Analysis of Lower Aliphatic Aldehyde

Applications | 2022 | ShimadzuInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Shimadzu

Summary

Importance of the Topic


Aldehydes are common airborne pollutants and process byproducts with health and environmental impacts. Accurate detection of lower aliphatic aldehydes such as acetaldehyde, propionaldehyde and butyraldehyde is essential for air quality monitoring, occupational safety and industrial process control. Derivatization with 2,4-dinitrophenylhydrazine (DNPH) enhances stability and detectability of these reactive compounds, enabling reliable quantitation at trace levels.

Study Objectives and Overview


This application note demonstrates a method for analysis of twelve DNPH derivatives of lower aliphatic aldehydes using gas chromatography coupled with a flame thermionic detector (GC-FTD). Key aims include establishing chromatographic separation, evaluating instrument parameters and achieving sensitive quantification at microgram-per-milliliter levels.

Methodology and Instrumentation


The method employs DNPH-derivatized standards prepared in ethyl acetate (1 µg/mL). A Shimadzu Nexis GC-2030 system equipped with an AOC-20i autosampler performs splitless injection (1 mL, 1 min sampling then 1:30 split). Chromatographic separation uses an SH-5 capillary column (30 m × 0.25 mm ID, 0.25 µm film) with a temperature program from 80 °C (1 min) ramped to 200 °C at 20 °C/min, held 10 min, then 250 °C at 5 °C/min for a total run of 27 min. Helium is the carrier gas under constant linear velocity control (41.7 cm/s). Detection relies on a flame thermionic detector (FTD-2030) at 280 °C with hydrogen (1.5 mL/min), air (145 mL/min) and helium makeup (27.5 mL/min).

Main Results and Discussion


The method achieved baseline separation of twelve DNPH derivatives, including two positional isomers for each aldehyde. Retention order reflected carbon chain length and branching, with acetaldehyde derivatives eluting first and n-valeraldehyde derivatives last. Flame thermionic detection provided selective and sensitive response to nitrogen-containing derivatives. Calibration curves demonstrated linearity across the tested range, with limits of detection in the low nanogram range. Chromatograms showed sharp peak shapes and minimal carryover, confirming robustness of the splitless injection and column conditioning.

Benefits and Practical Applications

  • High sensitivity: enables trace-level quantification for environmental and workplace monitoring.
  • Selective detection: FTD targets nitrogen in DNPH adducts, reducing matrix interferences.
  • Efficient workflow: automated sampling and temperature programming support high throughput.
  • Versatility: applicable to various matrices after DNPH derivatization.

Future Trends and Applications


Emerging approaches may integrate tandem mass spectrometry for even greater specificity and lower detection limits. Miniaturized sampling devices coupled with on-site DNPH cartridges and portable GC-FTD instruments could facilitate real-time monitoring in field environments. Advances in column technology and derivatization chemistries may expand the method to a broader range of volatile carbonyls.

Conclusion


The described GC-FTD method provides a reliable, sensitive and selective approach for analysis of lower aliphatic aldehyde-DNPH derivatives. Its straightforward sample preparation, robust chromatographic separation and targeted detection make it suitable for routine air quality assessments and industrial hygiene studies.

Instrumentation Used

  • Gas Chromatograph: Shimadzu Nexis GC-2030
  • Autosampler: Shimadzu AOC-20i
  • Column: SH-5 (30 m × 0.25 mm I.D., 0.25 µm)
  • Detector: FTD-2030 (Flame Thermionic Detector)
  • Software: LabSolutions LC/GC

Reference

  1. Shimadzu Application News G292 (JP, ENG), “Analysis of Lower Aliphatic Aldehydes” First Edition, Sep. 2022.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Lower Aliphatic Aldehydes using Nexis GC-2030
LAAN-A-GC-E057 Application News No. G292 Gas Chromatograph Analysis of Lower Aliphatic Aldehydes using Nexis GC-2030 Lower aliphatic aldehydes are known to be associated with sick building syndrome. Thus various regulations require strict control of their concentrations. A common method of…
Key words
aldehydes, aldehydesaliphatic, aliphaticgas, gasschematic, schematiccollection, collectionflowrate, flowratediagram, diagrammethod, methoddnphs, dnphscartridges, cartridgeslower, lowerflame, flamecartridge, cartridgethermionic, thermionicnews
Analyzing Organic Compound Species with Hydrogen Carrier Gas
ERAS-1000-0317 GC FID AOC SH Series SH-PolarWax Analyzing Organic Compound Species with Hydrogen Carrier Gas 317 Keywords: Organic Compounds, Aldehydes, Alcohols, Ketones, Esters, Hydrogen Carrier Gas 1. Acetaldehyde 2. Acetone 3. Ethyl acetate 4. Methanol 5. n-Propanol 6. i-Butanol 7.…
Key words
gas, gascarrier, carrierhydrogen, hydrogenfid, fidpolarwax, polarwaxaoc, aocsplit, splitmakeup, makeupdetector, detectormode, modeinjection, injectionspecies, speciesmodel, modelanalyzing, analyzingseries
Analysis of Thiophene in Benzene
Analysis of Thiophene in Benzene
2022|Shimadzu|Applications
ERAS-1000-0311 GC FPD(S) AOC SH Series SH-PolarWax Analysis of Thiophene in Benzene 311 Keywords: Impurity, Sulfur compound, Benzene 1.Thiophene 0.1 ppm Model Software Injection Unit Injection Volume Injection Temperature Column Column Temperature Purge Gas Carrier Gas Carrier Gas Control Detector…
Key words
wbi, wbiinjection, injectionlabsolutionstm, labsolutionstmgas, gaspolarwax, polarwaxaoc, aocunit, unitdetector, detectorthiophene, thiophenetemperature, temperaturefpd, fpdcarrier, carrierbenzene, benzenepurge, purgemodel
Shimadzu Capillary Gas Chromatograph GC-2010 Pro
Shimadzu Capillary Gas Chromatograph GC-2010 Pro
2020|Shimadzu|Brochures and specifications
C184-E049 Capillary Gas Chromatograph GC-2010 Pro GC-2010 Pro Shimadzu Gas Chromatograph The GC-2010 Pro represents a new generation in Routine GC. Fast oven heating and cooling reduces the analysis time and allow high sample throughput. Based on GC-2010 Plus™ technology,…
Key words
detector, detectorlabsolutions, labsolutionsanalysis, analysisgas, gassystem, systemfpd, fpdfir, firchromatograph, chromatographcapillary, capillaryhigh, highinjector, injectortechnology, technologyinjection, injectioncooling, coolingdual
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike