The 5973N inert MSD: Using Higher Ion Source Temperatures

Technical notes | 2004 | Agilent TechnologiesInstrumentation
GC/MSD, GC/SQ
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the topic


The performance of gas chromatography–mass spectrometry (GC–MS) for challenging, late-eluting analytes depends critically on ion source conditions. Raising the ion source temperature can enhance volatility of high-boiling compounds, reduce peak tailing, and improve detectability of persistent organic pollutants (POPs) and polycyclic aromatic hydrocarbons (PAHs). Understanding these effects helps analytical chemists optimize sensitivity and data quality in environmental, industrial, and quality-control laboratories.

Study objectives and overview


This application note evaluates the new 5973N inert mass selective detector (MSD) coupled with ChemStation software revision DA, focusing on extended ion source temperature capability (230 °C to 300 °C). The work investigates autotune spectra, chromatographic response for polychlorinated biphenyls (PCBs) and coronene, and develops automated bake-out routines to maintain instrument performance.

Methodology and instrumentation


A comprehensive approach was implemented:
  • Instrument: Agilent 5973N inert MSD equipped with modified ion source capable of 230 °C–300 °C operation.
  • Software: ChemStation G1701DA for temperature control, autotuning, and macro execution.
  • Tuning: Autotune at 230 °C and 300 °C using perfluorotributylamine (PFTBA); comparison of fragment abundances and isotopic ratios.
  • Chromatography: Full-scan (50–505 amu) and selected-ion-monitoring (SIM) analyses of six PCB congeners; full-scan analysis of coronene.
  • Bake-out procedure: Custom ChemStation macro to heat source (300 °C) and quadrupole (200 °C) overnight, followed by cool-down and system checkout.

Main results and discussion


Autotune spectra showed reduced relative abundance of high-mass PFTBA fragments at 300 °C, while isotopic ratios remained stable. Full-scan GC–MS of PCBs indicated increased total ion current for late-eluting congeners but also higher background noise, resulting in lower signal-to-noise (S/N) gains. In SIM mode, background remained unchanged, and late-eluting PCBs exhibited significant S/N improvements. Coronene analysis demonstrated that, despite similar peak areas, higher source temperature produced a more Gaussian peak shape, facilitating better detection of high-boiling PAHs.

Benefits and practical applications


The higher ion source temperature on the 5973N inert MSD offers:
  • Improved volatility and peak shapes for high-boiling analytes (POPs, PAHs).
  • Enhanced sensitivity and S/N in SIM mode for late-eluting compounds.
  • Stable isotopic abundance measurements despite increased fragmentation.
  • Automated bake-out routines to reduce air-water background and maintain consistent performance.

Future trends and opportunities


Continued development may include integration of dynamic source temperature programming to optimize fragmentation and sensitivity across a wider range of analytes. Automated maintenance workflows with remote monitoring and predictive bake-out scheduling could further streamline instrument uptime. Expanding high-temperature GC–MS methods will support emerging applications in environmental monitoring, metabolomics of nonvolatile compounds, and forensic analyses.

Conclusion


Leveraging higher ion source temperatures on the Agilent 5973N inert MSD can significantly benefit analysis of late-eluting, high-boiling compounds when applied judiciously. While full-scan methods may suffer increased background, SIM acquisitions maintain cleaner baselines and deliver improved detection limits. Automated bake-out protocols ensure reliable performance after source maintenance, making advanced temperature control a valuable asset in routine analytical workflows.

References


  1. Prest H, Thomson C. The 5973N inert MSD: Using Higher Ion Source Temperatures. Agilent Technologies Application Note 5989-0678EN, 2004.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
GC/MS Analysis of PCBs in Waste Oil Using the Backflush Capability of the Agilent QuickSwap Accessory
GC/MS Analysis of PCBs in Waste Oil Using the Backflush Capability of the Agilent QuickSwap Accessory Application Environmental Authors Frank David Research Institute for Chromatography Pres. Kennedypark 26 8500 Kortrijk Belgium Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington,…
Key words
backflush, backflushquickswap, quickswapscan, scanpcbs, pcbssim, simoil, oilwaste, wastepcb, pcbpressure, pressuretics, ticschromatograms, chromatogramsagilent, agilentoven, ovenprogram, programbiphenyls
Agilent JetClean: In-situ GC/MS Ion Source Cleaning and Conditioning
Agilent JetClean: In-situ GC/MS Ion Source Cleaning and Conditioning Application Note Author Abstract Harry Prest GC/MS maintenance is a universal requirement for maintaining analytical targets R&D Chemist during sample analysis. Rapid intra-column backflushing, one vital enhancement Life Sciences and in…
Key words
source, sourceclean, cleanjetclean, jetcleanacquire, acquirehydrogen, hydrogenacquisition, acquisitiontune, tuneemission, emissionbatches, batchesfile, filesetting, settingresponse, responseonly, onlyconsecutive, consecutivehydrogenic
SOLUTIONS THAT MEET YOUR DEMANDS FOR FORENSIC TOXICOLOGY
SOLUTIONS THAT MEET YOUR DEMANDS FOR FORENSIC TOXICOLOGY Excellent choices for forensic toxicology applications products I applications I software I services FORENSIC TOXICOLOGY > Search entire document Forensic toxicology analysis determines the absence or presence of drugs and their metabolites,…
Key words
toxicology, toxicologyforensic, forensicblood, blooddrugs, drugsmin, mintms, tmscocaine, cocainebenzodiazepines, benzodiazepinestime, timemass, massurine, urineagilent, agilentusing, usingalprazolam, alprazolamdiazepam
Solutions that meet your demands - Environmental Compendium - Emerging Contaminants
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Emerging Contaminants PPCPs Applications > Return to Table of Contents > Search entire document Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive…
Key words
pbde, pbdebde, bdepfoa, pfoapbdes, pbdesabundance, abundancepfos, pfospolybrominated, polybrominatedusa, usaion, ionmass, massdiphenyl, diphenylagilent, agilentfragmentor, fragmentorcompounds, compoundscollision
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike