240-MS GC/MS Ion Trap Mass Spectrometer

Manuals | 2008 | Agilent TechnologiesInstrumentation
GC/MSD, GC/IT
Industries
Manufacturer
Agilent Technologies

Summary

Significance of Topic


The Varian 240-MS GC/MS Ion Trap Mass Spectrometer is a versatile analytical tool for qualitative and quantitative analysis across pharmaceutical, environmental, and industrial laboratories. Proper understanding of its operation, maintenance, and safety procedures is essential to ensure reliable performance, accurate results, and long equipment life.

System Overview


The 240-MS integrates a gas chromatograph (GC) and an ion trap mass spectrometer (MS) within a controlled vacuum manifold. Sample components separate in the GC column, flow through a heated transfer line, and enter the MS for ionization (internal, external, or hybrid chemical ionization) and mass analysis. Ions are trapped, ejected in order of mass-to-charge ratio, and detected by a conversion dynode/electron multiplier.

Instrumentation and Methodology


  • Vacuum System: A turbo pump (230 L/s) backed by a foreline pump (DS-102 or dry scroll) maintains 10µTorr in the manifold.
  • Pneumatics: Helium damping gas, calibration gas (FC-43), and CI reagent gases are delivered through electronic flow controllers and needle valves.
  • Electronics: A controller board, RF generator/coil assembly, manifold electronics, and ion detection circuitry coordinate scan functions, RF trapping fields, supplemental waveforms, and data acquisition.
  • Ionization Modes: Internal electron ionization (EI), external EI, positive/negative chemical ionization (PCI/NCI), and hybrid CI allow selection of optimal ionization for analytes.
  • Safety: Comprehensive guidelines cover electrical, chemical, fire, explosion, and radiation hazards, plus GC and LC high-pressure and UV safety.

Maintenance and Cleaning Procedures


Regular pump oil changes (rotary vane or dry scroll), tip seal replacements, and filter checks sustain vacuum integrity. MS maintenance includes source cleaning (removing carbon deposits), ion trap cleaning (silica-coated electrodes, quartz spacers), and component replacement (filaments, electron multiplier, CI inlet). Detailed procedures for venting, bake-out, and reassembly ensure minimal downtime.

Troubleshooting


Systematic checks isolate issues in the data system, GC, or MS. Diagnostics mode monitors turbomolecular pump speed, ion gauge, air/water background, and mass spectrum quality. Leak detection with gas sprays, spectrum-based troubleshooting, and column test mixtures help resolve air leaks, water vapor problems, mass calibration drift, and chromatographic artifacts.

Benefits and Practical Applications


  • Enhanced uptime through preventive maintenance schedules.
  • Accurate mass spectral data by optimized ionization and vacuum conditions.
  • Flexible ionization modes for a wide range of compounds.
  • Rapid fault isolation via integrated diagnostics and synchronization options for automated workflows.

Future Trends and Potential Applications


Integration with automated sample handling, remote diagnostics, and advanced data processing (machine learning-based library searches) will streamline analyses. Micro-pump technology and novel ion trap designs may further reduce footprint and enhance sensitivity. Expanded hybrid ionization chemistries can improve the characterization of labile or high-polarity compounds.

Conclusion


The Varian 240-MS GC/MS provides robust, flexible analytical capabilities. Adherence to detailed safety, operation, and maintenance procedures maximizes instrument performance and ensures consistent, accurate results across diverse applications.

Reference


No external bibliographic references were cited in the original manual.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
MiniVac Controller
MiniVac Controller
2009|Agilent Technologies|Manuals
NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to www.agilent.com/chem. vacuum technologies MiniVac Controller Model 929-0196 Model 929-0197 87-900-053-01 (F) MARCH 2009 (I) MANUALE DI ISTRUZIONI…
Key words
controller, controllerminivac, minivacder, dervarian, variandes, descontrôleur, contrôleurles, lespour, pourvoltage, voltagedie, dievacion, vacionpump, pumpdel, delcontrollers, controllersvacuum
Restek Electronic Leak Detector
Instruction Manual Restek Electronic Leak Detector (Cat.# 28500) www.restek.com Table of Contents 1.0 Introduction.............................................................................................................................................. 3 2.0 Battery Charging..................................................................................................................................... 4 3.0 Powering Up............................................................................................................................................. 4 4.0 Zeroing the Unit...................................................................................................................................... 4 5.0 Prior to Operation................................................................................................................................... 5 6.0 Detecting Leaks....................................................................................................................................... 5 7.0 Specifications............................................................................................................................................ 6…
Key words
marche, marchebatterie, batterieled, ledl’unità, l’unitàoder, oderein, eindes, desdella, dellasonde, sondeund, undgerät, gerätdétecteur, détecteurrestek, restekaccesa, accesasonda
TurboVap (Safety Translations)
TurboVap (Safety Translations)
2019|Biotage|Manuals
TurboVap ® Safety Translations Contents TurboVap ® Safety Translations CONTENTS 1 3 6 9 12 15 2 安全 安全性について Sécurité Seguridad Sicherheitshinweise Sicurezza TurboVap Safety Translations | © Biotage 2019 ® 安全 安全 注意:这是英语版“TurboVap Installation and Safety”文档中 “Safety”一章的翻译。如有任何出入,以英语原版为准。 ® 使用目的…
Key words
システム, システムbiotage, biotagesistema, sistemadel, deldes, desund, undsystème, systèmeles, lesder, derturbovap, turbovapdie, diesie, sierohs, rohslas, lascon
Shimadzu GC Columns Installation Guide
ERAD-0001-9079 Columns Installation Guide 气相色谱柱安装指南 Instructions pour l’installation d’une colonne capillaire Einbau einer GC-Kapillarsäule Istruzioni per l’installazione di una colonna capillare Instalación de Columnas Capilares www.shimadzu.com/an 2 ERAD-0001-9079 GC Column Installation Instructions The Shimadzu Innovations and Technical Services specialists have…
Key words
カラム, カラムcolonna, colonnacolonne, colonnedella, dellacolumna, columnader, derpour, pourpara, paravecteur, vecteuruna, unatempérature, températurecapillaire, capillairel’installazione, l’installazioneauf, aufque
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike