Exploration of New Low-Pressure GC Columns for Food and Environmental Emerging Contaminants
Posters | 2024 | Restek | RAFAInstrumentation
Emerging food and environmental contaminants demand analytical methods that are both rapid and resource-efficient. Low-pressure gas chromatography (LPGC) delivers shorter runtimes and reduced carrier gas consumption while preserving the resolution needed for trace-level analysis.
This study assessed four new LPGC column chemistries—cyanopropylphenyl dimethylpolysiloxane; 50% dimethylsiloxane/50% diphenyl; 65% dimethylsiloxane/35% diphenyl; and trifluoropropylmethyl polysiloxane—for the determination of nitrosamines, alkylfurans, phthalates, arylamines, and fluorotelomer alcohols. Key performance metrics were runtime, helium usage, and chromatographic resolution compared to conventional GC columns.
The LPGC configuration included a GC inlet connected to a retention gap and restrictor, feeding a short analytical column directly coupled to a mass spectrometer. Main components and parameters:
LPGC methods exhibited an average 54% reduction in analysis time, with some applications achieving up to 3.3× faster runs. Helium consumption decreased by an average of 66%, reaching up to 81% savings. Chromatographic resolution remained acceptable for all target analytes, and lower elution temperatures (50–80 °C less) contributed to reduced column bleed.
Deployment of LPGC is expected to expand across environmental, food safety, and industrial laboratories. Integration with high-resolution MS will address complex sample matrices. Advances in column design and miniaturization may enable portable, field-deployable LPGC systems.
New low-pressure GC column kits provide a versatile, faster, and greener approach to analyze emerging contaminants without major changes to existing instrumentation. They achieve substantial time and helium savings while preserving chromatographic performance.
No specific references were provided in the source document.
Consumables, GC
IndustriesFood & Agriculture, Environmental
ManufacturerRestek
Summary
Importance of the Topic
Emerging food and environmental contaminants demand analytical methods that are both rapid and resource-efficient. Low-pressure gas chromatography (LPGC) delivers shorter runtimes and reduced carrier gas consumption while preserving the resolution needed for trace-level analysis.
Objectives and Study Overview
This study assessed four new LPGC column chemistries—cyanopropylphenyl dimethylpolysiloxane; 50% dimethylsiloxane/50% diphenyl; 65% dimethylsiloxane/35% diphenyl; and trifluoropropylmethyl polysiloxane—for the determination of nitrosamines, alkylfurans, phthalates, arylamines, and fluorotelomer alcohols. Key performance metrics were runtime, helium usage, and chromatographic resolution compared to conventional GC columns.
Methodology and Instrumentation
The LPGC configuration included a GC inlet connected to a retention gap and restrictor, feeding a short analytical column directly coupled to a mass spectrometer. Main components and parameters:
- Analytical columns: 0.32 mm or 0.53 mm i.d. capillaries with diverse stationary phases
- Retention gap and restrictor to maintain optimal inlet pressure
- Mass spectrometer detection enabling selective and sensitive quantification
- Standard injection techniques supporting high-volume injections
Main Results and Discussion
LPGC methods exhibited an average 54% reduction in analysis time, with some applications achieving up to 3.3× faster runs. Helium consumption decreased by an average of 66%, reaching up to 81% savings. Chromatographic resolution remained acceptable for all target analytes, and lower elution temperatures (50–80 °C less) contributed to reduced column bleed.
Benefits and Practical Applications
- Significantly shorter analysis times improve laboratory throughput
- Reduced carrier gas usage lowers operating costs and environmental impact
- Sharper peaks can enhance sensitivity in MS detection
- Compatibility with existing GC–MS systems and injection workflows
- Flexibility to select appropriate stationary phases for diverse contaminant classes
Future Trends and Applications
Deployment of LPGC is expected to expand across environmental, food safety, and industrial laboratories. Integration with high-resolution MS will address complex sample matrices. Advances in column design and miniaturization may enable portable, field-deployable LPGC systems.
Conclusion
New low-pressure GC column kits provide a versatile, faster, and greener approach to analyze emerging contaminants without major changes to existing instrumentation. They achieve substantial time and helium savings while preserving chromatographic performance.
References
No specific references were provided in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
An Introduction to Low-Pressure GC-MS (LPGC-MS)
2021|Restek|Technical notes
An Introduction to Low-Pressure GC-MS (LPGC-MS) Leverage Your MS Vacuum to Significantly Speed Up Analyses • 3x faster multiresidue pesticides analysis in foods. • Factory-coupled, leak-free kit makes setting up LPGC as simple as a column change. • Ideal for…
Key words
lpgc, lpgcsep, sepquechers, quecherskit, kitdspe, dspetubes, tubesaccelerator, acceleratororganonitrogen, organonitrogenmultiresidue, multiresiduepackets, packetsextraction, extractionorganophosphorus, organophosphorusdescription, descriptioncompounds, compoundstopaz
Speed Up Multiresidue Pesticides Analysis in Food with Low-Pressure GC-MS
2021|Thermo Fisher Scientific|Technical notes
Featured Application: LPGC-MS Pesticides Analysis in Strawberries Speed Up Multiresidue Pesticides Analysis in Food with Low-Pressure GC-MS • 3x faster than conventional methods. • Get the speed advantage of LPGC-MS with a simple column change. • Robust, factory-coupled column kit…
Key words
quechers, quechersdspe, dspeship, shipsep, sepampul, ampulshelf, shelfcolder, colderkit, kitdate, datelife, lifecleanup, cleanupshipping, shippingambient, ambientcrm, crmtubes
Rxi® GC Columns - Leverage the outstanding inertness, low bleed, and high reproducibility of Rxi® 3-in-1 technology to gain
2015|Restek|Brochures and specifications
Leverage the outstanding inertness, low bleed, and high reproducibility of Rxi® 3-in-1 technology to gain: • Accurate Data • The Right Results Fast • Maximized Instrument Uptime Pure Chromatography www.restek.com/rxi Rxi®-1ms............................................................................... 2 Rxi®-5ms............................................................................... 3 Rxi®-XLB................................................................................ 3 Rxi®-5Sil MS......................................................................... 4 Rxi®-35Sil…
Key words
price, pricerxi, rxinonpolar, nonpolarcrossbond, crossbondstabilwax, stabilwaxxlb, xlbmidpolarity, midpolarityrestek, restekgcxgc, gcxgcphases, phasessimilar, similarphase, phasepolysiloxane, polysiloxanecolumns, columnsguard
Zebron PLUS GC Columns - Improved Inertness, Better Peak Shape, Low Bleed, and Aqueous Stability
2019|Phenomenex|Brochures and specifications
Zebron PLUS GC Columns Improved Inertness, Better Peak Shape, Low Bleed, and Aqueous Stability ™ G C C o lu mns ZB-1plus ZB-5plus ZB-5MSplus ZB-624plus ZB-WAXplus ™ ™ ™ ™ ™ PLUS www.phenomenex.com/GCplus Exceptionally Versatile GC phases PLUS Exceptional Inertness…
Key words
zebron, zebronphenomenex, phenomenexbleed, bleedfilter, filterclick, clickwaxplus, waxplusindicator, indicatorgas, gasmin, mintrap, trapsolvents, solventsplus, pluselectronic, electronicbetter, betterwall