GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Blood Alcohol Determination Using Static Headspace Analysis with Optimized Sample Throughput

Applications |  | EST AnalyticalInstrumentation
HeadSpace, GC, GC columns, Consumables
Industries
Forensics
Manufacturer
EST Analytical, Restek, Shimadzu

Summary

Importance of the Topic


Accurate determination of blood alcohol concentration is critical in forensic toxicology and legal proceedings. Static headspace sampling coupled with gas chromatography–flame ionization detection (GC–FID) offers a reproducible and robust approach for quantifying ethanol and related volatile compounds in complex biological matrices, ensuring results withstand judicial scrutiny while meeting laboratory throughput demands.

Objectives and Overview of the Study


The primary goals were to optimize sample throughput without sacrificing accuracy or precision, to evaluate both primary and confirmation GC columns for blood alcohol analysis, and to implement software improvements on the autosampler to maintain a four-minute GC cycle time despite extended incubation periods.

Methodology and Instrumentation


  • Sample Preparation: Static headspace sampling of blood alcohol standards (0.01–0.40 g/dL) using n-propanol (0.20 g/dL) as an internal standard.
  • Autosampler: EST Analytical FLEX headspace autosampler with 2.5 mL syringe; incubation at 60 °C for 10 min with 80 % agitation; 4.1 min GC cycle time via optimized scheduling software.
  • Chromatography: Shimadzu GC-2010 Plus with split/splitless inlet (220 °C, 80:1 split, 39.1 cm/s helium); Restek Rtx-BAC Plus 1 and 2 columns (30 m × 0.32 mm × 1.8 µm); oven held at 40 °C for 4 min.
  • Detection: Flame ionization detector at 240 °C.

Main Results and Discussion


  • Linearity: All target volatiles exhibited R² ≥ 0.9999, with curve %RSD below 6 % across both columns.
  • Precision: Seven-replicate studies at 0.20 g/dL yielded %RSD ≤ 2.33 % for all analytes on both columns, demonstrating high reproducibility.
  • K-Factor Evaluation: Ethanol K-factor values remained within ±1.5 % acceptance criteria for both primary and confirmation columns.
  • Accuracy: Ethanol recoveries averaged 102 % (primary column) and 104 % (confirmation column), confirming method trueness.
  • Throughput Optimization: Advanced FLEX autosampler software enabled a 10-minute incubation without extending the 4-minute GC cycle, effectively doubling sample capacity per run.

Benefits and Practical Applications


  • Forensic Reliability: Provides legally defensible results with high accuracy and precision.
  • Enhanced Throughput: Software-driven scheduling allows more samples per batch without hardware changes.
  • Operational Flexibility: Dual-column confirmation ensures analytical confidence in routine and high-volume casework.

Future Trends and Applications


  • Automation Expansion: Integration of larger autosampler trays or multi-needle systems for further throughput gains.
  • Broader Volatile Profiling: Extension to other forensic or clinical analytes, such as drugs of abuse or metabolic markers.
  • Advanced Chromatographic Media: Utilization of shorter, narrower bore columns or fast GC techniques to reduce cycle times.
  • Data Analytics Integration: Real-time quality control and trend monitoring through AI-driven software for proactive instrument maintenance and method verification.

Conclusion


Static headspace GC–FID, combined with intelligent autosampler scheduling, delivers a high-throughput, precise, and accurate method for blood alcohol analysis that meets forensic standards and enhances laboratory efficiency.

Reference


  • Anne Jurek. Blood Alcohol Determination Using Static Headspace Analysis with Optimized Sample Throughput. EST Analytical Application Note.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Blood Alcohol Determination Using Static Headspace Analysis with Optimized Sample Throughput
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Blood Alcohol Determination Using Static Headspace Analysis with Optimized Sample Throughput Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH There are several ways to…
Key words
pass, passjurek, jurekanne, annebeer, beerethanol, ethanolflavor, flavorpropanol, propanolacetaldehyde, acetaldehydebutanol, butanolisopropanol, isopropanolsyringe, syringemicro, microneedle, needleagitation, agitationacetone
Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput
         JSB is an authorised partner of   Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput #129 Application Note - Environmental Author Abstract Anne Jurek Applications Chemist EST…
Key words
ethanol, ethanolpropanol, propanolacetaldehyde, acetaldehydebutanol, butanolsyringe, syringeisopropanol, isopropanoljsb, jsbneedle, needleagitation, agitationacetone, acetonealcohol, alcoholwait, waitheadspace, headspacedepth, depthdelay
Analysis of Yogurt by Dynamic Headspace and SPME
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Analysis of Yogurt by Dynamic Headspace and SPME Application Note Food and Flavor Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH This application note will compare dynamic…
Key words
count, countheadspace, headspaceyogurt, yogurtagitate, agitatefiber, fiberjurek, jurekpurge, purgeanne, annearea, areadynamic, dynamicspme, spmepropanoic, propanoicdiethylphthalate, diethylphthalateacetone, acetonedesorb
Determination of Benzene in Sunblock Employing Static Headspace Sampling
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Determination of Benzene in Sunblock Employing Static Headspace Sampling Application Note Consumer Products Authors Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Benzene is classified as a Group…
Key words
jurek, jurekanne, annebrand, brandincubate, incubatelotion, lotionincubator, incubatoragitate, agitatebeer, beerwait, waitsyringe, syringeflavor, flavorheadspace, headspacemicro, microinjection, injectionbenzene
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike