Improved Retention Time, Area Repeatability, and Sensitivity for Analysis of Residual Solvents
Applications | 2007 | Agilent TechnologiesInstrumentation
Residual organic solvents in pharmaceuticals present significant health hazards, prompting regulatory agencies such as the FDA, USP, and ICH to enforce strict limits. Accurate, sensitive, and repeatable GC methods are therefore essential for compliance and patient safety.
This study evaluates enhancements in static headspace GC analysis of residual solvents using the Agilent 7890A GC coupled with a G1888 headspace sampler. Key targets include improved retention time stability, peak area repeatability, sensitivity, and reduced analysis time.
The system integrates an Agilent 7890A GC with an electronic backpressure regulator and dual mode PCM, and an Agilent G1888 headspace sampler with controlled vial pressure. A DB-624 column (30 m x 0.45 mm x 2.55 μm) and FID are used under an oven program from 35 °C to 250 °C, total cycle time of 50 min including backflush. Injector temperature is 160 °C, split ratio 2:1–4:1, helium carrier at 9 mL/min.
Advances may include coupling with mass spectrometry, dynamic pressure modulation, and automated data workflows. Emerging pharmaceutical and environmental requirements will drive further reductions in detection limits and cycle times, broadening the application of advanced pneumatic and capillary technologies.
Integrating state-of-the-art pneumatic electronics and capillary flow features in the 7890A GC significantly enhances static headspace analysis of residual solvents, delivering superior precision, sensitivity, and throughput to support stringent regulatory and quality control demands.
GC, HeadSpace
IndustriesPharma & Biopharma
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Residual organic solvents in pharmaceuticals present significant health hazards, prompting regulatory agencies such as the FDA, USP, and ICH to enforce strict limits. Accurate, sensitive, and repeatable GC methods are therefore essential for compliance and patient safety.
Objectives and Study Overview
This study evaluates enhancements in static headspace GC analysis of residual solvents using the Agilent 7890A GC coupled with a G1888 headspace sampler. Key targets include improved retention time stability, peak area repeatability, sensitivity, and reduced analysis time.
Methodology
The system integrates an Agilent 7890A GC with an electronic backpressure regulator and dual mode PCM, and an Agilent G1888 headspace sampler with controlled vial pressure. A DB-624 column (30 m x 0.45 mm x 2.55 μm) and FID are used under an oven program from 35 °C to 250 °C, total cycle time of 50 min including backflush. Injector temperature is 160 °C, split ratio 2:1–4:1, helium carrier at 9 mL/min.
Instrumentation
- Agilent 7890A GC with advanced pneumatic control
- Agilent G1888 static headspace sampler with 1 mL loop
- DB-624 capillary column
- Flame ionization detector
Main Results and Discussion
- Area repeatability improved to 3% RSD from 9% by applying 5 psig backpressure in the HS loop
- Retention time variation reduced to ±0.001 min
- Sensitivity doubled via loop pressurization compared to atmospheric sampling
- Analysis cycle time halved for Class 1 solvents thanks to column backflush and fast oven cooldown
- Method detection limit of 9.8 ppm for o-xylene at 99% confidence (n=8)
Benefits and Practical Applications
- Improved reliability for regulatory QC in pharmaceuticals
- Enhanced detection capabilities meet tightening solvent thresholds
- Faster throughput supports high-volume laboratories
- Adaptable to a wide concentration range across solvent classes
Future Trends and Applications
Advances may include coupling with mass spectrometry, dynamic pressure modulation, and automated data workflows. Emerging pharmaceutical and environmental requirements will drive further reductions in detection limits and cycle times, broadening the application of advanced pneumatic and capillary technologies.
Conclusion
Integrating state-of-the-art pneumatic electronics and capillary flow features in the 7890A GC significantly enhances static headspace analysis of residual solvents, delivering superior precision, sensitivity, and throughput to support stringent regulatory and quality control demands.
References
- US EPA Method 524.2, Revision 4, 1992
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467>
2014|Agilent Technologies|Technical notes
System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467> Technical Overview Introduction The Agilent 7697A Headspace Sampler introduced onboard electronic pneumatics control and several instrument control enhancements to improve performance over earlier models…
Key words
vial, vialloop, looppressure, pressurefill, fillpressurization, pressurizationprocedure, procedurecontrol, controlsettings, settingsflow, flowshaking, shakingmin, minresidual, residualamount, amountmode, modearea
USP<467> residual solvents
2018|Agilent Technologies|Applications
Application Note Pharmaceuticals USP<467> residual solvents Applying the Agilent 5977A MSD with the Agilent 7697A headspace sampler and Agilent 7890B GC Authors Roger L. Firor and Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808-1610 USA Abstract The…
Key words
vial, vialheadspace, headspacesolvents, solventsresidual, residualloop, loopacquisition, acquisitiontetralin, tetralintetrahyrofuran, tetrahyrofurannitromethane, nitromethanesim, simmethylcyclohexane, methylcyclohexanecounts, countspressure, pressureusp, uspxylene
The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler
2004|Agilent Technologies|Applications
The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application Pharmaceuticals Author Introduction Roger L. Firor Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Organic volatile impurities (OVIs) can result from the manufacture…
Key words
dmi, dmiheadspace, headspacechloroform, chloroformtrichloroethylene, trichloroethylenebenzene, benzenesolvent, solventarea, areasolvents, solventscarryover, carryoverxylene, xylenecoelution, coelutiontemperature, temperaturecyclohexane, cyclohexaneconcentration, concentrationoven
Agilent G1888 Network Headspace Sampler
2007|Agilent Technologies|Brochures and specifications
Agilent G1888 Network Headspace Sampler Exceptional sensitivity, repeatability and productivity. Our measure is your success. products | applications | software | services Turn potential measurement obstacles into opportunities for success with Agilent’s G1888 Network Headspace Sampler. The Agilent G1888 Headspace…
Key words
agilent, agilenthdpe, hdpeyour, youryou, youxylene, xylenemfs, mfssim, simheadspace, headspacescan, scantime, timepressure, pressuresynchronous, synchronoushss, hsssample, samplebenzaldehyde