Rapid Contract Laboratory Program (CLP) Pesticides Analysis with 0.32 mm ID Capillary GC Columns Utilizing Hydrogen Carrier Gas
Applications | 2009 | Agilent TechnologiesInstrumentation
Rapid and reliable determination of pesticide residues is essential for environmental monitoring, food safety, and regulatory compliance. Contract laboratories face pressure to maximize sample throughput without compromising data quality. By optimizing chromatographic speed and resolution, laboratories can reduce turnaround times and handle large sample volumes more efficiently.
This study aims to demonstrate a high-throughput gas chromatography method for the U.S. EPA Contract Laboratory Program (CLP) pesticide panel using 30 m × 0.32 mm internal diameter capillary columns. The approach leverages hydrogen as carrier gas and flow programming to shorten analysis times while preserving the medium-bore column’s loading capacity and chromatographic performance. Dual-column operation enables simultaneous primary and confirmatory analyses in a single injection.
The experimental setup comprised an Agilent 7890A GC system equipped with a micro-electron capture detector (µECD), an Agilent 7683B autosampler, and Agilent capillary flow technology (CFT) two-way splitter (p/n G3181B) without makeup gas. Primary separation utilized a DB-17ms column (30 m × 0.32 mm, 0.25 µm), while confirmation employed a DB-XLB column (30 m × 0.32 mm, 0.5 µm), both with 1 m deactivated retention gaps. Hydrogen carrier gas was flow-programmed via electronic pneumatic control to accelerate elution of late-eluting analytes. Sample injection was splitless (0.5 µL) with pulsed pressure and optimized inlet conditions. Standards of 27 organochlorine pesticides and two surrogates were prepared in 2,2,4-trimethylpentane at 5 ng/mL.
The dual-column method achieved baseline resolution of all 27 target pesticides plus surrogates in under eight minutes—more than ten minutes faster than conventional helium-based methods on 0.32 mm columns. Chromatograms showed sharp, symmetric peaks and consistent retention for low-level (1.25 pg on-column) analytes. Hydrogen carrier gas reduced retention times without compromising resolution. Flow programming further improved peak elution for higher-boiling compounds.
Further improvements may include integration of advanced detectors (e.g., tandem MS), miniaturized column formats, and automated method translation software for rapid carrier gas switching. Machine learning algorithms could optimize temperature and flow programs in real time. The approach can be extended to other environmental, food, and forensic pesticide panels.
This application demonstrates a rapid, high-throughput GC-µECD method for CLP pesticide analysis using 0.32 mm ID columns with hydrogen carrier gas and flow programming. The protocol provides complete primary and confirmatory data in less than eight minutes, significantly increasing laboratory productivity while retaining analytical rigor.
1. K. Lynam, J. Henderson Jr. A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Agilent publication 5989-8031EN, 2008.
2. K. Lynam, W. Long. Contract Laboratory (CLP) Pesticide Analysis with 0.18 mm ID High Efficiency GC Columns Utilizing Helium Carrier Gas. Agilent publication 5989-7818EN, 2008.
3. D. Smith, K. Lynam. A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program (CLP) Pesticides Analysis. Agilent Technologies publication 5990-4069EN, 2009.
4. Agilent GC Method Translation Software. Agilent Technologies, 2009.
5. Agilent G3181B Two-Way Splitter Kit Without Makeup Gas Installation and Operation Guide. Agilent Technologies, 2009.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Rapid and reliable determination of pesticide residues is essential for environmental monitoring, food safety, and regulatory compliance. Contract laboratories face pressure to maximize sample throughput without compromising data quality. By optimizing chromatographic speed and resolution, laboratories can reduce turnaround times and handle large sample volumes more efficiently.
Study Objectives and Overview
This study aims to demonstrate a high-throughput gas chromatography method for the U.S. EPA Contract Laboratory Program (CLP) pesticide panel using 30 m × 0.32 mm internal diameter capillary columns. The approach leverages hydrogen as carrier gas and flow programming to shorten analysis times while preserving the medium-bore column’s loading capacity and chromatographic performance. Dual-column operation enables simultaneous primary and confirmatory analyses in a single injection.
Methodology and Instrumentation
The experimental setup comprised an Agilent 7890A GC system equipped with a micro-electron capture detector (µECD), an Agilent 7683B autosampler, and Agilent capillary flow technology (CFT) two-way splitter (p/n G3181B) without makeup gas. Primary separation utilized a DB-17ms column (30 m × 0.32 mm, 0.25 µm), while confirmation employed a DB-XLB column (30 m × 0.32 mm, 0.5 µm), both with 1 m deactivated retention gaps. Hydrogen carrier gas was flow-programmed via electronic pneumatic control to accelerate elution of late-eluting analytes. Sample injection was splitless (0.5 µL) with pulsed pressure and optimized inlet conditions. Standards of 27 organochlorine pesticides and two surrogates were prepared in 2,2,4-trimethylpentane at 5 ng/mL.
Key Results and Discussion
The dual-column method achieved baseline resolution of all 27 target pesticides plus surrogates in under eight minutes—more than ten minutes faster than conventional helium-based methods on 0.32 mm columns. Chromatograms showed sharp, symmetric peaks and consistent retention for low-level (1.25 pg on-column) analytes. Hydrogen carrier gas reduced retention times without compromising resolution. Flow programming further improved peak elution for higher-boiling compounds.
Benefits and Practical Applications
- High sample throughput: complete CLP panel in <8 min.
- Maintained sample loading capacity of medium-bore columns.
- Robust primary and confirmatory data from a single injection.
- Reduced helium dependency, lowering operating costs.
- Simplified maintenance via CFT splitter without makeup gas.
Future Trends and Potential Applications
Further improvements may include integration of advanced detectors (e.g., tandem MS), miniaturized column formats, and automated method translation software for rapid carrier gas switching. Machine learning algorithms could optimize temperature and flow programs in real time. The approach can be extended to other environmental, food, and forensic pesticide panels.
Conclusion
This application demonstrates a rapid, high-throughput GC-µECD method for CLP pesticide analysis using 0.32 mm ID columns with hydrogen carrier gas and flow programming. The protocol provides complete primary and confirmatory data in less than eight minutes, significantly increasing laboratory productivity while retaining analytical rigor.
Reference
1. K. Lynam, J. Henderson Jr. A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Agilent publication 5989-8031EN, 2008.
2. K. Lynam, W. Long. Contract Laboratory (CLP) Pesticide Analysis with 0.18 mm ID High Efficiency GC Columns Utilizing Helium Carrier Gas. Agilent publication 5989-7818EN, 2008.
3. D. Smith, K. Lynam. A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program (CLP) Pesticides Analysis. Agilent Technologies publication 5990-4069EN, 2009.
4. Agilent GC Method Translation Software. Agilent Technologies, 2009.
5. Agilent G3181B Two-Way Splitter Kit Without Makeup Gas Installation and Operation Guide. Agilent Technologies, 2009.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program (CLP) Pesticides Analysis
2010|Agilent Technologies|Applications
A 0.32 mm ID Capillary Column Approach to Contract Laboratory Program (CLP) Pesticides Analysis Application Note Environmental Authors Abstract Doris Smith and Ken Lynam Trace level organochlorine pesticide testing is a high volume critical analysis in the Agilent Technologies, Inc.…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorclp, clpdecachlorobiphenyl, decachlorobiphenylmethoxychlor, methoxychlorxlb, xlbaldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonetetrachloro
GC/μECD Analysis of Chlorinated Plaguicides Using Agilent J&W HP-1ms Ultra Inert and Agilent J&W DB-1301 Capillary GC Columns
2009|Agilent Technologies|Applications
GC/µECD Analysis of Chlorinated Plaguicides Using Agilent J&W HP-1ms Ultra Inert and Agilent J&W DB-1301 Capillary GC Columns Application Note Environmental Authors Abstract Doris Smith and Ken Lynam This application demonstrates a dual column GC method with microelectron capture Agilent…
Key words
bhc, bhcplaguicides, plaguicidesendosulfan, endosulfanendrin, endrinchlordane, chlordaneheptachlor, heptachlormethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonetetrachloro, tetrachlorosulfate, sulfatesurrogate
GC-μECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil
2012|Agilent Technologies|Applications
GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil Application Note Food Safety and Environmental Markets Authors Abstract Jennifer McCulley and Ken Lynam An olive oil sample obtained from a local grocery store is analyzed for 20…
Key words
bhc, bhcendosulfan, endosulfanendrin, endrinolive, oliveoil, oilchlordane, chlordaneheptachlor, heptachlorsurrogate, surrogatetetrachloro, tetrachloroµecd, µecdsulfate, sulfatemethoxychlor, methoxychlorxylene, xylenexlb, xlbquechers
Evaluating CLP and EPA Methods for Pesticides in Water Using Agilent J&W DB-CLP1/DB-CLP2 GC Columns 
2012|Agilent Technologies|Applications
Evaluating CLP and EPA Methods for Pesticides in Water Using Agilent J&W DB-CLP1/DB-CLP2 GC Columns Application Note Environmental Authors Abstract Doris Smith and Ken Lynam The Agilent J&W DB-CLP1 and DB-CLP2 column pair has the selectivity to separate Agilent Technologies,…
Key words
bhc, bhcmin, minester, estersurrogate, surrogateendrin, endrinmethyl, methylendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlordecachlorobiphenyl, decachlorobiphenyltetrachloro, tetrachloropermethrin, permethrinpeak, peakepoxide, epoxidedieldrin