Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059
Applications | 2004 | Agilent TechnologiesInstrumentation
Crude oil frequently contains trace amounts of methanol added to inhibit gas hydrate formation. Accurate determination of methanol at low ppm levels is critical to prevent oxygen-related issues during downstream refining. The complex hydrocarbon matrix complicates direct analysis by conventional one-dimensional GC, making two-dimensional heart-cutting GC a valuable tool for selective isolation and quantification of volatile impurities.
This study adapts ASTM Method D7059 for trace methanol in crude oil using a novel microfluidic Deans switch integrated into an Agilent 6890N GC. The primary goal is to demonstrate improved separation efficiency, sensitivity, and robustness for methanol levels ranging from 15 to 900 ppm in a challenging matrix.
Heart-cutting between 1.70 and 2.35 min successfully transferred methanol and 1-propanol to the secondary column, yielding retention at 4.72 min and 6.38 min respectively. Calibration over 5–1000 ppm achieved a correlation coefficient >0.9999. A 1 ppm methanol standard delivered a signal-to-noise ratio of 5:1, satisfying method detection limits. Analysis of quality control samples at 15 and 670 ppm returned recoveries within ±5 ppm and ±35 ppm, meeting ASTM acceptance criteria. Implementation of column backflushing after 7 min reduced cycle time and prevented residual high-boiling contamination.
The microfluidic switch’s reduced thermal mass ensures uniform heating and eliminates cold spots that could condense heavy hydrocarbons. Its inert internal surfaces and simplified plumbing minimize dead volume and adsorption, enhancing peak shape and sensitivity. Metal ferrule interfaces provide reliable capillary connections under thermal cycling, reducing maintenance and leak risk. This configuration streamlines trace alcohol quantification in petrochemical QA/QC and research laboratories.
Advances in microfluidic integration may extend two-dimensional GC to broader compound classes and multi-heart-cut schemes. Further miniaturization and automated valve control could enable field-deployable systems for on-site hydrocarbon monitoring. Coupling with mass spectrometry or novel detectors may broaden applicability to trace contaminants in fuels, environmental samples, and process streams.
The new microfluidic Deans switch integrated with 2-D GC on the Agilent 6890N platform meets and surpasses the performance criteria of ASTM Method D7059 for methanol analysis in crude oil. Its low thermal mass, inert flow paths, and robust column connections deliver repeatable, sensitive results, offering a reliable solution for industrial trace analysis.
GC, GCxGC
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Crude oil frequently contains trace amounts of methanol added to inhibit gas hydrate formation. Accurate determination of methanol at low ppm levels is critical to prevent oxygen-related issues during downstream refining. The complex hydrocarbon matrix complicates direct analysis by conventional one-dimensional GC, making two-dimensional heart-cutting GC a valuable tool for selective isolation and quantification of volatile impurities.
Aims and Study Overview
This study adapts ASTM Method D7059 for trace methanol in crude oil using a novel microfluidic Deans switch integrated into an Agilent 6890N GC. The primary goal is to demonstrate improved separation efficiency, sensitivity, and robustness for methanol levels ranging from 15 to 900 ppm in a challenging matrix.
Methodology and Instrumentation
- Gas chromatograph: Agilent 6890N with split/splitless inlet, dual flame ionization detectors, pneumatics control module, and autosampler
- Primary column: 10 m × 0.53 mm id DB-1 capillary (polydimethylsiloxane)
- Secondary column: 10 m × 0.53 mm id CP-Lowox for oxygenated compound separation
- Microfluidic Deans switch: etched stainless steel device featuring integrated flow paths, low thermal mass, surface-deactivated channels, and metal ferrule connections for leak-free high-temperature operation
- Data system: Agilent ChemStation for instrument control, heart-cut timing calculation, and quantitative analysis
- Sample preparation: crude oil spiked with methanol and 1-propanol internal standard following ASTM D7059 procedures
Main Results and Discussion
Heart-cutting between 1.70 and 2.35 min successfully transferred methanol and 1-propanol to the secondary column, yielding retention at 4.72 min and 6.38 min respectively. Calibration over 5–1000 ppm achieved a correlation coefficient >0.9999. A 1 ppm methanol standard delivered a signal-to-noise ratio of 5:1, satisfying method detection limits. Analysis of quality control samples at 15 and 670 ppm returned recoveries within ±5 ppm and ±35 ppm, meeting ASTM acceptance criteria. Implementation of column backflushing after 7 min reduced cycle time and prevented residual high-boiling contamination.
Benefits and Practical Applications
The microfluidic switch’s reduced thermal mass ensures uniform heating and eliminates cold spots that could condense heavy hydrocarbons. Its inert internal surfaces and simplified plumbing minimize dead volume and adsorption, enhancing peak shape and sensitivity. Metal ferrule interfaces provide reliable capillary connections under thermal cycling, reducing maintenance and leak risk. This configuration streamlines trace alcohol quantification in petrochemical QA/QC and research laboratories.
Future Trends and Potential Applications
Advances in microfluidic integration may extend two-dimensional GC to broader compound classes and multi-heart-cut schemes. Further miniaturization and automated valve control could enable field-deployable systems for on-site hydrocarbon monitoring. Coupling with mass spectrometry or novel detectors may broaden applicability to trace contaminants in fuels, environmental samples, and process streams.
Conclusion
The new microfluidic Deans switch integrated with 2-D GC on the Agilent 6890N platform meets and surpasses the performance criteria of ASTM Method D7059 for methanol analysis in crude oil. Its low thermal mass, inert flow paths, and robust column connections deliver repeatable, sensitive results, offering a reliable solution for industrial trace analysis.
References
- Annual Book of ASTM Standards Vol 05.04 Petroleum Products and Lubricants IV ASTM International 2004
- McCurry JD Quimby BD Two-dimensional Gas Chromatographic Analysis of Components in Fuel and Fuel Additives Using a Simplified Heart-Cutting GC System J Chromatogr Sci 2002 41 524–527
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Oxygenates in Fuels –Analysis Solutions From Trace Levels to Ethanol Fuels
2010|Agilent Technologies|Presentations
Oxygenates in Fuels – Analysis Solutions From Trace Levels to Ethanol Fuels Page 1 James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA October 13, 2010 Application Summary There is a need to measure trace oxygenates in: • Feedstocks…
Key words
oxygenates, oxygenatesethanol, ethanolgasoline, gasolineastm, astmoxygenate, oxygenateswitch, switchfluidic, fluidicfuel, fuelmethanol, methanolepc, epcether, ethercft, cftaux, auxtrace, tracedipe
Time-of-Flight Mass Spectrometry
2004|Agilent Technologies|Applications
Two-Dimensional Gas Chromatographic Analysis of Trace Benzene in Styrene Application HPI Authors Introduction Chunxiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai 200131 P.R. China The American Society of Testing and Materials (ASTM) method…
Key words
benzene, benzeneinnowax, innowaxstyrene, styreneprefractionator, prefractionatorheart, heartcutting, cuttingdeans, deansfid, fidhydrocarbons, hydrocarbonsswitch, switchmode, modecut, cutpcm, pcmarea, arearestrictor
Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection
2003|Agilent Technologies|Applications
Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract A two-dimensional gas chromatography system…
Key words
thiophene, thiopheneinnowax, innowaxbenzene, benzenedeans, deansfid, fidcolumn, columncut, cutswitch, switchplot, plotheart, heartsulfur, sulfurepc, epcexpensive, expensiveseparated, separatedrestrictor
Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS
2006|Agilent Technologies|Applications
Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS Application Note Forensic Toxicology Authors Dean F. Fritch, Ph.D., DABFT, DABCC-TC Orasure Technologies, Inc. 150 Webster Street Bethlehem, PA 18015 USA Bruce D. Quimby, Ph.D. Agilent Technologies, Inc. 2850 Centerville…
Key words
thc, thcdeans, deansoral, oralmsd, msdswitch, switchsolenoid, solenoidmin, minpcm, pcmtemp, tempwashes, washesfid, fidrestrictor, restrictorfluid, fluidtime, timecut