Flavors, Odors, and Contaminants in Alcoholic Beverages using Vacuum Assisted Sorbent Extraction (VASE) and GC/MS Analysis
Applications | 2017 | ENTECHInstrumentation
The analysis of volatile and semi-volatile compounds in alcoholic beverages is essential for quality control, flavor profiling, and contaminant monitoring. Traditional headspace and solvent-based extraction methods face challenges such as low recoveries for heavy analytes, matrix interferences from high ethanol content, and extensive sample handling. Vacuum Assisted Sorbent Extraction (VASE) addresses these limitations by using a static vacuum to enhance diffusion and capture a broad range of analytes with improved sensitivity and reproducibility.
This application note demonstrates VASE coupled with GC/MS for the comprehensive analysis of flavor, odor, and trace contaminants in rum, whiskey, cognac, and wine. Key aims include:
Samples (20–50 µL of beverage) were placed in 20–40 mL vials with Tenax-packed Sorbent Pens. A MicroQT valve and vacuum pump (<0.01 atm) evacuated the headspace in 15–30 s, then the closed system sat under static vacuum at controlled temperature (4–70 °C) and agitation (30–300 rpm) for 15 min to 48 hr to reach equilibrium. After extraction, Sorbent Pens were thermally desorbed in an Entech 5800 Sorbent Pen Desorption Unit. A filmless pre-column and dual-column configuration managed water and ethanol, using a split injection to deliver analytes onto an Agilent 7890/5977 GC/MS. Calibration standards (C8–C30 hydrocarbons; isotopically labeled geosmin-d3 and target analytes) ensured system performance and quantitation from 0.1 to 20 ng/L.
VASE achieved up to 50-fold higher recovery of low-volatility compounds than SPME. Key findings include:
VASE provides a robust, solvent-free headspace extraction method with the following advantages:
This makes VASE suitable for routine flavor profiling, QA/QC of alcoholic beverages, and trace contaminant surveillance.
Further developments may include:
VASE coupled with GC/MS demonstrates sensitive, reproducible extraction of a wide volatility range from high-alcohol matrices. By leveraging static vacuum diffusion and packed sorbent traps, the method overcomes the limitations of conventional headspace techniques, enabling efficient flavor and contaminant analysis without extensive sample preparation or isotope dilution.
GC/MSD, HeadSpace, Thermal desorption, GC/SQ
IndustriesFood & Agriculture
ManufacturerAgilent Technologies, ENTECH
Summary
Significance of the Topic
The analysis of volatile and semi-volatile compounds in alcoholic beverages is essential for quality control, flavor profiling, and contaminant monitoring. Traditional headspace and solvent-based extraction methods face challenges such as low recoveries for heavy analytes, matrix interferences from high ethanol content, and extensive sample handling. Vacuum Assisted Sorbent Extraction (VASE) addresses these limitations by using a static vacuum to enhance diffusion and capture a broad range of analytes with improved sensitivity and reproducibility.
Study Objectives and Overview
This application note demonstrates VASE coupled with GC/MS for the comprehensive analysis of flavor, odor, and trace contaminants in rum, whiskey, cognac, and wine. Key aims include:
- Evaluating extraction efficiency across volatilities (C8–C30) under static vacuum conditions.
- Comparing neat, water-diluted, and salt-enhanced matrices to optimize recoveries.
- Establishing calibration and quantitation for regulated trace analytes (trichloroanisoles, tribromoanisole, geosmin, 2-MIB) at ng/L levels.
- Assessing reproducibility, carryover, and matrix effects in complex alcohol-rich samples.
Methodology and Instrumentation
Samples (20–50 µL of beverage) were placed in 20–40 mL vials with Tenax-packed Sorbent Pens. A MicroQT valve and vacuum pump (<0.01 atm) evacuated the headspace in 15–30 s, then the closed system sat under static vacuum at controlled temperature (4–70 °C) and agitation (30–300 rpm) for 15 min to 48 hr to reach equilibrium. After extraction, Sorbent Pens were thermally desorbed in an Entech 5800 Sorbent Pen Desorption Unit. A filmless pre-column and dual-column configuration managed water and ethanol, using a split injection to deliver analytes onto an Agilent 7890/5977 GC/MS. Calibration standards (C8–C30 hydrocarbons; isotopically labeled geosmin-d3 and target analytes) ensured system performance and quantitation from 0.1 to 20 ng/L.
Main Results and Discussion
VASE achieved up to 50-fold higher recovery of low-volatility compounds than SPME. Key findings include:
- Clean blanks with negligible carryover after re-desorption.
- High reproducibility (RSD ≤ 15%) across duplicate rum and whiskey extracts.
- Water dilution flattened ethanol peaks and lowered baseline, aiding detection of early-eluting analytes.
- Salt addition enhanced phthalate recoveries by salting-out effects.
- Calibration curves for trichloroanisole isomers, tribromoanisole, geosmin, and 2-MIB were linear over 0.1–20 ng/L with 60–90% recovery.
- Detection limits were more influenced by chromatographic separation than by extraction efficiency.
Benefits and Practical Applications
VASE provides a robust, solvent-free headspace extraction method with the following advantages:
- Broad analyte range from light volatiles to semi-volatiles without cryogenic focusing.
- Low carryover and high throughput through multi-pen, parallel extractions.
- Reduced matrix interferences in high-ethanol samples.
- Capability to quantify targets lacking isotope standards.
- Minimal sample disturbance, preserving thermally labile compounds.
This makes VASE suitable for routine flavor profiling, QA/QC of alcoholic beverages, and trace contaminant surveillance.
Future Trends and Potential Applications
Further developments may include:
- Expanding VASE to semi-volatile contaminants in foods and environmental waters (PAHs, endocrine disruptors, drugs of abuse).
- Integration with high-resolution MS for non-target screening.
- New sorbent chemistries to tune selectivity for polar and thermally sensitive analytes.
- Automation enhancements for higher throughput in regulatory and industrial testing labs.
Conclusion
VASE coupled with GC/MS demonstrates sensitive, reproducible extraction of a wide volatility range from high-alcohol matrices. By leveraging static vacuum diffusion and packed sorbent traps, the method overcomes the limitations of conventional headspace techniques, enabling efficient flavor and contaminant analysis without extensive sample preparation or isotope dilution.
Instrumentation
- Entech 5600 Sorbent Pen Extraction System (agitation, temperature control)
- MicroQT Vial Evacuation Tool with oil-free diaphragm pump
- Entech 5800 Sorbent Pen Desorption Unit
- Entech 3801 Sorbent Pen Thermal Conditioner
- Agilent 7890 GC coupled to 5977 MSD
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Sorbent Pens™ For Next Generation Headspace Analysis Featured Chromatograms Abundance 15 2e+08 1.8e+08 1.6e+08 1.4e+08 1.2e+08 1e+08 13 6 2 8e+07 6e+07 8 5 4e+07 4 11 14 3 2e+07 Time--> 7 1 0 10.00 12.00 14.00 16.00 9 10…
Key words
vase, vasesorbent, sorbentassisted, assistedabundance, abundancevacuum, vacuumextraction, extractionsplitless, splitlessacid, acidtime, timefresh, freshcanned, cannedanalysis, analysisethyl, ethylfurfural, furfuralacetoin
VASE - Vacuum Assisted Sorbent Extraction Odor, Pesticide, & PAH Analysis for Dairy Products
2017|Agilent Technologies|Applications
VASE - Vacuum Assisted Sorbent Extraction Odor, Pesticide, & PAH Analysis for Dairy Products Extending Quantitative Headspace to Include Less-Volatile Compounds in Difficult Matrices Application Note Brief: V-3741-03 Authors Victoria L. Noad Daniel B. Cardin Entech Instruments Simi Valley, CA…
Key words
vase, vaseabundance, abundancesorbent, sorbentheadspace, headspacepen, penacid, acidrerun, reruntime, timeentech, entechdairy, dairydesorb, desorbcompounds, compoundsvacuum, vacuummaltol, maltolonto
Quantitative Headspace Measurement of Volatiles in Dairy Products using Vacuum Assisted Sorbent Extraction (VASE) and GCMS Analysis
2017|Agilent Technologies|Applications
Quantitative Headspace Measurement of Volatiles in Dairy Products using Vacuum Assisted Sorbent Extraction (VASE) and GCMS Analysis Application Note: V-3741-01 Authors Victoria L. Noad Daniel B. Cardin Entech Instruments Simi Valley, CA 93065 Abstract The headspace of common dairy products,…
Key words
pen, pensorbent, sorbentabundance, abundanceheadspace, headspacepens, pensvacuum, vacuumdairy, dairycompounds, compoundsentech, entechacid, acidadsorbent, adsorbentspdu, spduvase, vasesample, samplecholesterol
Sorbent Pens™ (Pat. Pend.) For Next Generation Headspace Analysis
2020|ENTECH|Brochures and specifications
Sorbent Pens™(Pat. Pend.) For Next Generation Headspace Analysis SORBENT PENS™ FOR NEXT GENERATION Headspace Analysis Headspace Sorbent Pens™ The Ultimate Extraction Power Introducing Entech’s exciting new Sorbent Pen™ technology. The most versatile extraction and sample preconcentration technology available for GC…
Key words
sorbent, sorbentpen, penpens, penspack, packvase, vaserail, railheadspace, headspacemounting, mountinghsp, hspunit, unitqty, qtyvial, vialspr, sprextraction, extractiondescription