GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Analysis of memantine hydrochloride in accordance with the USP

Applications |  | GL SciencesInstrumentation
GC, GC columns, Consumables
Industries
Pharma & Biopharma
Manufacturer
GL Sciences

Summary

Significance of the topic


Memantine hydrochloride analysis is critical in pharmaceutical quality control to ensure dosage safety efficacy and compliance with pharmacopeial standards. Precise quantification of the active substance and its organic impurities guards against substandard products and potential adverse effects.

Objectives and overview of the study


This study aimed to develop and validate a gas chromatography method in accordance with USP 39 for the assay of memantine hydrochloride and its organic impurities. Using an InertCap 5 capillary column the method was evaluated for resolution tailing factors and reproducibility.

Methodology and instrumentation


  • Instrument Gas chromatograph with flame ionization detector GC FID
  • Column InertCap 5 0.32 mm ID × 50 m film thickness 0.52 μm
  • Carrier gas Helium at 4.0 mL per minute
  • Temperature program 50 °C hold ramp to 145 °C at 5 °C per minute then to 250 °C at 10 °C per minute with a 20 minute hold
  • Injection Split mode 50 to 1 1 μL injection port at 220 °C
  • Detection FID at 300 °C auto range
  • Sample Standard solutions of memantine and related compounds

Main results and discussion


  • Resolution between memantine and adamantane reference was greater than 8.8 exceeding USP requirements of at least 6.0
  • Tailing factors for analytes ranged between 0.85 and 1.18 well within acceptable limits of no more than 2.0
  • Relative standard deviations were below 2.7 percent for impurity peaks demonstrating good reproducibility
  • Chromatograms exhibited sharp peak shapes and clear separation of the active substance from five related impurities facilitating reliable quantification

Benefits and practical applications


This GC FID method offers robust performance for routine quality control of memantine hydrochloride enabling reliable detection of low level impurities. Its reproducibility and compliance with USP standards make it suitable for pharmaceutical manufacturing laboratories focused on batch release stability testing and regulatory submissions.

Future trends and potential applications


  • Integration of faster temperature ramps and shorter columns for reduced analysis time
  • Use of alternative detectors or hyphenated techniques GC MS to enhance impurity identification
  • Automation and software driven data processing to improve throughput in high volume QC environments
  • Adaptation of the method for related compounds in pharmacokinetic and metabolite studies

Conclusion


The described GC FID procedure demonstrates high resolution precision and adherence to USP guidelines for memantine hydrochloride purity testing. The method supports rigorous quality assurance in pharmaceutical industries ensuring product safety and regulatory compliance.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Purity Test of Sucralose, Pharmaceutical Excipients Standard 2018
GC Technical Note Purity Test of Sucralose, Pharmaceutical Excipients Standard 2018 GT115 GL Sciences Inc. Sucralose is an artificial sweetener used as an excipient in pharmaceuticals. The purity test for sucralose listed in the Pharmaceutical Excipients Standard 2013 original required…
Key words
standard, standardstock, stockuvolts, uvoltssucralose, sucraloseinternal, internalheadspace, headspacesciences, sciencestest, testtechnical, technicalamended, amendedfid, fidsolution, solutionmethanol, methanolnote, noteautosampler
Quantitative Test for Hydroxypropylcellulose with Reference to the Japanese Pharmacopoeia, 17th Edition
GC Technical Note Quantitative Test for Hydroxypropylcellulose with Reference to the Japanese Pharmacopoeia, 17th Edition GT093 GL Sciences Inc. The test for hydroxypropyl cellulose was harmonized in the three pharmacopoeias of Japan, Europe and the United States. Based on the…
Key words
hydroxypropylcellulose, hydroxypropylcelluloseuvolts, uvoltsdistributors, distributorsnetherlands, netherlandsinternational, internationalassay, assayfactor, factortechnical, technicalresponse, responsetest, testquantitative, quantitativecarrier, carrierrsd, rsdanalyte, analytenote
Purity Test of Calcium Polystyrene Sulfonate, with reference to the Japanese Pharmacopoeia, 16th Edition - Comparisons of Chromosorb Series and Uniport Series of Siliceous Soil Support
GC Technical Note Purity Test of Calcium Polystyrene Sulfonate, with reference to the Japanese Pharmacopoeia, 16th Edition - Comparisons of Chromosorb Series and Uniport Series of Siliceous Soil Support GT083 GL Sciences Inc. Calcium polystyrene sulfonate is used as a…
Key words
uniport, uniportchromosorb, chromosorbwaw, wawdmcs, dmcsuvolts, uvoltssintered, sinteredseries, seriessiliceous, siliceouswashing, washingmin, minsymmetry, symmetrysciences, scienceshyperkalemia, hyperkalemiatime, timetreated
Purity Test and Fatty Acid Content Ratio Test of Polysorbate 80 - Supplement II to the Japanese Pharmacopoeia, 16th Edition
GC Technical Note GT086 GL Sciences Inc. Purity Test and Fatty Acid Content Ratio Test of Polysorbate 80 - Supplement II to the Japanese Pharmacopoeia, 16th Edition A harmonized test was agreed for Polysorbate 80 in the three pharmacopoeias of…
Key words
uvolt, uvoltmethyl, methyltest, testfatty, fattysciences, sciencesspecified, specifiedmyristate, myristateanalyte, analytesuitability, suitabilityadvice, advicemin, minconditions, conditionsheptane, heptanetime, timesystem
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike