What’s in Your Beer? GC/MS Static Head Space with an Agilent J&W DB-624 Ultra Inert Capillary GC Column
Posters | 2013 | Agilent Technologies | PittconInstrumentation
GC/MSD, HeadSpace, GC/SQ, GC columns, Consumables
IndustriesFood & Agriculture
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Profiling volatile and semi-volatile components in beer is essential for quality control, flavor development and authentication. Static headspace GC/MS allows rapid, solvent-free analysis of key flavor markers such as alcohols, esters, aldehydes and organic acids, enabling breweries and analytical laboratories to monitor brewing processes and ensure product consistency.Study Objectives and Overview
The study evaluates the performance of static headspace GC/MS coupled with the Agilent J&W DB-624 Ultra Inert capillary column. Typical beer styles—including light lager, pale ale, wheat beer and wet hopped ale—were profiled to compare separation efficiency, peak shape and sensitivity for a range of target analytes. A comparison with a competitor’s 624 phase column highlights improvements in organic acid detection.Methodology and Used Instrumentation
- Sample preparation: 10 mL beer aliquots in 20 mL crimp-cap headspace vials, equilibrated at 85 °C for 20 min.
- Headspace sampler: Agilent 7697A (oven 85 °C, loop 1 mL at 85 °C, transfer line 100 °C).
- Gas chromatograph: Agilent 7890 equipped with an Agilent J&W DB-624UI column (30 m × 0.25 mm × 1.4 µm).
- Carrier gas: Helium at ~1.8 mL/min, constant flow.
- Inlet: Split 5:1 at 220 °C with Ultra Inert liner.
- Detector: MSD (scan range 30–400 amu) and FID (H₂, air and N₂ makeup flows).
Main Results and Discussion
- Separation of fusel alcohols, esters, aldehydes and organic acids was achieved within a 26 min program, with clear resolution of amyl and iso-amyl compounds.
- The DB-624UI column delivered sharper peaks and higher signal for organic acids (acetic, propionic, butanoic, octanoic) compared to a competitor’s 624ms phase, which exhibited severe tailing.
- Distinct headspace profiles were observed for different beer styles, underlining the method’s ability to detect both common and unique volatile markers.
Benefits and Practical Applications
This approach provides a standardized, high-throughput method for:- Routine quality control and batch consistency checks.
- Flavor stability studies during shelf life and storage.
- Product development to evaluate raw material or process changes.
- Authentication and detection of adulteration in craft and commercial beers.
Future Trends and Opportunities
Advances in headspace automation, combined with data-rich GC×GC or high-resolution MS, will enhance fingerprinting of complex matrices. Integration with chemometric and machine learning tools can accelerate untargeted screening and pattern recognition, supporting real-time process monitoring and predictive quality control in brewing operations.Conclusion
Static headspace GC/MS using the Agilent J&W DB-624 Ultra Inert column offers robust, reproducible profiling of beer volatiles, especially organic acids that challenge traditional phases. The enhanced column inertness and sensitivity support comprehensive flavor analysis across diverse beer styles.Reference
Application Note 5991-1136EN (Agilent Technologies).Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
What’s in Your Beer? GC/MS Static Head Space with a DB-624 Ultra Inert Capillary GC Column
2012|Agilent Technologies|Posters
What’s in Your Beer? GC/MS Static Head Space with a DB-624 Ultra Inert Capillary GC Column AOAC 2012 Authenticity P-T-013 Joan Stevens1, Trisa Robarge1,, and Kenneth Lynam (Agilent Technologies Inc., Lake Forest, CA1,, Wilmington DE2) Abstract Static head space GC/MS…
Key words
ethyl, ethylacetate, acetatealcohol, alcoholisoamyl, isoamylisobutyl, isobutylacid, acidoctanoic, octanoicdecanoate, decanoatealdehyde, aldehydelinolool, linoloolacetyl, acetylhexanoate, hexanoatebeer, beeramyl, amylpropyl
What’s in Your Beer? Part 2: GC/MS Static Head Space with a Highly Inert 624 Capillary GC Column
2018|Agilent Technologies|Posters
What’s in Your Beer? Part 2: GC/MS Static Head Space with a Highly Inert 624 Capillary GC Column Introduction Some ingredients in beer can be challenging to separate and chromatograph. Common beer flavor components include; fusel oils, aldehydes, esters, and…
Key words
ethyl, ethylacetate, acetatealdehyde, aldehydeisobutyl, isobutylacid, acidalcohol, alcoholacetyl, acetylbeer, beerisoamyl, isoamylethanol, ethanolbeers, beersoctanoic, octanoicamly, amlychromatogram, chromatogramacetic
Using GC/MSD with High Efficiency Source and JetClean to detect low level contaminants in ethylene and propylene
2018|Agilent Technologies|Posters
What’s in Your Beer? Part 2: GC/MS Static Head Space with a Highly Inert 624 Capillary GC Column Introduction Some ingredients in beer can be challenging to separate and chromatograph. Common beer flavor components include; fusel oils, aldehydes, esters, and…
Key words
ethyl, ethylacetate, acetatealdehyde, aldehydeisobutyl, isobutylacid, acidalcohol, alcoholacetyl, acetylbeer, beerisoamyl, isoamylethanol, ethanolbeers, beersoctanoic, octanoicamly, amlychromatogram, chromatogramacetic
Screen Beer by GC/MS Static Headspace with the Agilent J&W DB-624 Ultra Inert Capillary Column 
2012|Agilent Technologies|Applications
Screen Beer by GC/MS Static Headspace with the Agilent J&W DB-624 Ultra Inert Capillary Column Application Note Food Testing & Agriculture Author Ken Lynam Agilent Technologies, Inc. Abstract This application note highlights the utility of Agilent J&W DB-624UI columns for…
Key words
ethyl, ethylacetate, acetateacid, acidaldehyde, aldehydebeer, beeroctanoic, octanoicisoamyl, isoamylflavor, flavorisobutyl, isobutylamyl, amylfermentation, fermentationbutanoic, butanoicalcohol, alcoholpropionic, propionicactive