Dynamic Headspace- Glass vs Metal
Applications | | CDS AnalyticalInstrumentation
Dynamic headspace sampling with adsorptive concentration has revolutionized trace analysis of volatile organic compounds across environmental, pharmaceutical, polymer, and food industries. Its high enrichment capacity enables detection of analytes at parts per trillion levels, improving sensitivity and selectivity in gas chromatographic assays.
This work compares glass-lined versus conventional metal flow paths in dynamic headspace systems to assess their impact on recovery of polar and heat-labile compounds. A case study involving volatile profiling of heated garlic illustrates practical performance under real sample conditions.
A dynamic headspace purge-and-trap protocol was implemented using glass-lined stainless steel, quartz, or fused silica tubing to minimize adsorption and degradation of sensitive analytes. Volatiles were purged from liquids or solids with helium at 30 ml min-1, trapped on ambient Tenax, thermally desorbed at 250°C, and cryogenically refocused at –100°C before transfer to the GC.
The glass-lined dynamic headspace approach offers:
Advances may include integration with mass spectrometry for structural identification, development of new sorbents for selective trapping, and automation of dynamic headspace systems for high-throughput screening in industrial quality control.
Replacing metal tubing with glass-lined or inert fused silica paths in dynamic headspace systems significantly boosts recovery of challenging analytes and expands the technique’s applicability. Cryofocusing onto capillary columns further enhances chromatographic performance, making this approach a powerful tool for trace volatile analysis.
GC, HeadSpace
IndustriesMaterials Testing
ManufacturerCDS Analytical
Summary
Significance of the Topic
Dynamic headspace sampling with adsorptive concentration has revolutionized trace analysis of volatile organic compounds across environmental, pharmaceutical, polymer, and food industries. Its high enrichment capacity enables detection of analytes at parts per trillion levels, improving sensitivity and selectivity in gas chromatographic assays.
Objectives and Overview
This work compares glass-lined versus conventional metal flow paths in dynamic headspace systems to assess their impact on recovery of polar and heat-labile compounds. A case study involving volatile profiling of heated garlic illustrates practical performance under real sample conditions.
Methodology and Instrumentation
A dynamic headspace purge-and-trap protocol was implemented using glass-lined stainless steel, quartz, or fused silica tubing to minimize adsorption and degradation of sensitive analytes. Volatiles were purged from liquids or solids with helium at 30 ml min-1, trapped on ambient Tenax, thermally desorbed at 250°C, and cryogenically refocused at –100°C before transfer to the GC.
Main Results and Discussion
- Glass-lined pathways enhanced recovery of polar compounds including 1,3-butanediol, 1,4-butanediol, methyl hexanoate, octanol, 2,6-dimethylphenol, and 2,4-dimethylaniline compared to metal tubing.
- Certain analytes undetected in the metal system were successfully quantified with the glass-lined configuration.
- In the garlic application, warming 2 mg of raw garlic to 70°C and purging for 10 minutes yielded a complex profile of sulfur and oxygen containing volatiles effectively trapped and analyzed.
Benefits and Practical Applications
The glass-lined dynamic headspace approach offers:
- Improved analyte recovery for polar and thermally sensitive compounds
- Enhanced sensitivity with no split injection losses
- Broad applicability to environmental, pharmaceutical, polymer, and food analyses
Future Trends and Potential Applications
Advances may include integration with mass spectrometry for structural identification, development of new sorbents for selective trapping, and automation of dynamic headspace systems for high-throughput screening in industrial quality control.
Conclusion
Replacing metal tubing with glass-lined or inert fused silica paths in dynamic headspace systems significantly boosts recovery of challenging analytes and expands the technique’s applicability. Cryofocusing onto capillary columns further enhances chromatographic performance, making this approach a powerful tool for trace volatile analysis.
Instrumentation
- Valve oven and transfer line: 250°C
- Sample vessel heating: 700°C for 10 minutes
- Trap desorption: 250°C for 10 minutes
- Cryofocus: –100°C for 10 minutes
- GC-FID: 0.53 mm × 20 m SE-54 column, helium carrier at 2 psi, oven program 50°C for 2 minutes, ramp 7°C/min to 250°C
Reference
- T. Wampler, W. Bowe, E. Levy, Splitless capillary GC analysis of herbs and spices using cryofocusing, American Lab, October 1985.
- T. Wampler, W. Bowe, J. Higgins, E. Levy, Systems approach to automatic cryofocusing in purge and trap, headspace, and pyrolytic analysis, American Lab, August 1985.
- S. Jacobsson, Analysis of volatile organic compounds in polymers by dynamic headspace and gas chromatography mass spectrometry, HRC & CC7, pages 185-190, 1984.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Dynamic Headspace Analysis of Food Volatiles Coffee and Tea
|CDS Analytical|Applications
#12a Dynamic Headspace Analysis of Food Volatiles Coffee and Tea Application Note Food & Flavor Author: T. Wampler Because the aroma and flavor of foods are due largely to volatile organic compounds, headspace analysis is very popular in food research…
Key words
chromatograms, chromatogramscompounds, compoundsfood, foodheadspace, headspacerefocused, refocuseddynamic, dynamicswept, sweptaccompanying, accompanyingconstantly, constantlycoffee, coffeetenax, tenaxorganics, organicsconcentrator, concentratoritself, itselftea
Dynamic Headspace of Citrus Peels
|CDS Analytical|Applications
Dynamic Headspace of Citrus Peels Application Note Food and Flavor Author: T. Wampler Headspace sampling of volatiles from food samples is a convenient way of introducing aroma constituents into a gas chromatograph. Traditional or “static” headspace techniques involve placing the…
Key words
peels, peelscryogenic, cryogenicrefocusing, refocusingbackflushed, backflushedpyroprobe, pyroprobecitrus, citrustenax, tenaxlimonene, limonenetrapped, trappedflavor, flavorespecially, especiallyheated, heatedambient, ambienttemp, tempheadspace
Dynamic Headspace Analysis of Spices
|SCION Instruments|Applications
#2a Dynamic Headspace Analysis of Spices Application Note Environmental Industry Author: T. Wampler Headspace sampling for gas chromatography is a process by which the volatile organic compounds from a sample are collected and injected onto the GC. In static headspace…
Key words
chemical, chemicalpermitting, permittingsensitivity, sensitivitycryofocuser, cryofocuserheadspace, headspaceminutes, minutesspices, spicestenax, tenaxlost, lostconcentrator, concentratornone, noneroom, roomexamples, examplespacked, packedprimary
Analysis of Waste Solvent by Purge and Trap
|CDS Analytical|Applications
#3a Analysis of Waste Solvent by Purge and Trap Application Note Environmental Industry The purge and trap technique was originally developed for the analysis of organic volatiles in water samples. The process involves passing a stream of inert gas through…
Key words
technique, techniquepurge, purgetrap, traporiginally, originallyindustry, industrywaste, wasteenvironmental, environmentaldeveloped, developedorganic, organicanalysis, analysissolvent, solventnote, noteapplication, applicationsample