Routine-grade quantitative performance of TriPlus 500 Headspace Autosampler coupled to TRACE 1310 GC-FID
Applications | 2018 | Thermo Fisher ScientificInstrumentation
Testing and controlling residual solvents in pharmaceutical products is critical to ensure patient safety and compliance with regulatory standards. Residual solvents may pose health risks, and international guidelines such as ICH Q3C and USP <467> define acceptable limits and analytical procedures. Headspace gas chromatography coupled with flame ionization detection is a preferred approach due to its efficiency in analyzing volatile organic compounds with minimal sample preparation.
This study evaluates the quantitative performance of a Thermo Scientific TriPlus 500 Headspace Autosampler combined with a TRACE 1310 GC equipped with an Instant Connect FID for routine analysis of Class 1, Class 2A, and Class 2B residual solvents according to USP <467> procedure C. The aim is to assess linearity, sensitivity (LOD/LOQ), and repeatability under standard operating conditions.
Sample and standard solutions of residual solvents were prepared in dimethyl sulfoxide (DMSO) following USP <467> matrix ratios. Calibration standards at 12.5%, 25%, 50%, and 200% of the specified concentration limits were spiked into pharmaceutical product matrices and analyzed in triplicate. Data processing adhered to FDA Title 21 CFR Part 11 using Chromeleon CDS for automated quantitation and validation workflows based on ICH Q2(R1) guidelines.
All target solvents exhibited excellent linearity across their calibration ranges with coefficients of determination (R²) ≥ 0.997 and average response factor RSDs below 12%. LODs and LOQs calculated per ICH Q2(R1) recommendations fell well below the regulatory limits for each class of solvent. Calibration curves for representative Class 1, Class 2A, and Class 2B compounds demonstrated consistent performance at low and high concentration levels. Automated data handling in Chromeleon CDS streamlined validation and reporting, enhancing throughput.
Advancements may include integration with mass spectrometry detectors for enhanced specificity, further miniaturization of headspace units, and implementation of advanced automation and artificial intelligence for predictive maintenance and real-time data analysis. Expanded eWorkflow libraries and cloud-based data management could further accelerate method development and regulatory compliance.
The TriPlus 500 Headspace Autosampler coupled with a TRACE 1310 GC-FID provides robust, sensitive, and reproducible analysis of USP <467> residual solvents. Its high linearity, low detection limits, and compliance with regulatory data standards make it a valuable tool for pharmaceutical quality control.
GC, HeadSpace
IndustriesPharma & Biopharma
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
Testing and controlling residual solvents in pharmaceutical products is critical to ensure patient safety and compliance with regulatory standards. Residual solvents may pose health risks, and international guidelines such as ICH Q3C and USP <467> define acceptable limits and analytical procedures. Headspace gas chromatography coupled with flame ionization detection is a preferred approach due to its efficiency in analyzing volatile organic compounds with minimal sample preparation.
Study Objectives and Overview
This study evaluates the quantitative performance of a Thermo Scientific TriPlus 500 Headspace Autosampler combined with a TRACE 1310 GC equipped with an Instant Connect FID for routine analysis of Class 1, Class 2A, and Class 2B residual solvents according to USP <467> procedure C. The aim is to assess linearity, sensitivity (LOD/LOQ), and repeatability under standard operating conditions.
Methodology
Sample and standard solutions of residual solvents were prepared in dimethyl sulfoxide (DMSO) following USP <467> matrix ratios. Calibration standards at 12.5%, 25%, 50%, and 200% of the specified concentration limits were spiked into pharmaceutical product matrices and analyzed in triplicate. Data processing adhered to FDA Title 21 CFR Part 11 using Chromeleon CDS for automated quantitation and validation workflows based on ICH Q2(R1) guidelines.
Used Instrumentation
- Thermo Scientific TriPlus 500 Headspace Autosampler
- Thermo Scientific TRACE 1310 Gas Chromatograph with Instant Connect Split/Splitless injector
- Thermo Scientific Instant Connect Flame Ionization Detector
- TraceGOLD TG-624 column, 30 m × 0.32 mm × 1.8 µm
Main Results and Discussion
All target solvents exhibited excellent linearity across their calibration ranges with coefficients of determination (R²) ≥ 0.997 and average response factor RSDs below 12%. LODs and LOQs calculated per ICH Q2(R1) recommendations fell well below the regulatory limits for each class of solvent. Calibration curves for representative Class 1, Class 2A, and Class 2B compounds demonstrated consistent performance at low and high concentration levels. Automated data handling in Chromeleon CDS streamlined validation and reporting, enhancing throughput.
Benefits and Practical Applications
- Reliable routine analysis of residual solvents in QC laboratories
- High sensitivity and wide dynamic range for diverse solvent classes
- Reduced sample preparation time through headspace sampling
- Regulatory compliance with automated data integrity and reporting
Future Trends and Opportunities
Advancements may include integration with mass spectrometry detectors for enhanced specificity, further miniaturization of headspace units, and implementation of advanced automation and artificial intelligence for predictive maintenance and real-time data analysis. Expanded eWorkflow libraries and cloud-based data management could further accelerate method development and regulatory compliance.
Conclusion
The TriPlus 500 Headspace Autosampler coupled with a TRACE 1310 GC-FID provides robust, sensitive, and reproducible analysis of USP <467> residual solvents. Its high linearity, low detection limits, and compliance with regulatory data standards make it a valuable tool for pharmaceutical quality control.
Reference
- Impurities: Guideline for Residual Solvents Q3C(R6), ICH Harmonized Guidelines, 2016.
- United States Pharmacopeia General Chapter <467> Organic Volatile Impurities, USP, 2012.
- Thermo Fisher Scientific Application Note: Residual Solvent Analysis, 2018.
- Validation of Analytical Procedures: Text and Methodology Q2(R1), ICH Harmonized Guidelines, 2005.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10676 Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method Authors Giulia Riccardino1, Paolo Magni2, Stefano Pelagatti2, Manuela Bergna2, Davide Bressanello2, and Cristian Cojocariu1 Thermo Fisher Scientific, Runcorn,…
Key words
xylene, xyleneresidual, residualsolvents, solventstetralin, tetralinprocedure, procedurechlorobenzene, chlorobenzenepharmaceutical, pharmaceuticalaccording, accordingtoluene, toluenechloroform, chloroformnitromethane, nitromethaneaspirin, aspirinmethycyclohexane, methycyclohexanepyridine, pyridinetetrahydrofuran
Enhanced Productivity for Residual Solvent Analysis in Pharmaceutical Products According to USP 467 by Using a New Valve and Loop Static Headspace Sampler
2019|Thermo Fisher Scientific|Posters
Enhanced Productivity for Residual Solvent Analysis in Pharmaceutical Products According to USP 467 by Using a New Valve-and-Loop Static Headspace Sampler Xiaoteng Gong1, Benjamin Webber1, Gayatri Trivedi1, Danny Hower1, Dujuan Lu1, Julian Gulbinski1, Cristian I. Cojocariu2, Giulia Riccardino2, Daniela Cavagnino3,…
Key words
xylene, xylenedmso, dmsowater, waterresidual, residualpharmaceutical, pharmaceuticalprocedure, proceduredichloromethane, dichloromethanesoluble, solublesolvents, solventsmethylcyclohexane, methylcyclohexanewere, werediluted, dilutedinsoluble, insolublerepeatability, repeatabilitycyclohexane
Equivalent GC systems performance for regulatory method compliance and validation
2021|Thermo Fisher Scientific|Technical notes
WHITE PAPER 74062 Equivalent GC systems performance for regulatory method compliance and validation Author: Giulia Riccardino, David Lee, and Cristian Cojocariu, Thermo Fisher Scientific, Runcorn, UK Keywords: Gas Chromatography, TRACE 1300 Series, TRACE 1600 Series, method transfer, method portability, system…
Key words
methyl, methylinstant, instantequivalency, equivalencyfid, fidgro, grochromatographic, chromatographicprocedure, procedureconnect, connectsuitability, suitabilityresidual, residualwisconsin, wisconsinmethod, methodssl, sslrequirements, requirementsheadspace
Simplified, cost-effective headspace GC method for residual solvents analysis in pharmaceutical products
2018|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 10681 Simplified, cost-effective headspace GC method for residual solvents analysis in pharmaceutical products Authors Goal Giulia Riccardino and Cristian Cojocariu, Thermo Fisher Scientific, Runcorn, UK The aim of this work was to develop a rapid, cost-effective, modified USP…
Key words
residual, residualvial, vialmethycyclohexane, methycyclohexanexylene, xyleneincubation, incubationloop, loopdichloromethane, dichloromethanechromatographic, chromatographiccumene, cumenesolvent, solventrelief, reliefpain, painheadspace, headspaceshaking, shakingfid