Derivatization Procedure and Negative Chemical Ionization GC/MS/MS Conditions for the Analysis of Steroidal Analogs

Applications | 2012 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Clinical Research
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


The accurate measurement of steroidal analogs such as estrogens and androgens is essential in clinical research, forensic toxicology and environmental monitoring. These compounds serve as biomarkers for diseases (for example, estrogen dysregulation in breast cancer and hypertension) and as emerging contaminants affecting ecosystems. A robust, sensitive analytical method enables reliable detection at trace levels in complex matrices.

Objectives and Overview of the Study


This application note presents a two-step chemical derivatization protocol coupled with negative chemical ionization gas chromatography tandem mass spectrometry (NCI-GC/MS/MS). It focuses on optimizing derivatization reagents and conditions for estrogens and androgens to achieve picogram-level detection limits and high selectivity in biological and environmental samples.

Methodology and Derivatization Procedure


The workflow transforms phenolic and alcoholic groups into pentafluorobenzoyl esters and ketones into pentafluorobenzyl oximes. Key steps include:
  • Esterification: Add 1.0% pentafluorobenzoyl chloride in anhydrous ethyl acetate and 100 µL pyridine to the dried extract. Incubate at 60 °C for 30 minutes. Quench with sodium bicarbonate, centrifuge, transfer organic layer.
  • Oximation: Evaporate, then add 0.1% pentafluorobenzyl hydroxylamine hydrochloride in pyridine. Incubate at 60 °C for 30 minutes, evaporate to dryness.
  • Reconstitution: Redissolve in 50 µL dodecane or isooctane prior to analysis.

The protocol uses polar, aprotic solvents and tertiary amine bases to maximize conversion and suppress hydrolysis of acid chlorides.

Instrumentation Used


  • Gas Chromatograph: Agilent 7890A with dual Agilent J&W DB-17ht columns (1 m × 0.18 mm, 0.1 µm and 15 m × 0.18 mm, 0.1 µm) joined by a Purged Ultimate Union for timed backflush.
  • Mass Spectrometer: Agilent 7000 Triple Quadrupole in NCI mode using methane reagent gas.
  • Injection: Pulsed splitless (40 psi, 0.9 min), 2 µL volume, inlet at 280 °C.
  • Carrier Gas: Helium in constant flow (1.0 mL/min on precolumn then 99 mL/min backflush; 1.2 mL/min on analytical column).
  • Oven Program: 205 °C (1 min), ramp 10 °C/min to 295 °C, then 2.5 °C/min to 305 °C.
  • MS Parameters: Source and quadrupoles at 150 °C; selected reaction monitoring transitions optimized for each steroid; collision gas nitrogen (1.5 mL/min).

Main Results and Discussion


Estradiol (E2) was used to validate sensitivity. Eight replicates of a 10 pg/mL standard yielded:
  • Average signal-to-noise ratio of 441 (RMS algorithm over 0.2 min noise).
  • Relative standard deviation of 3.7 % for area counts.
  • Instrument detection limit (IDL) calculated at approximately 1.1 pg/mL (99 % confidence).

A calibration curve from 0.5 to 250 pg/mL showed linearity (R² = 0.998). Retention times for derivatized analytes ranged from 8.96 to 11.98 minutes under the optimized GC program.

Benefits and Practical Applications


This method delivers:
  • Ultrahigh sensitivity at low-pg/mL levels.
  • Selective electron capture via pentafluorobenzoyl modification.
  • Efficient removal of matrix interferences through backflush.
  • Extended column lifetime and stable background.
  • Versatility across sample types: serum, plasma, wastewater, soil and biosolids.

Future Trends and Opportunities


Potential developments include:
  • Automation of derivatization for high-throughput clinical assays.
  • Expansion to broader panels of endogenous and synthetic steroids.
  • Integration with alternative ionization techniques to compare sensitivity and selectivity.
  • Application in large-scale environmental monitoring of endocrine disruptors.
  • Use of novel fluorinated reagents for faster or more stable derivatives.

Conclusion


The described two-step derivatization combined with NCI-GC/MS/MS achieves highly sensitive and reproducible analysis of steroidal analogs. With pg/mL detection capability and robust chromatography, it supports diverse research needs from clinical biomarker quantification to environmental surveillance.

References


  1. Churley M., Macherone A., White R. Analysis of estrone and 17β-estradiol in ground water by GC-NCI-MS/MS. 59th ASMS Conference, Denver, CO, June 5–8, 2011.
  2. Macherone A., Churley M., White R. Ultra-low Detection of Estrogenic Compounds by GC-NCI/MS/MS. LC/GC Special Issues, Advanstar Communications, 2010.
  3. Wells G., Prest H. Why use signal-to-noise as a measure of MS performance when it’s often meaningless. Agilent Application Note 5990-8341EN, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Approaching The Ultimate Limits of Detection For Endocrine Disruptors in Wastewater Effluent Using GC-NCI-MS/MS
Approaching The Ultimate Limits of Detection For Endocrine Disruptors in Wastewater Effluent Using GC-NCI-MS/MS Melissa Churley1, Ruth-Marfil-Vega2, Marc Mills3 and Anthony Macherone4,5 1Agilent Technologies, Inc., Santa Clara, CA; 2 American Water, Belleville, IL; 3 National Risk Management Research Laboratory (NRMRL),…
Key words
steroidal, steroidaltestosterone, testosteroneestrogens, estrogensanalogs, analogsnci, nciendocrine, endocrinedev, devamenable, amenableequlilin, equlilininection, inectionwastewater, wastewatergcnci, gcnciagency, agencymarfil, marfilpentafluorobenzoyl
Analysis of Estrone and 17β-Estradiol in Ground Water by GC-NCI-MS/MS
Analysis of Estrone and 17β 17β-Estradiol in Ground Water by GCGC-NCI NCI--MS/MS ASMS 2011 WP 316 2Robert Melissa M li Ch l 1Anthony Churley, A th M h Macherone, R b t Whit White 1 Agilent Technologies, Technologies Inc Inc.…
Key words
andd, anddlev, levnci, ncispiked, spikedoff, offdht, dhtmpa, mpanor, norestrogens, estrogensstudyy, studyynon, nonpfb, pfbspike, spikeect, ecteffluent
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum Technical Overview Clinical Research Authors Introduction Anthony Macherone, Ph.D. Estrone (E1) and estradiol (E2) are the two primary estrogens in females and males. Agilent Technologies, Inc.…
Key words
wide, widepfbcl, pfbcltriple, triplemrm, mrmderivatization, derivatizationpyridine, pyridineprocedure, procedurefalse, falsecentrivap, centrivaplabconco, labconcoquadrupole, quadrupoletrue, truemsd, msdcollision, collisionestrone
Androgenic and Estrogenic Activity in Water Bodies Receiving Cattle Feedlot Effluent in Eastern Nebraska, USA
Research | Article Androgenic and Estrogenic Activity in Water Bodies Receiving Cattle Feedlot Effluent in Eastern Nebraska, USA Ana M. Soto,1 Janine M. Calabro,1 Nancy V. Prechtl,1 Alice Y. Yau,2 Edward F. Orlando,3 Andreas Daxenberger,4 Alan S. Kolok,5 Louis J.…
Key words
estrogenic, estrogenicactivity, activityfeedlot, feedlotandrogenic, androgenicandrogen, androgenanabolic, anaboliccattle, cattleeia, eiaestrogen, estrogenrunoff, runoffwere, werecafos, cafoscell, cellperspectives, perspectiveswater
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike