Simplified Analysis of Aqueous Short Chain Fatty Acids
Applications | 2022 | ShimadzuInstrumentation
Aqueous short-chain fatty acids (SCFAs) play essential roles in metabolic processes, microbial ecology, and functional foods. Accurate quantification of SCFAs in aqueous matrices supports research in nutrition, gut health, and food quality control. Streamlined methods that balance sensitivity, speed, and reproducibility are highly sought after in analytical laboratories.
This application note presents a simplified gas chromatography–mass spectrometry (GC-MS) approach for simultaneous analysis of ten SCFAs in water. The primary goals were to achieve baseline separation, reliable quantitation under selected ion monitoring (SIM), and a total run time below ten minutes to enhance laboratory throughput.
Chromatographic separation was performed on a 60 m × 0.25 mm ID, 0.5 µm SH-Wax capillary column. A constant linear velocity of 34 cm/s was maintained with helium as carrier gas. The oven ramp started at 80 °C (2 min), increased at 40 °C/min to 200 °C, then at 25 °C/min to a final temperature, completing the cycle in 8.6 min.
Injection was done in split mode (5:1) at 240 °C with a 1 µL injection volume. The Shimadzu GCMS-QP2020 NX system, coupled with the AOC-20i plus auto-injector and AOC-20s Plus autosampler, delivered stable sample introduction. Mass spectrometric detection in SIM mode targeted characteristic ions for each SCFA, optimizing signal-to-noise ratios.
All ten analytes—formic, acetic, propionic, isobutyric, butyric, isovaleric, valeric, 4-methylpentanoic, caproic, and heptanoic acids—were resolved with symmetrical peak shapes. Retention times ranged from approximately 5.8 min (acetic acid) to 8.2 min (caproic acid). Detection limits were in the low-µg/L range under SIM conditions, demonstrating sufficient sensitivity for most aqueous applications. The total ion chromatogram and individual extracted ion traces confirmed minimal interference and consistent quantitation.
Advances may include automated sample derivatization, multiplexed injection systems, and integration with high-resolution MS for broader metabolomic profiling. Miniaturized GC-MS platforms and online coupling with extraction techniques could further streamline SCFA monitoring in complex matrices.
The described GC-MS SIM method combines speed, sensitivity, and simplicity for comprehensive SCFA analysis in water. It supports quality control, nutritional studies, and microbiome research through reliable quantitation of key organic acids within an 8.6 min run time.
Consumables, GC columns, GC/MSD, GC/SQ
IndustriesManufacturerShimadzu
Summary
Importance of the Topic
Aqueous short-chain fatty acids (SCFAs) play essential roles in metabolic processes, microbial ecology, and functional foods. Accurate quantification of SCFAs in aqueous matrices supports research in nutrition, gut health, and food quality control. Streamlined methods that balance sensitivity, speed, and reproducibility are highly sought after in analytical laboratories.
Study Objectives and Overview
This application note presents a simplified gas chromatography–mass spectrometry (GC-MS) approach for simultaneous analysis of ten SCFAs in water. The primary goals were to achieve baseline separation, reliable quantitation under selected ion monitoring (SIM), and a total run time below ten minutes to enhance laboratory throughput.
Applied Methodology and Instrumentation
Chromatographic separation was performed on a 60 m × 0.25 mm ID, 0.5 µm SH-Wax capillary column. A constant linear velocity of 34 cm/s was maintained with helium as carrier gas. The oven ramp started at 80 °C (2 min), increased at 40 °C/min to 200 °C, then at 25 °C/min to a final temperature, completing the cycle in 8.6 min.
Injection was done in split mode (5:1) at 240 °C with a 1 µL injection volume. The Shimadzu GCMS-QP2020 NX system, coupled with the AOC-20i plus auto-injector and AOC-20s Plus autosampler, delivered stable sample introduction. Mass spectrometric detection in SIM mode targeted characteristic ions for each SCFA, optimizing signal-to-noise ratios.
Main Results and Discussion
All ten analytes—formic, acetic, propionic, isobutyric, butyric, isovaleric, valeric, 4-methylpentanoic, caproic, and heptanoic acids—were resolved with symmetrical peak shapes. Retention times ranged from approximately 5.8 min (acetic acid) to 8.2 min (caproic acid). Detection limits were in the low-µg/L range under SIM conditions, demonstrating sufficient sensitivity for most aqueous applications. The total ion chromatogram and individual extracted ion traces confirmed minimal interference and consistent quantitation.
Practical Benefits and Applications
- Rapid analysis enables high sample throughput in research and QA/QC laboratories.
- Simplified SIM method reduces data processing complexity while maintaining selectivity.
- The robust column and GC-MS configuration ensure reproducible retention and quantification.
- Applicable to diverse aqueous matrices, including functional foods, beverages, and fermentation broths.
Future Trends and Potential Applications
Advances may include automated sample derivatization, multiplexed injection systems, and integration with high-resolution MS for broader metabolomic profiling. Miniaturized GC-MS platforms and online coupling with extraction techniques could further streamline SCFA monitoring in complex matrices.
Conclusion
The described GC-MS SIM method combines speed, sensitivity, and simplicity for comprehensive SCFA analysis in water. It supports quality control, nutritional studies, and microbiome research through reliable quantitation of key organic acids within an 8.6 min run time.
References
- Application News 01-00027 (JP, ENG), Shimadzu Corporation, 2022.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Simplified Analysis of Aqueous Short Chain Fatty Acids by GC/MS
2021|Shimadzu|Applications
GC-MS GCMS-QP™2020 NX, GCMS-TQ™ NX series Simplified Analysis of Aqueous Short Chain Fatty Acids by GC/MS Application News Y. Nakagawa User Benefits No derivatization or special consumables required Total run time of about 10 mins including the cooling…
Key words
acid, acidpropionic, propionicisobutyric, isobutyricacetic, aceticfatty, fattymin, minhexanoic, hexanoicacids, acidsisovaleric, isovalericvaleric, valericderivatization, derivatizationbutyric, butyricchain, chainshort, shortbrim
Analysis of Neutral PFAS in Ambient Air Using Thermal Desorption GC-MS
2025|Shimadzu|Applications
GC-MS TD-30R/GCMS-QP2020 NX Application News Analysis of Neutral PFAS in Ambient Air Using Thermal Desorption GC-MS Nami Iwasa User Benefits A thermal desorption GC-MS system can accurately measure quantities of volatile and semivolatile per- and polyfluoroalkyl substances (PFAS) in…
Key words
etfosa, etfosamefosa, mefosaarea, areapfas, pfasconcentration, concentrationfosas, fosasftohs, ftohsfluorotelomer, fluorotelomersim, simfavorable, favorableambient, ambientsemivolatile, semivolatilenews, newsmodebased, modebasedair
Analysis of Phenethylamines Using On-Column TFA Derivatization
2024|Shimadzu|Applications
GC-MS GCMS-QP 2050 and GCMS NX series Application News Analysis of Phenethylamines Using On-Column TFA Derivatization Mahiro Yoshida,1 Yuki Sakamoto,1 and Maiko Kusano2 1 Shimadzu Corporation and 2 Department of Legal Medicine, School of Medicine, Showa University User Benefits …
Key words
tfa, tfamethamphetamine, methamphetaminephenethylamines, phenethylaminesderivatization, derivatizationamphetamine, amphetaminemin, minurine, urineretention, retentionnews, newslabsolutions, labsolutionsconcentration, concentrationanhydride, anhydridemdma, mdmamda, mdatime
Rapid quantification of 12 nitrosamines in metformin using triple quadrupole GC-MS/MS with Advanced Electron Ionization (AEI)
2021|Thermo Fisher Scientific|Applications
APPLICATION NOTE 000173 Rapid quantification of 12 nitrosamines in metformin using triple quadrupole GC-MS/MS with Advanced Electron Ionization (AEI) Authors: David Lee1, Aaron Lamb1, Andrea Romano2, and Giulia Riccardino2 Thermo Fisher Scientific, Runcorn, UK 2 Thermo Fisher Scientific, Milan, Italy…
Key words
nitroso, nitrosophenylamine, phenylaminenpyr, npyrnepha, nephandma, ndmanmpa, nmpahigh, highndpa, ndpanpip, npipndpha, ndphandea, ndeanmea, nmeandba, ndbandipa, ndipanmor