Round-the-Clock, On-Line, and Cryogen-Free Monitoring of Hydrocarbons from Acetylene to Trimethylbenzene in Ambient Air

Applications | 2017 | Agilent TechnologiesInstrumentation
GC, Thermal desorption
Industries
Environmental
Manufacturer
Markes

Summary

Significance of the Topic


Ground-level ozone formation in urban atmospheres is driven by volatile organic compounds (VOCs), which serve as precursors in photochemical smog formation.
Continuous, unattended monitoring of these compounds is essential for regulatory compliance, source attribution, and real-time assessment of pollution episodes and meteorological influences.

Objectives and Study Overview


This study validates a cryogen-free thermal desorption and gas chromatographic system for online, round-the-clock measurement of 30–56 VOCs ranging from acetylene to trimethylbenzenes in ambient air.
Key goals include evaluating breakthrough volumes, retention stability, calibration reliability, detection limits below 0.5 ppb, and system performance under field conditions.

Methodology


Automated sampling is conducted by pulling ambient air onto a cooled, sorbent-packed focusing trap at –30 °C, followed by rapid desorption at 325 °C into a GC system. A membrane dryer removes water and polar interferences. Retention tests use certified gas standards from the National Physical Laboratory and the US EPA. Performance metrics include breakthrough volume determination, retention time reproducibility, multipoint calibration, and a stringent F-test for calibration fidelity.

Instrumentation Used


UNITY 2 Air Server with electronic mass flow control and cryogen-free, electrically cooled sorbent trap (U-T17O3P-2S).
Gas chromatograph with two-column Deans switch configuration:
  • Dimethylpolysiloxane column (60 m × 0.25 mm) for C6+ analytes, FID1 detection.
  • Alumina PLOT column (50 m × 0.32 mm) for C2–C6 analytes, FID2 detection.


Main Results and Discussion


Breakthrough experiments showed quantitative retention of acetylene up to 1.5 L sampled volume without loss. Chromatographic separation of 30- and 56-component standards achieved distinct, well-resolved peaks.
Retention time stability was excellent (RSD < 0.01 % for C6+ and < 0.32 % for C2–C6). Calibration reproducibility yielded RSDs below 1 % for most compounds and R2 values above 0.99 in over 98 % of cases. Calculated detection limits were approximately 0.05 ppb for C2–C3 and 0.03 ppb for C4+.
A field trial in a semi-rural location demonstrated clear diurnal VOC profiles correlated with traffic patterns and continuous unattended operation for extended periods.

Benefits and Practical Applications

  • Cryogen-free, automated, round-the-clock monitoring reduces operational costs and complexity.
  • High sensitivity and low detection limits enable trace-level quantitation of ozone precursors.
  • Robust retention and reproducibility support regulatory compliance and long-term environmental studies.
  • Real-time data allow analysis of emission episodes and meteorological influences.


Future Trends and Opportunities


Integration of mass spectrometric detection for expanded compound identification.
Development of portable, networked sensor arrays for urban air quality mapping.
Advanced data analytics and predictive algorithms to forecast pollution events.
Extension to additional trace gas species and coupling with meteorological and remote sensing data.

Conclusion


The validated cryogen-free UNITY 2 Air Server system offers reliable, sensitive, and fully automated monitoring of VOC ozone precursors in ambient air, fulfilling regulatory requirements and providing valuable insights into urban air quality dynamics.

References

  • Council Directive 96/62/EC of 27 September 1996 on ambient air quality assessment and management and Directive 2000/69/EC of 16 November 2000 on limit values for benzene and carbon monoxide in ambient air. European Parliament and Council.
  • US Clean Air Act Amendments of 1990, provisions for attainment and maintenance of national ambient air quality standards. US Environmental Protection Agency.
  • AFNOR NF T90-210. Qualité de l’eau. Protocole d’évaluation initiale des performances d’une méthode dans un laboratoire. AFNOR, 2009.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Using the CIA Advantage for automated cryogen-free analysis of canister air and gas
Thermal Desorption Technical Support Note 99: Using the CIA Advantage for automated cryogen-free analysis of canister air and gas Key Words: Canisters, US EPA Method TO-15, ozone precursors, air toxics, greenhouse gases Summary The CIA Advantage is an advanced system…
Key words
cia, ciaadvantage, advantagecanister, canisterair, airtoxics, toxicsisopropanol, isopropanolsampling, samplingmfc, mfcpropene, propenecryogen, cryogenlow, lowvinyl, vinyluoromethane, uoromethanehigh, highlowconcentration
Round-the-clock, online and cryogen-free monitoring of VOC hydrocarbons in ambient air using GC-MS
APPLICATION NOTE Terry Jeffers, Suresh Seethapathy, Dwain Cardona Thermo Fisher Scientific, Austin, TX Key words EPA air quality, PAMS assessment, air sampling, ambient air monitoring, VOC emissions, thermal desorption, continuous monitoring Goal To describe the evaluation and validation of a…
Key words
ethyltoluene, ethyltolueneozone, ozoneunity, unitypams, pamskori, koritrap, trapfocusing, focusingair, aircompounds, compoundssystem, systemmonitoring, monitoringretention, retentionmarkes, markesethyne, ethynesorbent
Cryogen-free analysis of VOCs in car exhaust
Cryogen-free analysis of VOCs in car exhaust
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 73566 Cryogen-free analysis of VOCs in car exhaust Thermal desorption coupled to gas chromatography with dual flame ionization detectors Authors: David Lee1, Terry Jeffers2, Cristian Cojocariu1, and David Wevill3 Thermo Fisher Scientific, Runcorn, UK 1 Thermo Fisher Scientific,…
Key words
none, noneethyltoluene, ethyltoluenedetected, detectedppbc, ppbckori, koritrap, trapdual, dualcia, ciavocs, vocsmarkes, markesdeans, deanspams, pamspurge, purgecompound, compoundpropane
Cryogen-free analysis of VOCs in car exhaust
Cryogen-free analysis of VOCs in car exhaust
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 73566 Cryogen-free analysis of VOCs in car exhaust Thermal desorption coupled to gas chromatography with dual flame ionization detectors Authors: David Lee1, Terry Jeffers2, Cristian Cojocariu1, and David Wevill3 Thermo Fisher Scientific, Runcorn, UK 1 Thermo Fisher Scientific,…
Key words
none, noneethyltoluene, ethyltoluenedetected, detectedppbc, ppbckori, koritrap, trapcia, ciavocs, vocsdual, dualmarkes, markesdeans, deanspams, pamspurge, purgepropane, propanecompound
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike