Static Headspace Sampling of Volatile Organic Compounds

Applications |  | EST AnalyticalInstrumentation
GC/MSD, GC/SQ, HeadSpace
Industries
Environmental
Manufacturer
EST Analytical, Agilent Technologies

Summary

Importance of the Topic


Volatile organic compounds (VOCs) are pervasive environmental contaminants affecting water quality and public health. Analytical methods that balance sensitivity, robustness and operational simplicity are essential for routine monitoring and regulatory compliance.

Objectives and Study Overview


This study evaluates static headspace sampling coupled with GC/MS using SIM/Scan acquisition as an alternative to traditional purge and trap methods for the analysis of over 50 VOCs according to EPA Method 8260. Method linearity, detection limits, precision and accuracy were assessed.

Methodology


Samples were equilibrated at 60 ºC for 20 minutes after addition of sodium chloride to improve analyte partitioning. A headspace aliquot of 1 mL was withdrawn and injected in splitless mode. Calibration covered 0.5–200 ppb with midpoint and low-level replicates to determine detection limits, precision and recovery.

Instrumentation


  • Autosampler: EST Analytical FLEX headspace system with 2.5 mL syringe
  • Gas Chromatograph: Agilent 7890 with Restek Rxi-624 Sil MS 30 m × 0.25 mm ID, 1.4 µm film column
  • Mass Spectrometer: Agilent 5975 operated in SIM/Scan mode (m/z 35–265)

Main Results and Discussion


Calibration curves exhibited RSDs below 15% in compliance with EPA 8260B. Method detection limits ranged from 0.10 to 0.67 ppb. Repeatability at 50 ppb showed average precision of 5.6% RSD and mean recoveries of 101%. Static headspace sampling provided consistent analyte response without active-site artifacts or trap degradation.

Benefits and Practical Applications


  • Streamlined sample preparation without purge/trap hardware
  • Wide linear dynamic range and reliable calibration
  • Elimination of foaming and trap maintenance issues
  • Seamless integration with automated workflows in environmental laboratories

Future Trends and Potential Applications


Advances in MS sensitivity and ion acquisition strategies (SIM/Scan) will further lower detection limits for headspace analysis. Coupling headspace with solid phase microextraction (SPME) and expanded automation promise broader applications in environmental and food matrices.

Conclusion


Static headspace sampling combined with SIM/Scan GC/MS meets or exceeds EPA 8260 requirements for a wide range of VOCs, offering a robust and efficient alternative to purge and trap for water quality analysis.

Reference


  • United States Environmental Protection Agency Method 8260B, Revision 2, December 1996

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Optimizing Volatile Organic Compound Determination by Static Headspace Sampling
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimizing Volatile Organic Compound Determination by Static Headspace Sampling Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Volatile contaminants in drinking, ground and wastewaters…
Key words
jurek, jurekanne, anneheadspace, headspacevocs, vocssyringe, syringesampling, samplingdelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzenebeer, beerdepth, depthflavor, flavorstatic, staticinjection
Optimizing Volatile Organic Compound Determination by Static Headspace Sampling
         JSB is an authorised partner of   Optimizing Volatile Organic Compound Determination by Static Headspace Sampling #133 Application Note - Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati,…
Key words
syringe, syringesampling, samplingheadspace, headspacevocs, vocsjsb, jsbdelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzenedepth, depthstatic, staticjurek, jurekfill, fillanne, anneinjection
Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling
         JSB is an authorised partner of   Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling #134 Anne Jurek Introduction: Static headspace sampling has always been a viable option…
Key words
syringe, syringedelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzeneheadspace, headspacedepth, depthjsb, jsbfill, fillinjection, injectionspeed, speedoff, offsampling, samplingvocs, vocsflex
A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Environmental laboratories are…
Key words
jurek, jurekanne, annecenturion, centuriondesorb, desorbacetate, acetateest, estsampling, samplingpurge, purgebake, bakebeer, beertert, tertflavor, flavorbutylbenzene, butylbenzeneextraction, extractionmicro
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike