Agilent AI Peak Integration for GC/MS Analysis of Phthalates

Applications | 2024 | Agilent TechnologiesInstrumentation
GC/MS/MS, GC/QQQ, GC/SQ, GC/Q-TOF, GC/HRMS, GC/MSD, Software
Industries
Other
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Manual integration of GC/MS chromatograms remains one of the most labor-intensive and subjective steps in analytical workflows. Phthalates are common plasticizers regulated for their health effects, but their analysis is complicated by isomeric mixtures, broad peak shapes, and matrix interferences. An AI-assisted integration strategy can address these challenges by automating baseline correction and peak detection while preserving accuracy.

Study Objectives and Overview


This application note demonstrates the implementation of Agilent AI Peak Integration within MassHunter Quantitative Analysis software for GC/MS phthalate measurements. The goals were to train a machine learning model on real-world manual integrations, validate its reproducibility, evaluate processing speed, assess accuracy across concentration ranges, and examine its ability to correct false positive and negative detections.

Methodology and Instrumentation


• Agilent GC/MSD instruments operating in synchronous SIM/Scan mode
• MassHunter Quantitative Analysis v10.2 enhanced with AI Peak Integration plugin
• Model training based on over 1,000 manually integrated phthalate samples uploaded to a cloud database
• Continuous learning enabled by optional user adjustments after AI integration

Main Results and Discussion


• Reproducibility: Five replicate integrations of diisodecyl phthalate (DIDP) at 0.4 ppm yielded identical peak areas (1050) using a fixed model version.
• Speed: AI integration processed individual samples in an average of 12 seconds, compared to 60–120 seconds for manual review and adjustment.
• Accuracy: Across 576 samples from 18 batches, the quantifier ion integration accuracy exceeded 98% relative to manual results.
• False positives/negatives: The model successfully removed spurious integrations (e.g., overlap between bis(2-ethylhexyl) phthalate and dicyclohexyl phthalate) and detected low-level peaks down to 0.1 ppm within calibration curves.

Benefits and Practical Applications


Integrating AI-assisted peak detection delivers:
  • Significant reduction in data analysis time and labor
  • Consistent results and minimized operator bias
  • Adaptive performance via continuous model refinement

Future Trends and Opportunities for Use


Advancements in AI-driven integration are expected to extend beyond phthalates into diverse compound classes and complex matrices. Integration with laboratory information management systems (LIMS) and regulatory compliance workflows may streamline quality assurance. Ongoing model development could enable real-time chromatographic decision support and improved handling of unresolved peaks.

Conclusion


Agilent AI Peak Integration for MassHunter Quantitative Analysis significantly enhances GC/MS phthalate workflows by delivering rapid, reproducible, and accurate peak integrations. The machine learning approach reduces manual workload, supports continuous improvement, and ensures consistent quantification across diverse sample sets.

References


  1. Net S, et al. Occurrence, fate, behavior and ecotoxicological state of phthalates in different environmental matrices. Environmental Science & Technology. 2015;49(7):4019–4035. doi:10.1021/es505233b
  2. Dong C, et al. Detecting phthalate esters in sludge particulates from wastewater treatment plants. Journal of Environmental Science and Health, Part A. 2020;55(10):1233–1240. doi:10.1080/10934529.2020.1780850
  3. Sosnovshchenko O, Luo R, King T. Agilent AI Peak Integration for MassHunter. Agilent Technologies Technical Overview, Publication 5994-6728EN, 2023.
  4. Bushey J. Phthalate Analysis Using an Agilent 8890 GC and an Agilent 5977A GC/MSD. Agilent Technologies Application Note, Publication 5994-0483EN, 2018.
  5. Andrianova A, Quimby B. Rapid Screening Workflow for Phthalates in Plastics by GC/MSD in Under Six Minutes. Agilent Technologies Application Note, Publication 5994-2727EN, 2020.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
AI Peak Integration for MassHunter software automates manual peak integration during the data analysis process in GC/SQ
Poster Reprint ASMS 2023 Poster number TP 062 AI Peak Integration for MassHunter software automates manual peak integration during the data analysis process in GC/SQ Thomas Bispham, Tamas King, Winnie Chau, and Ruoji Luo Agilent Technologies, Inc. Introduction GC/MS data…
Key words
integration, integrationlearning, learningmanual, manualmachine, machinepeak, peakbatch, batchphthalate, phthalatemodel, modeldinp, dinpdata, datachemist, chemistadd, addautomates, automatescompounds, compoundstrained
Agilent AI Peak Integration for MassHunter
Agilent AI Peak Integration for MassHunter
2023|Agilent Technologies|Technical notes
Technical Overview Agilent AI Peak Integration for MassHunter A tool to optimize GC-MS Phthalate Test Quantitative Analysis Authors Oleksandr Sosnovshchenko, Ruoji Luo, and Tamas King, Agilent Technologies, Inc. Abstract Agilent AI Peak Integration for MassHunter is a software tool designed…
Key words
integration, integrationlearning, learningmodel, modelmachine, machinepeak, peakmasshunter, masshuntertraining, trainingmanual, manualaccuracy, accuracypractitioners, practitionerscorrectness, correctnessmodels, modelsintegrator, integratoragilent, agilentprocess
Phthalate Analysis Using an Agilent 8890 GC and an Agilent 5977A GC/MSD
Application Note Consumer Products Phthalate Analysis Using an Agilent 8890 GC and an Agilent 5977A GC/MSD Author Jared Bushey Agilent Technologies, Inc. Wilmington, DE, USA Abstract Mixtures of seven phthalates were analyzed using the Agilent 8890 GC coupled to an…
Key words
phthalate, phthalatedetected, detectedphthalates, phthalatesdidp, didpdibp, dibpdehp, dehpdinp, dinpdnop, dnopdbp, dbpbbp, bbppen, penacquisition, acquisitionwere, werecounts, countslevel
Analysis of Phthalates Using the Agilent 5977C GC/MSD with Hydrogen Carrier Gas
Application Note Materials Testing & Research Analysis of Phthalates Using the Agilent 5977C GC/MSD with Hydrogen Carrier Gas Author Bruce D. Quimby, Ph.D. Agilent Technologies, Inc. Abstract An Agilent 8890 GC system coupled with an Agilent 5977C Inert Plus GC/MSD…
Key words
phthalate, phthalatelinear, linearhydrogen, hydrogendphp, dphpdibp, dibpcarrier, carrierbmpp, bmppdhxp, dhxpdmep, dmepdep, depdbep, dbepdehp, dehpsource, sourcednp, dnpdbp
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike