Fast Distilled Spirit Analysis with an Agilent 8850 Gas Chromatograph and Long-Term Stability Test

Applications | 2025 | Agilent TechnologiesInstrumentation
GC
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Rapid and precise quantification of alcohols, aldehydes, esters and organic acids in distilled spirits is essential for quality control, flavor profiling and process optimization. Traditional gas chromatographic methods often require more than 30 minutes per sample, limiting throughput and increasing energy consumption. Fast, reliable assays help producers respond to quality issues in real time, reduce laboratory bottlenecks and lower operational costs.

Objectives and Study Overview


The study aimed to develop and validate a high-throughput GC method using the Agilent 8850 system equipped with a flame ionization detector (FID) and a 10 m DB-WAX Ultra Inert column. Key goals included:
  • Reducing analysis cycle times by at least threefold versus conventional methods
  • Evaluating performance under nitrogen, helium and hydrogen carrier gases
  • Verifying system inertness and stability over more than 2 000 real-sample injections

Methodology and Instrumentation


Instrumentation and conditions:
  • Agilent 8850 GC/FID with rapid oven heating/cooling
  • 10 m×0.18 mm, 0.18 μm Agilent J&W DB-WAX Ultra Inert column
  • Agilent 7650A automatic liquid sampler (5 μL syringe)
  • Carrier gases tested: nitrogen (0.5–20 mL/min ramped flow), helium (0.16–20 mL/min), hydrogen (0.2–10 mL/min)
  • Oven program: initial 35 °C hold, ramp to 70 °C then to 210–240 °C at up to 172 °C/min
  • Internal standards: tert-pentanol, n-pentyl acetate, 2-ethylbutyric acid

Sample preparation:
  • Seventeen target analytes prepared in ethanol/water (60:40) solvent
  • Calibration levels L1–L6 covering the expected concentration range
  • Quantification based on correlation coefficients ≥ 0.9992

Main Results and Discussion


Analysis time and stability:
  • Nitrogen carrier gas: 11 min total cycle vs. > 35 min traditional, > 3× throughput
  • Helium carrier gas: 9 min cycle
  • Hydrogen carrier gas: 7 min cycle, enabling ultra-fast screening
  • Retention time RSD ≤ 0.02%, peak area RSD ≤ 2.7% across all analytes
  • Limits of detection between 0.5 and 4.3 ppm
  • Carryover of acetic and hexanoic acids < 2.5% after high-level injection, < 10% after two blanks
  • Consistent chromatographic performance over 2 000 injections of aqueous liquor samples

Energy and cost savings:
  • 8850 GC consumes 0.077 kWh per run (nitrogen) vs. 0.221 kWh for conventional GC
  • Single 8850 system can process 32 500 samples/year vs. three conventional systems
  • Annual energy cost savings of > $560 (at $0.12/kWh) with one 8850 unit
  • Even greater throughput and savings with helium or hydrogen carrier gases

Practical Benefits of the Method


By combining rapid oven cycling and a short, highly inert column, the Agilent 8850 GC method offers:
  • Substantial increases in sample throughput for QC labs
  • Reliable quantification with low detection limits and high repeatability
  • Reduced energy consumption and operating costs
  • Minimized downtime through robust long-term stability

Future Trends and Potential Applications


Emerging directions include:
  • Integration of automated data processing and AI-guided method optimization
  • Expanded analyte panels to include volatile sulfur compounds and phenolics
  • Miniaturized or portable GC platforms for in-line process monitoring
  • Alternative carrier gases or gas mixtures to further accelerate separations

Conclusion


The Agilent 8850 GC/FID combined with a 10 m DB-WAX UI column enables fast, energy-efficient analysis of 17 key compounds in distilled spirits. The method delivers three- to five-fold reductions in cycle time, excellent chromatographic performance and demonstrated stability over thousands of injections. Laboratories can substantially improve throughput, cut costs and maintain rigorous quality standards using this approach.

Reference


  1. Kenneth, L.; Zhou, Y. Analysis of Distilled Spirits Using an Agilent J&W DB-WAX Ultra Inert Capillary GC Column. Agilent Technologies Application Note 5991-6638EN, 2016.
  2. Denoyer, E. Improving Return on Innovation in Gas Chromatography. Agilent Technologies Application Note 5991-7833EN, 2017.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Flavor Compounds in Beer using the Integrated Agilent 8697 Headspace Sampler with the Agilent 8890 GC System
Application Note Food Testing and Agriculture Analysis of Flavor Compounds in Beer using the Integrated Agilent 8697 Headspace Sampler with the Agilent 8890 GC System Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd., Shanghai 200131 P. R. China Abstract This…
Key words
butanedione, butanedionebeer, beerethyl, ethylecd, ecdpentanedione, pentanedioneflavor, flavorcompounds, compoundsacetate, acetateisoamylol, isoamylolformate, formateisobutyraldehyde, isobutyraldehydestock, stockconcentration, concentrationcaprylate, caprylateused
Analysis of Some Alcohols, Aldehydes, and Esters in Distilled Spirits with the Agilent 8860 Gas Chromatograph
Application Note Spirits Analysis Analysis of Some Alcohols, Aldehydes, and Esters in Distilled Spirits with the Agilent 8860 Gas Chromatograph Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd. Shanghai 200131 P. R. China Abstract The Agilent 8860 GC configured with…
Key words
liquor, liquorspirits, spiritschinese, chineseflavor, flavorethyl, ethylratio, ratiowere, wereisovaleraldehyde, isovaleraldehydeacetal, acetalhexanoate, hexanoateethanol, ethanolstandard, standarddistilled, distilledfid, fidacetate
GC/MS detection of short chain fatty acids from mammalian feces using automated sample preparation in aqueous solution
Application Note Metabolomics GC/MS detection of short chain fatty acids from mammalian feces using automated sample preparation in aqueous solution Authors Takeshi Furuhashi and Genki Ishihara Anicom Specialty Medical Institute Inc., 8-17-1, Nishi Shinjuku, Shinjuku-ku, Tokyo, Kuniyo Sugitate, Agilent Technologies…
Key words
acid, acidderivatization, derivatizationchloroformate, chloroformatefeces, fecesscfa, scfacounts, countsfatty, fattyisobutanoic, isobutanoicchain, chainisobutanol, isobutanolshort, shortscfas, scfasisocaproic, isocaproicprotocol, protocolsamples
Method Translation and Evaluation to Implement Nitrogen Carrier Gas in the Dual-Flame Ionization Detector Configuration for Blood Alcohol Analysis
Application Note Forensics Method Translation and Evaluation to Implement Nitrogen Carrier Gas in the Dual-Flame Ionization Detector Configuration for Blood Alcohol Analysis Using the Agilent 8697 Headspace Sampler/8890 GC System Author Abstract Abbey Fausett Agilent Technologies, Inc. A dual-column headspace…
Key words
nitrogen, nitrogenhelium, heliumcarrier, carriergas, gasrrf, rrfitr, itrsplit, splitunder, underresolution, resolutionloop, loopboth, bothile, ileethanol, ethanolflame, flametranslation
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike