Optimizing Volatile Organic Compound Determination by Static Headspace Sampling

Applications |  | ZOEX/JSBInstrumentation
GC/MSD, HeadSpace, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies, EST Analytical

Summary

Significance of the Topic


Volatile organic compounds (VOCs) in drinking, ground and wastewaters are widespread environmental pollutants with potential impacts on human health and ecosystems. Reliable detection of VOCs at trace levels is critical for regulatory compliance, water quality monitoring and risk assessment.

Objectives and Study Overview


This study evaluates an automated static headspace sampling approach coupled to gas chromatography–mass spectrometry (GC–MS) for quantifying over 50 priority VOCs in water. It aims to optimize sampling and analysis parameters, establish detection limits and precision, and benchmark performance against USEPA Method 8260 requirements.

Methodology


• Sample preparation: 10 mL water standard spiked with VOCs and 2 g sodium chloride in a headspace vial.
• Headspace conditions: incubate at 60 °C for 20 min with agitation (80% speed), equilibrate, and withdraw 1 000 µL headspace for injection.
• GC oven program: hold at 45 °C for 2 min, ramp 15 °C/min to 220 °C, hold 1.33 min (total run 15 min).
• MS acquisition: SIM/Scan mode covering m/z 35–265, segmented ion lists for target compounds.
• Calibration: linear range 0.5–200 ppb; replicate injections at low and mid‐level standards to assess method detection limits (MDLs), precision and recovery.

Instrumentation


  • Autosampler: EST Analytical FLEX headspace sampler with 2.5 mL syringe.
  • Gas chromatograph: Agilent 7890 with split/splitless inlet and Restek Rxi-624 Sil MS column (30 m × 0.25 mm × 1.4 µm).
  • Mass spectrometer: Agilent 5975, source at 230 °C, transfer line 180 °C.

Main Results and Discussion


All 50+ VOCs produced linear calibration curves with curve RSD averages of 8.1% (well below the 15% requirement). MDLs ranged typically below 1 ppb for most compounds. Precision at mid‐level standards averaged <6% RSD and overall recoveries averaged 101.4%. Performance met or exceeded USEPA 8260 criteria across the analyte panel.

Benefits and Practical Applications


  • Eliminates need for purge‐and‐trap hardware and potential trap artefacts or foaming issues.
  • Simplifies sample preparation with automated vial handling directly on the GC.
  • Extends linear range and maintains competitive sensitivity.
  • Suitable for routine environmental monitoring under European and Canadian regulations, and adaptable to other regions.

Future Trends and Opportunities


Advances may include integration of high‐resolution mass spectrometry, miniaturized headspace systems for field deployment, online real‐time monitoring for process control, and expansion to semi‐volatile or polar organic compounds using modified headspace or extraction techniques.

Conclusion


Automated static headspace sampling coupled with SIM/Scan GC–MS provides a robust, accurate and efficient alternative to traditional purge‐and‐trap methods for VOC analysis in water. It delivers reliable detection limits, precision and recoveries in compliance with USEPA 8260, while offering operational simplicity and broad applicability.

References


1. Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry; United States Environmental Protection Agency Method 8260B, Revision 2, December 1996.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling
         JSB is an authorised partner of   Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling #134 Anne Jurek Introduction: Static headspace sampling has always been a viable option…
Key words
syringe, syringedelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzeneheadspace, headspacedepth, depthjsb, jsbfill, fillinjection, injectionspeed, speedoff, offsampling, samplingvocs, vocsflex
Optimizing Volatile Organic Compound Determination by Static Headspace Sampling
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimizing Volatile Organic Compound Determination by Static Headspace Sampling Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Volatile contaminants in drinking, ground and wastewaters…
Key words
jurek, jurekanne, anneheadspace, headspacevocs, vocssyringe, syringesampling, samplingdelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzenebeer, beerdepth, depthflavor, flavorstatic, staticinjection
Static Headspace Sampling of Volatile Organic Compounds
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Static Headspace Sampling of Volatile Organic Compounds Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Volatile contaminants in drinking, ground and wastewaters are an…
Key words
jurek, jureksyringe, syringeanne, annedelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzenedepth, depthfill, fillinjection, injectionspeed, speedheadspace, headspacebeer, beeroff, offflavor
A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
         JSB is an authorised partner of   #101 A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek Introducttion: The Unite ed States Environmental Protection…
Key words
centurion, centurionacetate, acetatedesorb, desorbest, estpurge, purgesampling, samplingtert, tertpling, plingtime, timebutylbenzene, butylbenzenebake, bakesamp, sampencon, enconwater, watermethyl
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike