Improved Performance And Dynamic Range For EPA Method TO-15 using the Entech 7200 and Shimadzu QP2010 Ultra GCMS

Applications |  | ENTECHInstrumentation
GC/MSD, Thermal desorption, Sample Preparation, GC/SQ
Industries
Environmental
Manufacturer
Shimadzu, ENTECH

Summary

Importance of the Topic


Accurate measurement of volatile organic compounds (VOCs) in ambient air is essential for environmental monitoring, regulatory compliance, and public health protection. Traditional canister-based TO-15 methods face challenges of limited dynamic range, high detection limits, and sample carryover. This study introduces advances in preconcentration and GC-MS analysis that extend sensitivity to the parts-per-trillion (ppt) level while maintaining a broad quantitation range.

Objectives and Study Overview


The primary goal was to demonstrate the performance of a new Air Toxics TO-15 Analyzer combining the Entech 7200 preconcentrator with a Shimadzu QP2010 Ultra GC-MS. Key objectives included:
  • Extending linear dynamic range to ≥100-fold with TO-15 calibration criteria
  • Lowering detection limits to 10–20 ppt in full-scan mode
  • Minimizing water, CO₂, and matrix interferences
  • Reducing carryover and sample preparation downtime

Methodology and Instrumentation


Samples collected in Summa canisters were processed through a three-stage Extended Cold Trap Dehydration protocol: an empty Silonite-D trap at ‑40 °C to freeze out water, a Tenax second stage at ‑40 °C for VOC capture, and a final focusing trap for GC injection. Volumes from 10 to 1 000 mL of a blended 10 ppbv standard were sampled under Electronic Volume Control. Separation was performed on a Shimadzu QP2010 Ultra GC-MS (35 °C hold, 6 °C/min to 120 °C, 10 °C/min to final) with full-scan acquisition (28–280 amu). Method Detection Limits (MDLs) were determined from seven replicates of a 2 ppbv standard in 100 mL aliquots.

Used Instrumentation


  • Entech 7200 Preconcentrator with Extended Cold Trap Dehydration
  • Silonite-D coated traps and transfer lines
  • Entech 7650 Robotic Autosampler for sample isolation
  • Shimadzu QP2010 Ultra GC-MS
  • Entech 4600A Dynamic Dilution System and 4700 Precision Diluter

Key Findings and Discussion


Linearity was maintained over a 100- to 2 000-fold range, meeting EPA TO-15 criteria with RSDs generally below 10–20%. Full-scan MDLs were 10–20 ppt, while single-ion monitoring achieved signal-to-noise ratios >20 at 0.2 ppbv. The inert Silonite-D surfaces prevented compound losses and reduced tailing. Sub-ambient trapping virtually eliminated carryover, enabling rapid recovery after high-concentration exposures.

Benefits and Practical Applications


  • Expanded dynamic range reduces dilution steps and enhances throughput
  • Ppt-level sensitivity addresses vapor intrusion and air toxics regulations
  • Robust performance in the presence of water, CO₂, and variable sample matrices
  • Zero cross-contamination through digital rotary valves and autosampler isolation

Future Trends and Opportunities


Further integration of inert materials and precision flow control promises even greater sensitivity and robustness. Future developments may include cooling systems that obviate liquid nitrogen, advanced automation for higher sample throughput, and real-time data analytics using AI. Extensions to complex matrices such as soil gas or indoor air mixtures will broaden the method’s applicability.

Conclusion


The Entech 7200 coupled with the Shimadzu QP2010 Ultra GC-MS significantly advances EPA TO-15 analysis by achieving ppt detection limits, a wide dynamic range, and improved reliability. Enhanced water management, inert flow paths, and precise volume control deliver high confidence in VOC quantitation, reducing laboratory downtime and meeting demanding environmental monitoring needs.

References


Application Note A-3736-01, Entech Instruments, Inc., Simi Valley, CA

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Improved Performance And Dynamic Range For EPA Method TO-15 using the Entech 7200 and Thermo Scientific ISQ QD GCMS
Improved Performance And Dynamic Range For EPA Method TO-15 using the Entech 7200 and Thermo Scientific ISQ QD GCMS INSTRUMENTS INC. Application Note: A-3736-02 Authors Doug Markle Chris Casteel Tom Robinson Dan Cardin Entech Instruments, Inc. Simi Valley, CA 93065…
Key words
canisters, canisterstrapping, trappingsilonite, siloniteair, aircanister, canistertrap, trapisq, isqpreconcentration, preconcentrationvolume, volumeconcentration, concentrationsample, sampleaccu, accuvery, veryentech, entechgcms
Improved Performance And Dynamic Range For EPA Method TO-15 using the Entech 7200 and Agilent 7890B/5977 GCMS
Improved Performance And Dynamic Range For EPA Method TO-15 using the Entech 7200 and Agilent 7890B/5977 GCMS Application Note: A-3737-01 Authors Tom Robinson Dan Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Abstract Canister sampling and analysis for measurement of…
Key words
trap, trapcanister, canistercanisters, canistersair, airtrapping, trappinglow, lowcompounds, compoundspreconcentration, preconcentrationvolume, volumewhich, whichnext, nextsilonite, siloniteinertness, inertnessfocusing, focusingwater
Fast US EPA TO15 Analysis Providing Higher Productivity using the new Entech 7200A and Agilent 7890B/5977 GCMS
Fast US EPA TO15 Analysis Providing Higher Productivity using the new Entech 7200A and Agilent 7890B/5977 GCMS Application Note: A-3742-01 Authors Daniel B. Cardin Tom Robinson Victoria L. Noad Entech Instruments Simi Valley, CA 93065 Introduction Canister sampling and analysis…
Key words
tert, tertether, etherchloride, chlorideentech, entechbutyl, butylvinyl, vinylmethyl, methylbromoethene, bromoethenetrichlorotrifluoroethane, trichlorotrifluoroethaneisopropyl, isopropylpreconcentrators, preconcentratorsethyl, ethylbenzene, benzeneacetate, acetatepropene
Expanding the Molecular Weight Range of Whole Air Sampling with Stainless Steel Canisters using Active SPME Sample Preparation and GCMS Analysis
Expanding the Molecular Weight Range of Whole Air Sampling with Stainless Steel Canisters using Active SPME Sample Preparation and GCMS Analysis Application Note: A-3732-03 Authors Chris Casteel Dan Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Abstract A new preconcentration…
Key words
spme, spmecanisters, canisterscanister, canisterthermal, thermaltrap, trapdesorb, desorbbakeout, bakeoutrotary, rotarycompounds, compoundsactive, activetenax, tenaxair, airdean, deaneliminated, eliminatedthermally
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike