Fast analysis of coal tar polycyclic aromatic hydrocarbons on Agilent J&W Select PAH
Applications | 2011 | Agilent TechnologiesInstrumentation
Polycyclic aromatic hydrocarbons (PAHs) are priority environmental pollutants with known carcinogenic and mutagenic properties. Coal tar is a highly complex matrix containing dozens of PAH isomers that share identical masses, making their reliable separation and quantitation by gas chromatography–mass spectrometry (GC–MS) extremely challenging. High selectivity columns and optimized temperature programs are crucial to resolving critical isomer pairs and achieving accurate results in environmental monitoring, industrial quality control, and regulatory compliance.
This application note demonstrates a fast, single-run GC–MS method for analyzing a certified coal tar reference material (NIST SRM 1597a) using the Agilent J&W Select PAH capillary column. Key objectives include:
Instrumentation and conditions were as follows:
The optimized method achieved complete separation of regulated PAH isomers in a 29-minute run. Highlights include:
The Agilent J&W Select PAH column, combined with an optimized oven program, provides a reliable, fast, and high-resolution solution for comprehensive coal tar PAH analysis. This method delivers accurate separation and quantitation of all regulated PAH isomers in a single 29-minute run, meeting the demands of modern analytical laboratories.
GC/MSD, GC/MS/MS, GC/QQQ, GC columns, Consumables
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Significance of the topic
Polycyclic aromatic hydrocarbons (PAHs) are priority environmental pollutants with known carcinogenic and mutagenic properties. Coal tar is a highly complex matrix containing dozens of PAH isomers that share identical masses, making their reliable separation and quantitation by gas chromatography–mass spectrometry (GC–MS) extremely challenging. High selectivity columns and optimized temperature programs are crucial to resolving critical isomer pairs and achieving accurate results in environmental monitoring, industrial quality control, and regulatory compliance.
Aims and Overview of the Study
This application note demonstrates a fast, single-run GC–MS method for analyzing a certified coal tar reference material (NIST SRM 1597a) using the Agilent J&W Select PAH capillary column. Key objectives include:
- Evaluating column selectivity for over 30 PAHs and isomers in a single injection.
- Optimizing oven temperature programming to achieve baseline separation of critical PAH pairs.
- Demonstrating method speed and robustness in full-scan (FS) and selected ion monitoring (SIM) modes.
Methodology and Instrumentation
Instrumentation and conditions were as follows:
- GC–MS System: Triple quadrupole GC–MS.
- Column: Agilent J&W Select PAH, 15 m × 0.15 mm ID, 0.10 μm film thickness.
- Carrier Gas: Helium, constant flow at 1.2 mL/min.
- Injection: 1.0 μL in SIM mode and 0.1 μL in FS mode, splitless at 300 °C.
- Oven Program:
• Hold at 70 °C for 0.40 min
• Ramp to 180 °C at 70 °C/min
• Ramp to 230 °C at 7 °C/min, hold 7 min
• Ramp to 280 °C at 50 °C/min, hold 7 min
• Ramp to 350 °C at 30 °C/min, hold 4 min - Detection: Electron ionization (EI) in FS or SIM, ion source at 275 °C, transfer line at 300 °C.
Main Results and Discussion
The optimized method achieved complete separation of regulated PAH isomers in a 29-minute run. Highlights include:
- Clear resolution of chrysene (66 mg/kg) from triphenylene (12 mg/kg).
- Separation of benzo[b,k,j]fluoranthenes (66, 37, 41 mg/kg) from benzo[a]pyrene (94 mg/kg).
- Stable mass spectra with predominant molecular ions (e.g., m/z 228 for chrysene) and limited fragmentation.
- Sensitivity in SIM mode allowed quantitation of compounds spanning four orders of magnitude (e.g., naphthalene at 1030 mg/kg down to dibenzo[a,e]pyrene at 9 mg/kg).
- Critical isomer groups—benzofluorenes, methylchrysenes, dibenzoanthracenes and dibenzopyrenes—were baseline separated, demonstrating column selectivity.
Benefits and Practical Applications
- Single-shot analysis of complex PAH mixtures without false positives.
- High throughput with rapid temperature ramps reducing total cycle time.
- Robust high-temperature stability for extended column lifetime.
- Applicable to environmental, food, and industrial QA/QC laboratories for regulatory PAH monitoring.
Future Trends and Opportunities
- Integration with high-resolution mass spectrometry for enhanced selectivity in ultra-trace analysis.
- Automation and coupling with sample extraction systems for higher sample throughput.
- Development of shorter, narrower-bore selective phases to further reduce runtimes.
- Application to emerging PAH derivatives and alkylated homologs in environmental matrices.
Conclusion
The Agilent J&W Select PAH column, combined with an optimized oven program, provides a reliable, fast, and high-resolution solution for comprehensive coal tar PAH analysis. This method delivers accurate separation and quantitation of all regulated PAH isomers in a single 29-minute run, meeting the demands of modern analytical laboratories.
References
- Poster DL, Schantz MM, Sander LC, Wise SA (2006) Analysis of polycyclic aromatic hydrocarbons (PAHs) in environmental samples: a critical review of gas chromatographic methods. Anal Bioanal Chem 386:859–881.
- Sander LC, Wise SA (1997) Polycyclic Aromatic Hydrocarbon Structure Index. NIST Special Publication. National Institute of Standards and Technology, Gaithersburg, MD.
- Wenzl T, Simon R, Kleiner J, Anklam E (2006) Analytical methods for polycyclic aromatic hydrocarbons (PAHs) in food and the environment needed for new food legislation in the European Union. Trends Anal Chem 25(7):716–725.
- Ziegenhals K, Hübschmann HJ, Speer K, Jira W (2008) Fast-GC/HRMS to quantify the EU priority PAH. J Sep Sci 31:1779–1786.
- Beekman M, Boersma AHR, Sijm DTHM (2008) Coal-tar pitch high temperature (CTPHT), transitional arrangements and way forward under REACH. REACH-SEA scoping study report. National Institute for Public Health and the Environment, The Netherlands.
- Bordajandi LR et al. (2008) Optimisation of the GC-MS conditions for the determination of the 15 EU foodstuff priority PAHs. J Sep Sci 31:1769–1778.
- Gómez-Ruiz JA, Wenzl T (2009) Evaluation of gas chromatography columns for the analysis of the 15+1 EU-priority PAHs. Anal Bioanal Chem 393:1697–1707.
- Lerda D (2009) Polycyclic Aromatic Hydrocarbons Factsheet. European Commission Joint Research Centre.
- NIST SRM 1597a. Complex mixture of PAHs from coal tar. National Institute of Standards and Technology, Gaithersburg, MD.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Complex polycyclic aromatic hydrocarbons from coal tar on Agilent J&W Select PAH
2011|Agilent Technologies|Applications
Complex polycyclic aromatic hydrocarbons from coal tar on Agilent J&W Select PAH Application Note Author Introduction John Oostdijk Agilent Technologies, Inc. Coal tar is a brown or black liquid of high viscosity, which smells of naphthalene and aromatic hydrocarbons and…
Key words
benzo, benzobap, bapdbaep, dbaepbghip, bghipbjfa, bjfabkfa, bkfafluoranthene, fluoranthenebaa, baapolycyclic, polycyclictar, tarbbfa, bbfacoal, coalaromatic, aromaticchrysene, chrysenebafa
Analysis of Low-level Polycyclic Aromatic Hydrocarbons (PAHs) in Rubber and Plastic Articles Using Agilent J&W DB-EUPAH GC column
2010|Agilent Technologies|Applications
Analysis of Low-level Polycyclic Aromatic Hydrocarbons (PAHs) in Rubber and Plastic Articles Using Agilent J&W DB-EUPAH GC column Application Note Gas Chromatography/Mass Spectrometry Authors Abstract Yun Zou and Chongtian Yu Agilent J&W DB-EUPAH GC columns are designed for the analysis…
Key words
benzo, benzofluoranthene, fluoranthenepyrene, pyreneanthracene, anthracenepahs, pahsrubber, rubberabundance, abundanceacenaphthylene, acenaphthyleneextender, extenderacenaphthene, acenaphthenefluorene, fluorenephenanthrene, phenanthrenechrysene, chryseneeupah, eupahplastic
Quantification of 15+1 EU Priority PAH in representative German smoked meat products using a Fast GC/HRMS Method
|Thermo Fisher Scientific|Posters
Quantification of 15+1 EU Priority PAH in representative German smokedAnalytik meat products using a Fast GC/HRMS Method. Arbeitsgruppe Wolfgang Jira1, Fredi Schwägele1, Hans-Joachim Huebschmann2, *Heinz Mehlmann2 1. Analysis Division, Max Rubner-Institut (MRI), Kulmbach, Germany 2. POPs Center of Excellence, Thermo…
Key words
benzo, benzopyrene, pyrenedibenzo, dibenzobap, bapxuoranthene, xuoranthenepah, pahsmoked, smokedmeat, meatbgp, bgpdlp, dlpdip, dipbjf, bjfbbf, bbfbkf, bkfbaa
Fast and Accurate GC-MS Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) in Rubber and Plastic using a Zebron™ ZB-PAH-EU GC Column
2020|Phenomenex|Applications
TN-2091 APPLICATIONS Ramkumar Dhandapani, Ph.D. Product Manager - Gas Chromatography Fast and Accurate GC-MS Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) in Rubber and Plastic using a Zebron™ ZB-PAH-EU GC Column He has a PhD in Analytical Chemistry and over 15…
Key words
zebron, zebronbenzo, benzopah, pahfluoranthene, fluoranthenepyrene, pyrenerubber, rubberplastic, plasticreproducing, reproducinganthracene, anthracenelove, lovetrouble, troubleliner, linerbenz, benztaper, tapermeter