GC/μECD Analysis of Chlorinated Plaguicides Using Agilent J&W HP-1ms Ultra Inert and Agilent J&W DB-1301 Capillary GC Columns

Applications | 2009 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Chlorinated plaguicides represent a major class of persistent agrochemicals used worldwide to control pests. Their widespread use and tendency to contaminate water systems raise significant environmental and health concerns. Sensitive and reliable analytical methods are essential to monitor trace levels of these toxic compounds in surface and groundwater and to meet stringent regulatory limits such as those outlined by the EU Water Framework Directive.

Objectives and Study Overview


This study develops and validates a dual-column gas chromatography method coupled with micro-electron capture detection for quantifying low-level chlorinated plaguicides. The dual arrangement uses an Agilent J&W HP-1ms Ultra Inert column for primary analysis and a DB-1301 column for confirmation, enabling robust quantitation down to sub-picogram levels.

Methodology and Instrumentation


The analytical platform comprises an Agilent 7890A GC with dual micro-ECDs and an Agilent 7683B autosampler. Key chromatographic conditions include splitless injection at 250 °C, hydrogen carrier gas, and a temperature program from 125 to 280 °C. A capillary flow technology two-way splitter directs the flow to both columns simultaneously. The primary column is a 30 m × 0.25 mm, 1.0 µm HP-1ms Ultra Inert, chosen for its exceptional inertness and low bleed. The confirmation column is a 30 m × 0.25 mm, 1.0 µm DB-1301 phase. Sample preparation involved serial dilutions of organochlorine pesticide standards in 2-2-4-trimethylpentane to yield calibration levels from 1 to 40 ng/mL.

Main Results and Discussion


Calibration across six concentration levels from 1 to 40 ng/mL produced a linear response with R2 >= 0.998 on both columns. The Ultra Inert column delivered sharp, symmetrical peaks with minimal tailing at trace loads of 0.25 pg on column, yielding signal-to-noise ratios greater than 11. Tailoring the temperature program and using the inert column minimized analyte adsorption and column bleed, enhancing sensitivity. Dual-column confirmation improved analyte identification and quantitation confidence.

Benefits and Practical Applications


  • High sensitivity and reproducibility at trace levels support compliance testing and environmental monitoring.
  • A dual-column strategy ensures confirmatory analysis from a single injection, improving laboratory efficiency.
  • Advanced column chemistry reduces active site interactions, yielding reliable quantitation across diverse chlorinated analytes.

Future Trends and Opportunities


Advances in column inertness and detection technology will further lower detection limits for persistent organic pollutants. Integration with high-throughput autosamplers and automated data processing will accelerate routine environmental screening. Emerging tandem configurations and alternative detectors may expand the suite of target analytes, enabling multiplexed trace analysis in complex matrices.

Conclusion


The dual-column GC method utilizing an ultra inert column demonstrates robust, sensitive, and confirmatory analysis of chlorinated plaguicides at trace levels. Its high linearity, low detection limits, and simplified maintenance make it a valuable approach for environmental laboratories and regulatory compliance.

Reference


  1. European Parliament and Council Directive 2000/60/EC establishing a framework for Community water policy.
  2. Agilent Technologies. Agilent J&W Ultra Inert GC Columns: A New Tool to Battle Challenging Active Analytes. Publication 5989-8685EN, 2008.
  3. Agilent Technologies. G3181B Two-Way Splitter Kit Without Makeup Gas Installation and Operation Guide.
  4. Smith D, Lynam K. A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program Pesticides Analysis. Publication 5990-4069EN, 2009.
  5. Smith D, Lynam K. Chlorinated Solvents and Disinfection By-Product Analysis Using HP-1ms Ultra Inert and DB-1301 Columns. Publication 5990-3737EN, 2009.
  6. Agilent Technologies. Understanding Your ChemStation. Manual G2070-91125, 2008.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program (CLP) Pesticides Analysis
A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program (CLP) Pesticides Analysis Application Note Environmental Authors Abstract Doris Smith and Ken Lynam Trace level organochlorine pesticide testing is a high volume critical analysis in the Agilent Technologies, Inc.…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorclp, clpdecachlorobiphenyl, decachlorobiphenylmethoxychlor, methoxychlorxlb, xlbaldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonetetrachloro
Agilent J&W GC Columns For Environmental Applications
Agilent J&W GC Columns For Environmental Applications
2012|Agilent Technologies|Brochures and specifications
Agilent J&W GC Columns For Environmental Applications MEASURE TRACE-LEVEL CONTAMINANTS AND MEET REGULATORY REQUIREMENTS WITH CONFIDENCE Reliably and efficiently analyze increasingly small quantities of active solutes Every day, you are on the front lines in the battle to safeguard our…
Key words
identification, identificationbenzo, benzobhc, bhcpeak, peakendrin, endrinagilent, agilentpesticides, pesticidescolumns, columnsendosulfan, endosulfanmin, minheptachlor, heptachlorxylene, xyleneinert, inertcolumn, columnultra
Rapid Contract Laboratory Program (CLP) Pesticides Analysis with 0.32 mm ID Capillary GC Columns Utilizing Hydrogen Carrier Gas
Rapid Contract Laboratory Program (CLP) Pesticides Analysis with 0.32 mm ID Capillary GC Columns Utilizing Hydrogen Carrier Gas Application Brief Doris Smith and Kenneth Lynam Environmental Analysis time is often a limiting factor in sample turnaround for a laboratory. In…
Key words
bhc, bhcendosulfan, endosulfanendrin, endrinclp, clpchlordane, chlordaneheptachlor, heptachlortelodrin, telodrinpesticides, pesticideskepone, keponetetrachloro, tetrachlorosplitter, splittersurrogate, surrogatedecachlorobiphenyl, decachlorobiphenylmirex, mirexmethoxychlor
Chlorinated Pesticides Analysis by Capillary GC: Alternative Approaches
Bulletin 841C Chlorinated Pesticides Analysis by Capillary GC: Alternative Approaches We manufacture and test SPB-608 capillary columns specifically for analyses of chlorinated pesticides. 30m x 0.25mm ID columns can be used in capillary GC systems; 15m x 0.53mm ID columns…
Key words
bhc, bhcendosulfan, endosulfanendrin, endrinheptachlor, heptachlorpesticides, pesticidesepoxide, epoxidedieldrin, dieldrinaldrin, aldrinaldehyde, aldehydesupelco, supelcosulfate, sulfateextraction, extractionchlordane, chlordaneenvi, envimethoxychlor
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike