Reduced carryover and increased reproducibility in the analysis of cold‑pressed pink grapefruit essential oil

Applications | 2018 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The analysis of cold-pressed grapefruit essential oil is important in food, fragrance and quality control due to its complex mixture of high-molecular-weight aromatic compounds. Conventional polyethylene glycol (WAX) columns often struggle with thermal stability and carryover when analyzing late-eluting, higher boiling constituents, compromising reproducibility and accuracy.

Study Objectives and Overview


This application note evaluates the performance of Agilent J&W DB-HeavyWAX GC column for cold-pressed pink grapefruit essential oil, focusing on reducing sample carryover, improving reproducibility, and maintaining resolution over extended high-temperature operation. Results are compared with traditional DB-WAX and competitor WAX phases.

Methodology and Instrumentation


Sample Preparation:
  • Cold-pressed pink grapefruit essential oil injected neat.

Gas Chromatography Conditions:
  • Columns: Agilent J&W DB-HeavyWAX (30 m × 0.25 mm, 0.25 µm), DB-WAX and competitor WAX (identical dimensions).
  • Carrier Gas: Helium at 1 mL/min constant flow.
  • Oven Programs: 60 °C hold 2 min, ramp 5 °C/min to final hold at 250 °C or 280 °C for 30 min.
  • Inlet: Split 200:1, 250 °C, Ultra Inert liner.
  • Detector: FID at 280 °C with standard H₂, air and makeup flows.

Instrumentation:
  • Agilent 7890B GC with FID;
  • Agilent 7693 autosampler;
  • Mass spectrometer used for compound identification.

Main Results and Discussion


Final Oven Temperature and Carryover:
  • At 250 °C final hold, heavy coumarins Meranzin and Isomeranzin failed to elute, resulting in carryover into subsequent runs.
  • Raising final temperature to 280 °C enabled sharp peak shapes and full elution of late-eluting compounds.

High-Molecular-Weight Compound Profiling:
  • Five high-molecular-weight aromatics (Meranzin, Isomeranzin, Osthole, Bergapten, Auraptenol) were resolved only at the elevated final temperature.

Thermal Stability and Resolution Over Time:
  • Traditional WAX column run at 280 °C for 50 h and 100 h showed retention time shifts, co-elution of linalool and γ-amorphene, and altered elution order, indicating phase degradation.
  • DB-HeavyWAX maintained consistent retention, resolution and elution order for four monitored peaks even after 100 h at 280 °C.
  • Selectivity of DB-HeavyWAX closely matched that of DB-WAX, facilitating straightforward method translation.

Benefits and Practical Applications


The DB-HeavyWAX column offers:
  • Reduced carryover and consistent peak shapes for high-boiling essential oil constituents.
  • Enhanced reproducibility across extended sequences of analyses.
  • Extended operating range (280 °C isothermal, 290 °C programmed) without compromising selectivity.

This makes it well suited for routine quality control of citrus and other complex essential oils in food, fragrance and agricultural testing.

Future Trends and Potential Applications


Emerging directions include:
  • Application to other cold-pressed oils with heavy aromatics (e.g., bergamot, neroli).
  • Coupling high-temperature WAX phases with mass spectrometry for trace-level quantitation.
  • Standardization of extended-temperature GC methods in QA/QC workflows.

Conclusion


The Agilent J&W DB-HeavyWAX column demonstrates superior thermal stability and consistent selectivity for cold-pressed pink grapefruit essential oil analysis. Operating at up to 280 °C eliminates carryover of late-eluting heavy compounds, ensures reproducible retention times over long runs, and maintains elution order, outperforming traditional WAX phases while requiring minimal method adjustments.

Reference


  1. Naha Aromatherapy. How are essential oils extracted? Available online. Accessed January 2018.
  2. Lin J., Rouseff R.L. Characterization of aroma-impact compounds in cold-pressed grapefruit oil using time–intensity GC–olfactometry and GC–MS. Flavor Fragrance J. 2001;16:457–463.
  3. Abercrombie V., Provoost L. Increased Thermal Stability and Maximum Temperature of the Agilent J&W DB-HeavyWAX Column. Agilent Technologies Application Note 5991-9035EN, 2018.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
WAX Column Technology Update  and Recent Applications
WAX Column Technology Update and Recent Applications
2021|Agilent Technologies|Presentations
WAX Column Technology Update and Recent Applications Gustavo Serrano Izaguirre, Ph.D., Global Product Manager – GC Columns 1 March 29, 2021 DE44274.5728240741 WAX Columns • Also referred to as Polyethylene Glycol (PEG) columns • Used to analyze polar compounds •…
Key words
wax, waxconfidentiality, confidentialitytitle, titlelabel, labelheavywax, heavywaxfatwax, fatwaxacid, acidinert, inertacids, acidsultra, ultracompetitor, competitorinertness, inertnessdha, dhacolumn, columnxylene
Now There’s A WAX Column That Can Take The Heat
Now There’s A WAX Column That Can Take The Heat Introducing: Agilent J&W DB-HeavyWAX GC columns Extend operating temperature limits while maintaining WAX selectivity Proof of thermal stability: Standard WAX column vs. Agilent J&W DB-HeavyWAX column WAX GC columns can…
Key words
heavywax, heavywaxwax, waxagilent, agilentcolumn, columncolumns, columnsmaximum, maximumbleed, bleedtemperature, temperatureindustrial, industrialisomeranzin, isomeranzinmeranzin, meranzinosthole, ostholeoperating, operatingtemperatures, temperaturesincreased
WAX Column Technology Update
WAX Column Technology Update
2018|Agilent Technologies|Presentations
WAX Column Technology Update Mark Sinnott August 14, 2018 1 August 14, 2018 Agilent J&W DB-HeavyWAX WAX Column Update • What is a WAX column? • Traditional WAX Phases • • • • • • 2 DB-WAX HP-INNOWAX VF-WAX MS…
Key words
heavywax, heavywaxwax, waxfatwax, fatwaxagilent, agilentcolumn, columninert, inertacids, acidscompetitor, competitordha, dhaacid, acidfame, famebleed, bleedffap, ffapultra, ultratraditional
Selecting the Right WAX Column
Selecting the Right WAX Column
2018|Agilent Technologies|Posters
Selecting the Right WAX Column Agilent J&W WAX GC columns deliver excellent inertness with greater sensitivity than competitive WAX columns. With the increasing number of available Agilent J&W WAX columns, it can be challenging deciding which one is best suited…
Key words
wax, waxheavywax, heavywaxinert, inertacid, acidultra, ultrafatwax, fatwaxyou, youpropionic, propionicyes, yescompetitor, competitoranalytes, analytesfames, fameswhat, whatacids, acidsdentical
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike