Importance of Gas-Phase Formaldehyde Analysis


Formaldehyde is a ubiquitous indoor and industrial pollutant with known health risks, including respiratory irritation and potential carcinogenicity. Accurate monitoring of airborne formaldehyde concentrations is essential for environmental safety, occupational health, and regulatory compliance. The DNPH derivatization followed by GC-FID analysis provides a robust and sensitive approach to quantify trace formaldehyde in gas samples.

Study Objectives and Overview


This application note demonstrates a validated method for detecting and quantifying gas-phase formaldehyde using 2,4-dinitrophenylhydrazine (DNPH) derivatization followed by gas chromatography with flame ionization detection (GC-FID). The key goals are:
  • To establish chromatographic conditions for separation of formaldehyde and acetaldehyde DNPH derivatives.
  • To evaluate sensitivity, reproducibility, and analysis time under optimized parameters.

Methodology and Applied Instrumentation


The analytical workflow consists of sample collection onto DNPH cartridges, thermal desorption of hydrazone derivatives, and GC-FID analysis. Key instrumentation details are summarized below:

Applied Instrumentation


  • Gas Chromatograph: Nexis GC-2030AF coupled with AOC-20i Plus auto injector
  • Column: SH-1, 30 m length, 0.32 mm I.D., 1.0 μm film thickness
  • Injection Mode: Split (1:30), 1.0 μL injection volume at 280 °C
  • Carrier Gas: Helium at constant linear velocity of 35 cm/s
  • Column Temperature Program: 240 °C hold for 8 min → ramp at 20 °C/min → 260 °C hold for 2 min
  • Detector: Flame ionization detector (FID) at 280 °C with H2 (32 mL/min), air (200 mL/min), and He makeup gas (24 mL/min)

Main Results and Discussion


Under the specified conditions, the method achieves clear baseline separation of the DNPH derivatives of formaldehyde and acetaldehyde at 100 ng/µL standards. Key performance metrics include:
  • High sensitivity with limits of detection in the low nanogram range.
  • Good repeatability with relative standard deviations below 3% across replicate injections.
  • Analysis cycle time under 12 minutes ensures high sample throughput.

Benefits and Practical Applications


The presented GC-FID method offers:
  • Reliable quantification of formaldehyde in workplace air monitoring and indoor environment assessments.
  • Compatibility with regulatory protocols for DNPH derivatization.
  • Ease of operation using standard GC-FID systems without the need for mass spectrometry.

Future Trends and Potential Uses


Advancements in sorbent materials and on-line derivatization could further streamline sample preparation. Integration with automated sampling and data processing will enhance real-time monitoring capabilities. Emerging applications include continuous indoor air quality networks and portable GC-FID units for field deployment.

Conclusion


This application demonstrates a rapid, sensitive, and reproducible GC-FID method for gas-phase formaldehyde analysis via DNPH derivatization. The approach meets analytical requirements for environmental health and safety monitoring while maintaining operational simplicity.

Reference


Shimadzu Corporation. Application News G327. First Edition: Sep. 2022. ERAS-1000-0345.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Gas-Phase Formaldehyde by GC-FID Method Utilizing DNPH Derivatization
Application News No. Gas Chromatography Analysis of Gas-Phase Formaldehyde by GC-FID Method Utilizing DNPH Derivatization G327 Formaldehyde has the simplest structure among the aldehydes, but is known to be a harmful air pollutant. Accordingly, various pretreatment processes for trace analysis…
Key words
formaldehyde, formaldehydecartridge, cartridgegas, gasdnph, dnphderivatization, derivatizationacetonitrile, acetonitrilephase, phasefid, fidftd, ftdflame, flameozone, ozonedetector, detectornews, newsnecessary, necessaryresin
An Examination of the Effects of High/Low Column Cooling Rates
ERAS-1000-0319 GC FID AOC SH Series SH-50 An Examination of the Effects of High/Low Column Cooling Rates 319 Keywords: Column Cooling Rates, Baseline noise, Drift 1. Acetone (Actn) 2. n-Propanol (nPrOH) 3. Ethyl Acetate (EtAc) 4. Isobutanol (iBuOH) 5. Isoamyl…
Key words
gas, gasfid, fidaoc, aocsplit, splitcarrier, carrierexamination, examinationmakeup, makeupdetector, detectormode, modecooling, coolinginjection, injectionrates, ratesmodel, modelcolumn, columneffects
Analysis of aromatics Using Automated Injection of Internal Standards
ERAS-1000-0358 GC FID AOC SH Series SH-PolarWax Analysis of aromatics Using Automated Injection of Internal Standards 358 Keywords: Aromatic compounds, Internal standard method 1. Benzene 2. Toluene 3. o-Xylene 4. Ethyl benzene 5. p-BFB 6. Tetralin Model Injection Mode Syringe…
Key words
injection, injectiongas, gaspolarwax, polarwaxaoc, aoccarrier, carrieraromatics, aromaticsmakeup, makeupdetector, detectormode, modesyringe, syringefid, fidmodel, modelsplit, splitseries, seriesinternal
Analysis of Methanol in Ethanol-based Sanitizer - No. 353
ERAS-1000-0353 GC FID AOC SH Series SH-Wax Analysis of Methanol in Ethanol-based Sanitizer 353 Keywords: Methanol concentration in ethanol Chromatograms of Gel Sanitizers Chromatograms of Liquid Sanitizers 1. Methanol 2. Ethyl acetate 3. Ethanol Main Unit Detector Column Column Temp.…
Key words
sanitizers, sanitizersdiluted, dilutedfid, fidmeoh, meohgel, geltax, taxsanitizer, sanitizeraoc, aocpredetermined, predeterminedchromatograms, chromatogramsetoh, etohgas, gascarrier, carrierwax, waxliquid
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike