Agilent Reversed Flow Modulator
Technical notes | 2018 | Agilent TechnologiesInstrumentation
Two-dimensional gas chromatography (GC×GC) delivers exceptional resolution for complex mixtures encountered in petrochemical, environmental, fragrance and metabolomic analyses. The Reversed Flow Modulator (RFM) based on Agilent capillary flow technology optimizes transfer between columns while maintaining chromatographic integrity and reducing operational costs.
This technical overview presents the design and performance of Agilent’s second-generation reversed injection flow modulator. It aims to demonstrate how precise flow control and inert construction enhance repeatability and sensitivity in GC×GC separations without requiring cryogenic cooling.
GC×GC separation is achieved by coupling a nonpolar first column (boiling-point based separation) with a shorter polar second column (polarity based separation). The RFM employs a three-way solenoid valve synchronized with a pressure control module (PCM) to alternate load and inject states. A proprietary inert coating on the flow paths ensures zero carryover and negligible thermal lag. Key aspects:
The system configuration includes:
Performance evaluation demonstrated:
Illustrations of flow states (load vs. inject) confirm precise timing and minimal band broadening during transfer.
The RFM offers significant advantages:
Advancements may include miniaturized modulators for portable GC×GC, integration with high-resolution mass spectrometry, and application of machine learning algorithms for automated peak deconvolution. Emerging fields such as real‐time process monitoring and advanced petrochemical fingerprinting are set to benefit.
Agilent’s reversed flow modulator delivers reliable, high‐performance GC×GC separations with simplified operation and lower costs. Its inert design and precise flow control preserve chromatographic integrity and expand analytical capabilities across diverse industries.
GC, GCxGC
IndustriesManufacturerAgilent Technologies
Summary
Importance of the topic
Two-dimensional gas chromatography (GC×GC) delivers exceptional resolution for complex mixtures encountered in petrochemical, environmental, fragrance and metabolomic analyses. The Reversed Flow Modulator (RFM) based on Agilent capillary flow technology optimizes transfer between columns while maintaining chromatographic integrity and reducing operational costs.
Study objectives and overview
This technical overview presents the design and performance of Agilent’s second-generation reversed injection flow modulator. It aims to demonstrate how precise flow control and inert construction enhance repeatability and sensitivity in GC×GC separations without requiring cryogenic cooling.
Methodology and Instrumentation
GC×GC separation is achieved by coupling a nonpolar first column (boiling-point based separation) with a shorter polar second column (polarity based separation). The RFM employs a three-way solenoid valve synchronized with a pressure control module (PCM) to alternate load and inject states. A proprietary inert coating on the flow paths ensures zero carryover and negligible thermal lag. Key aspects:
- No cryogenics required due to low thermal mass
- Precisely timed valve switching at defined carrier gas flow rates
- Preservation of first-dimension resolution during transfer
Instrumentation
The system configuration includes:
- Agilent 7890B GC with firmware version A.04.07 or higher
- G4573A reversed inject flow modulator kit
- Pressure control module (PCM) for carrier gas delivery
- Flame ionization detector (FID) capable of 200 Hz data acquisition
- Split/splitless inlet or multi-mode inlet (MMI)
- High-purity helium or hydrogen carrier gas
- Two-dimensional data processing software
Key results and discussion
Performance evaluation demonstrated:
- High repeatability of modulation cycles under varying temperature programs
- Complete fractionation of effluent into the second column for enhanced peak capacity
- Absence of chromatographic artifacts or sample discrimination
- Robust operation over extended sequences without maintenance
Illustrations of flow states (load vs. inject) confirm precise timing and minimal band broadening during transfer.
Benefits and practical applications
The RFM offers significant advantages:
- Reduced operational costs by eliminating cryogens
- Enhanced detection of coeluting compounds in fuel profiling, polymer analysis and flavor studies
- Capability to analyze a wide volatility range without modification
- Seamless integration into existing GC×GC workflows for environmental screening and metabolomics
Future trends and potential applications
Advancements may include miniaturized modulators for portable GC×GC, integration with high-resolution mass spectrometry, and application of machine learning algorithms for automated peak deconvolution. Emerging fields such as real‐time process monitoring and advanced petrochemical fingerprinting are set to benefit.
Conclusion
Agilent’s reversed flow modulator delivers reliable, high‐performance GC×GC separations with simplified operation and lower costs. Its inert design and precise flow control preserve chromatographic integrity and expand analytical capabilities across diverse industries.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Comprehensive GC System Based on Flow Modulation for the 7890A GC
2008|Agilent Technologies|Technical notes
Comprehensive GC System Based on Flow Modulation for the 7890A GC Application Brief Introduction A hardware solution is available on the 7890A for the practice of comprehensive GC. The system uses a capillary flow modulator controlled by the 7890A GC.…
Key words
modulator, modulatormodulation, modulationflow, flowsecond, secondpcm, pcmvalve, valvetechnology, technologychannel, channelcapillary, capillaryaccessory, accessorycollection, collectionband, bandcryogen, cryogencollect, collectsolenoid
Capillary Flow Technology: GCxGC Flow Modulator
2013|Agilent Technologies|Brochures and specifications
Capillary Flow Technology: GCxGC Flow Modulator GET A SECOND DIMENSION OF INFORMATION ON COMPLEX MIXTURES Maximize information while collecting and analyzing GC data Comprehensive two-dimensional GC, or GCxGC, is a powerful technique that can be used to separate very complex…
Key words
modulator, modulatorflow, flowgcxgc, gcxgcsecond, secondtechnology, technologycapillary, capillarychannel, channelaromatics, aromaticsmodulation, modulationagilent, agilentcollection, collectionfeedstock, feedstockpcm, pcminlet, inletmixtures
Hydrocarbon Class Analysis of Conventional and Synthetic Aviation Turbine Fuels by ASTM D8396
2024|Agilent Technologies|Applications
Application Note Energy and Chemicals Hydrocarbon Class Analysis of Conventional and Synthetic Aviation Turbine Fuels by ASTM D8396 Flow-modulated GCxGC-FID using the Agilent Reverse Flow Modulator and 8890 GC Author Abstract Scott Hoy Agilent Technologies, Inc. This application note demonstrates…
Key words
naphthenes, naphthenesparaffins, paraffinsdimension, dimensionrfm, rfmretention, retentiontotal, totalisoparaffins, isoparaffinsgcxgc, gcxgchydrogen, hydrogensec, secmodulation, modulationtemplate, templatecyclohexane, cyclohexanefuel, fuelmolecular
Analysis of Essential Oils Using Comprehensive Two-Dimensional Gas Chromatography (GC × GC) and High-Resolution Mass Spectrometer
2025|Agilent Technologies|Applications
Application Note Food Testing and Agriculture Analysis of Essential Oils Using Comprehensive Two-Dimensional Gas Chromatography (GC × GC) and High‑Resolution Mass Spectrometer Authors Nick Harden, Sofia Nieto, Scott Hoy, Kai Chen, and Matthew Curtis Agilent Technologies, Inc. Abstract The analysis…
Key words
fid, fidrfm, rfmrestrictor, restrictoroils, oilsmodulation, modulationessential, essentialmodulator, modulatorunpurged, unpurgedsplitter, splitterspliter, spliterpurged, purgedcurcumene, curcumenetrans, transcedrene, cedrenecis