapping pp g compounds p…
Key words
backflush, backflushrestrictor, restrictormidrofluidic, midrofluidicinnowax, innowaxbenzene, benzenepcm, pcmfid, fidcolumn, columndeans, deansinlet, inletdevices, devicessolenoid, solenoidrestricted, restrictedpuu, puuethanol
Agilent 7890B Gas Chromatograph
Agilent 7890B Gas Chromatograph
2018|Agilent Technologies|Brochures and specifications
Resolve Your Search For Reliability Agilent 7890B Gas Chromatograph Resolve Your Search For The Next Evolution In Gas Chromatography At Agilent, we believe that building the world’s most trusted GC system is an ongoing process. With every step, we increase…
Key words
your, yourbackflush, backflushagilent, agilentltm, ltmdetector, detectormsd, msdinert, inertethanol, ethanolinlet, inletcolumn, columnmmi, mmibuoh, buohflow, flowcft, cftfid
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike