Get Your GC Methods In-Line with the Correct Liner
Presentations | 2022 | Agilent TechnologiesInstrumentation
Accurate sample introduction in gas chromatography (GC) is essential for achieving reliable peak shapes, reproducible quantification and high resolution. The inlet liner plays a central role by providing a controlled vaporization space, minimizing sample discrimination and preventing system contamination. Proper liner selection ensures robust performance across diverse sample matrices and analytical needs.
This work provides a comprehensive guide to selecting and using the correct GC inlet liner. It covers:
Liner selection is based on key performance criteria:
Instrumentation covered includes standard split/splitless GC inlets, multimode inlet (MMI) systems, programmed temperature vaporization (PTV) and cross-lab liners for Thermo, Shimadzu, Varian/Bruker and PerkinElmer systems.
Comparative testing demonstrates that fritted liners provide:
Split injection studies highlight the importance of glass wool to promote vapor mixing at high carrier flows and to minimize discrimination, whereas splitless injections benefit from snug-fitting liners and optimized purge timing to prevent solvent tailing and achieve sample refocusing.
Proper liner choice delivers:
Developments expected include:
Choosing the right GC liner is as critical as column selection. Key recommendations:
These guidelines help laboratories optimize GC methods, improve data quality and reduce downtime related to inlet issues.
GC, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Importance of Topic
Accurate sample introduction in gas chromatography (GC) is essential for achieving reliable peak shapes, reproducible quantification and high resolution. The inlet liner plays a central role by providing a controlled vaporization space, minimizing sample discrimination and preventing system contamination. Proper liner selection ensures robust performance across diverse sample matrices and analytical needs.
Objectives and Study Overview
This work provides a comprehensive guide to selecting and using the correct GC inlet liner. It covers:
- Core liner characteristics (volume, deactivation, features)
- Comparison of split and splitless injection modes
- Specialty liners (direct connect, dimpled, cups) and cross-vendor compatibility
- Practical examples demonstrating liner performance
Methodology and Instrumentation
Liner selection is based on key performance criteria:
- Inner volume: prevents backflash by matching solvent vapor volume to liner capacity
- Surface deactivation: silylation treatments (Ultra Inert, Original) reduce adsorption of active analytes
- Physical features: glass wool or frit for nonvolatile filtration and vapor mixing; tapered profiles for needle wiping and controlled vapor flow
Instrumentation covered includes standard split/splitless GC inlets, multimode inlet (MMI) systems, programmed temperature vaporization (PTV) and cross-lab liners for Thermo, Shimadzu, Varian/Bruker and PerkinElmer systems.
Main Results and Discussion
Comparative testing demonstrates that fritted liners provide:
- More consistent porosity and longer lifetime (up to twice that of wool) in semivolatile analyses (EPA 8270 soil tests)
- Improved peak symmetry and higher signal response for active pesticides (deltamethrin, azoxystrobin) in food matrices
Split injection studies highlight the importance of glass wool to promote vapor mixing at high carrier flows and to minimize discrimination, whereas splitless injections benefit from snug-fitting liners and optimized purge timing to prevent solvent tailing and achieve sample refocusing.
Benefits and Practical Applications
Proper liner choice delivers:
- Enhanced sensitivity and resolution through efficient vaporization and sharp injections
- Extended maintenance intervals by filtering nonvolatiles before column entry
- Consistent quantitative performance across concentration ranges and matrices
- Flexibility to adapt to manual injections, headspace, purge-and-trap or SPME applications
Future Trends and Potential Uses
Developments expected include:
- Advanced inert coatings to further reduce active site interactions
- Customized geometries (spirals, laminar flow cups) for specific sample classes
- Integration of smart inlet liners with pressure and temperature feedback for automated optimization
- Broader cross-platform compatibility to standardize workflows across instrumentation brands
Conclusions
Choosing the right GC liner is as critical as column selection. Key recommendations:
- Match liner features to injection mode: use wool or frit for split, snug inert liners for splitless
- Employ Ultra Inert deactivation for trace, active analytes and fritted designs for extended lifetime
- Leverage specialty liners (direct connect, dimpled, Jennings cup) where matrix or sensitivity demands arise
- Maintain a consistent configuration and perform regular liner inspection and replacement
These guidelines help laboratories optimize GC methods, improve data quality and reduce downtime related to inlet issues.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fundamentals and Troubleshooting of the Split/Splitless Inlet
2020|Agilent Technologies|Presentations
The picture can't be display ed. Fundamentals and Troubleshooting of the Split/Splitless Inlet Rachael Ciotti GCMS Applications Scientist Shannon Coleman GC & GCMS Applications Scientist Kirk Lokits GCMS Applications Scientist DE3988773148 Split/Splitless Inlet GC Sample Introduction Kirk Lokits, Ph.D. Agilent…
Key words
split, splitliner, linersplitless, splitlessinjection, injectioninlet, inletwool, woolliners, linerspurge, purgevent, ventseptum, septumglass, glasssample, samplebackflash, backflashinjections, injectionscolumn
Comparison of Fritted and Wool Liners for Analysis of Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry
2020|Agilent Technologies|Applications
Application Note Environmental Comparison of Fritted and Wool Liners for Analysis of Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is commonly used in the analysis of semivolatile…
Key words
wool, woolliner, linerfrit, fritliners, linersglass, glassbreakdown, breakdownddt, ddtccv, ccvagilent, agilenttrim, trimfailing, failingtaper, taperinitial, initialcalibration, calibrationfritted
Selecting the Right Inlet Liner for Efficient Sample Transfer
2019|Agilent Technologies|Guides
Selecting the Right Inlet Liner for Efficient Sample Transfer GC inlets convert an injected liquid phase sample into gas for transfer onto the GC column. Selecting the wrong liner can lead to incomplete transfer, poor separation, and peak tailing—often incorrectly…
Key words
deactivated, deactivatedfritted, frittedliner, linerdirty, dirtymatrix, matrixinlet, inletpressure, pressurebackflash, backflashtaper, taperlow, lowglass, glassspecialized, specializedclean, cleanstraight, straightissues
Endrin/DDT Stability for Fritted and Wool Liners
2020|Agilent Technologies|Applications
Application Note Environmental Endrin/DDT Stability for Fritted and Wool Liners Authors Lukas Wieder and Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract This application note demonstrates the lifetime of Agilent Ultra Inert fritted liners for environmental applications with complex, nonvolatile…
Key words
wool, woolfrit, fritmid, midliner, linerendrin, endrinlow, lowbreakdown, breakdownagilent, agilentddt, ddtliners, linersfritted, frittedmixture, mixtureddd, dddinert, inertdde