Expanding the Molecular Weight Range of Whole Air Sampling with Stainless Steel Canisters using Active SPME Sample Preparation and GCMS Analysis
Applications | 2010 | ENTECHInstrumentation
Whole-air canister sampling combined with thermal preconcentration and GC–MS detection is widely used for trace analysis of volatile and semi-volatile organic compounds at sub-ppb levels. However, conventional methods struggle to recover high molecular-weight semi-volatiles (beyond C18–C20) and thermally labile analytes. The development of an Active SPME-based three-stage preconcentrator addresses these limitations, extending the method’s applicability to C24 compounds and improving recovery of sensitive species.
This application note by Chris Casteel and Dan Cardin (Entech Instruments) demonstrates the integration of an Active SPME stage into a three-trap preconcentrator (Entech 7150) to expand the molecular weight range of whole-air sampling. The study compares recoveries of a C10–C24 hydrocarbon standard and 65 EPA Method TO-15 analytes, evaluates system linearity, carryover, and reproducibility, and benchmarks performance against direct GC injection.
The system achieved quantitative recovery of C2–C24 compounds. Direct injection of the C10–C24 standard into GC–MS showed less than 5% variance, while preconcentration through the 7150 delivered comparable or improved responses, likely due to reduced thermal degradation. The 65-component TO-15 calibration exhibited excellent linearity (R2 > 0.995) across 0.4–20 ppb. No carryover was detected in blanks following high-level hydrocarbon injections, demonstrating effective cleanup and trap protection. An observed C22 enrichment was traced to canister heating heterogeneity and is being addressed in oven redesign.
Emerging techniques such as helium diffusion sampling promise time-integrated collection without external flow controllers, further reducing sampling artifacts for heavy semi-volatiles. Ongoing improvements in oven uniformity will ensure consistent analyte enrichment. The Active SPME approach may be extended to polar, reactive, or biogenic VOCs, supporting applications in indoor air quality, hazardous chemical detection, and process monitoring.
The three-stage Active SPME preconcentrator effectively broadens the molecular weight range of canister-based air sampling to C24 while preserving quantitative performance for EPA TO-15 analytes. The design’s reduced dead volume, inert surfaces, and dynamic trapping stages deliver high sensitivity, reproducibility, and zero carryover, making it a versatile tool for trace organic analysis in diverse environmental and industrial settings.
None specified in the source document.
GC/MSD, SPME, Thermal desorption, Sample Preparation, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, ENTECH
Summary
Importance of the Topic
Whole-air canister sampling combined with thermal preconcentration and GC–MS detection is widely used for trace analysis of volatile and semi-volatile organic compounds at sub-ppb levels. However, conventional methods struggle to recover high molecular-weight semi-volatiles (beyond C18–C20) and thermally labile analytes. The development of an Active SPME-based three-stage preconcentrator addresses these limitations, extending the method’s applicability to C24 compounds and improving recovery of sensitive species.
Objectives and Study Overview
This application note by Chris Casteel and Dan Cardin (Entech Instruments) demonstrates the integration of an Active SPME stage into a three-trap preconcentrator (Entech 7150) to expand the molecular weight range of whole-air sampling. The study compares recoveries of a C10–C24 hydrocarbon standard and 65 EPA Method TO-15 analytes, evaluates system linearity, carryover, and reproducibility, and benchmarks performance against direct GC injection.
Methodology and Instrumentation
- Sampling media: 6 L and 1.4 L Silonite coated stainless steel canisters pre-cleaned to <50 mtorr
- Preconcentrator: Entech 7150 three-stage system with Dean’s Switching valves for reduced dead volume and elimination of rotary valve contamination
- Stage 1 Active SPME trap: traps C10–C25 analytes at 50 °C and refocuses at –52 °C
- Stage 2 dehydration trap: cold trap at –40 °C for vapor-to-solid water removal, then heated to 160 °C to release water without analyte loss
- Stage 3 Tenax trap: captures C2–C12 at –50 °C, desorbs at 200 °C for quantitative VOC recovery
- Carrier: helium delivered via Dean’s Switching, eliminating liquid nitrogen and reducing cost
- GC–MS: Agilent 7890 GC coupled to 5975 MSD with HP-1 column (60 m x 0.32 mm x 1 μm film), full-scan mode
- Calibration: 5-point TO-15 standard (0.4–20 ppb) and C10–C24 hydrocarbon mix (50 ng/μL in methanol)
Major Results and Discussion
The system achieved quantitative recovery of C2–C24 compounds. Direct injection of the C10–C24 standard into GC–MS showed less than 5% variance, while preconcentration through the 7150 delivered comparable or improved responses, likely due to reduced thermal degradation. The 65-component TO-15 calibration exhibited excellent linearity (R2 > 0.995) across 0.4–20 ppb. No carryover was detected in blanks following high-level hydrocarbon injections, demonstrating effective cleanup and trap protection. An observed C22 enrichment was traced to canister heating heterogeneity and is being addressed in oven redesign.
Benefits and Practical Applications
- Extended molecular weight coverage up to C24 without overheating adsorbents
- Improved recovery of thermally labile and polar compounds through low-temperature SPME trapping
- Sub-ppb sensitivity and minimal carryover enable reliable monitoring of environmental and industrial air pollutants
- Streamlined field sampling: vacuum-based canisters with no need for flow controllers
- Cost and maintenance savings via helium switching and elimination of liquid nitrogen
Future Trends and Potential Applications
Emerging techniques such as helium diffusion sampling promise time-integrated collection without external flow controllers, further reducing sampling artifacts for heavy semi-volatiles. Ongoing improvements in oven uniformity will ensure consistent analyte enrichment. The Active SPME approach may be extended to polar, reactive, or biogenic VOCs, supporting applications in indoor air quality, hazardous chemical detection, and process monitoring.
Conclusion
The three-stage Active SPME preconcentrator effectively broadens the molecular weight range of canister-based air sampling to C24 while preserving quantitative performance for EPA TO-15 analytes. The design’s reduced dead volume, inert surfaces, and dynamic trapping stages deliver high sensitivity, reproducibility, and zero carryover, making it a versatile tool for trace organic analysis in diverse environmental and industrial settings.
References
None specified in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Formaldehyde and VOC’s in Indoor Air Quality Determinations by GCMS
2007|Agilent Technologies|Applications
Formaldehyde and VOC’s in Indoor Air Quality Determinations by GCMS Application Note: 101 Application Note: A-3727-01 Authors Dan Cardin Tom Robinson Entech Instrurments, Inc Simi Valley, CA 93065 Abstract A stainless steel canister based method has been developed to measure…
Key words
formaldehyde, formaldehydeppbv, ppbvair, airpropanal, propanalcanister, canisterindoor, indoorcarbonyl, carbonylgcms, gcmsminicans, minicanssilonite, siloniteacetone, acetoneacetalde, acetaldeformalde, formaldesampling, samplingminican
Improving Canister Sampling Accuracy and Reliability Using Helium Diffusion Sampling into Silonite Coated Stainless Steel Canisters
|Agilent Technologies|Applications
Improving Canister Sampling Accuracy and Reliability Using Helium Diffusion Sampling into Silonite Coated Stainless Steel Canisters INSTRUMENTS INC. Application Note: A-3731-01 Authors Daniel B. Cardin Thomas X. Robinson Daniel J. Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Abstract A…
Key words
sampling, samplingcanister, canisterhds, hdsdiffusion, diffusioncanisters, canistersair, airhelium, heliumover, overgauge, gaugesampler, samplervacuum, vacuumsilonite, silonitevac, vacatmospheric, atmosphericorifice
5800 Sorbent Pen™ Desorption Unit
2020|ENTECH|Brochures and specifications
5800 Sorbent Pen™ Desorption Unit Compatibility • Compatible with Agilent®, Thermo Scientific®, and Shimadzu® GC/GCMS systems GC GC Oven Column 1 MS Temperature Programming • Two temperature desorption setpoints • Heating rates up to 450 °C/min • Initial temperature 50-350…
Key words
sorbent, sorbentdesorption, desorptionpens, pensentech, entechtenax, tenaxspr, sprthermal, thermalheadspace, headspacesvocs, svocsvase, vasetubes, tubescarbopack, carbopackair, airfeve, feverail
A Simple, More Reliable Time Integrated Whole Air Sampling Approach To EPA Method TO15 Using Helium Diffusion Sampling
|Agilent Technologies|Applications
A Simple, More Reliable Time Integrated Whole Air Sampling Approach To EPA Method TO15 Using Helium Diffusion Sampling INSTRUMENTS INC. Application Note: A-3732-01 Authors Thomas X. Robinson Chris Casteel Dan Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Abstract A…
Key words
sampling, samplingcanisters, canistersair, aircanister, canisteratmospheric, atmosphericpressure, pressurepropane, propanevacuum, vacuumdiffusion, diffusionsampled, sampledgauge, gaugehelium, heliumvirtually, virtuallyintegrated, integratedagain