Analytical and Retention Index of 61 Components of Organic Solvents - Using InertCap WAX

Applications |  | GL SciencesInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
GL Sciences

Summary

Importance of Retention Index Analysis


This study focuses on the retention index—a standardized measure that relates chromatographic retention times of unknown compounds to those of straight-chain alkanes. By expressing elution behaviour on a scale proportional to carbon number, retention indices enable reliable compound identification in complex organic solvent mixtures. Such indices are invaluable in analytical laboratories for qualitative screening, method development, and building robust spectral libraries.

Objectives and Study Overview


The primary goal was to determine the retention indices of 61 common organic solvent components using an InertCap WAX capillary column under both temperature programming and isothermal conditions. This dual-mode approach evaluates the linearity of retention behaviour against the number of carbons, verifying the method’s suitability for routine qualitative GC analysis.

Methodology and Instrumentation


The experiments were performed on a gas chromatograph equipped with a flame ionization detector (GC-FID). Key chromatographic parameters included:
  • Column: InertCap WAX, 60 m × 0.25 mm I.D., 0.25 µm film thickness
  • Carrier gas: Helium at 160 kPa
  • Oven programming: 40 °C hold, ramp at 5 °C/min to 220 °C
  • Isothermal runs: maintained at 40 °C, 80 °C, 120 °C, and 160 °C with adjusted pressures for constant flow
  • Injection: split mode, 150 mL/min at 240 °C
  • Detection: FID at 240 °C, range 10^0
  • Sample: 0.2 µL of an evenly mixed standard solution

Results and Discussion


Retention indices measured under temperature programming ranged from approximately 599 for n-hexane to 2121 for m-cresol, illustrating a near-linear increase with carbon number. Isothermal analysis at different oven temperatures produced similar index values with minor shifts, confirming the robustness of the InertCap WAX phase for solvent analysis. Key observations:
  • Linear relationship between retention index and carbon count held true for both thermal ramp and isothermal conditions.
  • Light solvents (C6–C8) yielded indices near 600–900, while aromatic and oxygenated compounds (e.g., toluene, benzene, acetone) occupied the mid-range (900–1050).
  • Heavier and multifunctional solvents (e.g., cyclohexanone, dimethylacetamide, phenols) delivered indices above 1300, up to ~2121.
  • Isothermal runs at higher temperatures compressed retention times but preserved index order, facilitating faster screening when throughput is critical.

Benefits and Practical Applications


Determining accurate retention indices for common solvents supports:
  • Rapid qualitative identification in environmental, pharmaceutical, and industrial laboratories.
  • Development of in-house retention index libraries compatible with vendor or open-source databases.
  • Method transferability across instruments by referencing a universal alkane scale.
  • Quality control workflows where unknown solvent contaminants must be flagged quickly.

Future Trends and Opportunities


Advances likely to expand retention index utility include:
  • Coupling GC–MS libraries with retention index filters for enhanced deconvolution and false-positive reduction.
  • Automated retention index calculators within data processing software to streamline method development.
  • Application of high-temperature and high-flow columns for ultra-fast screening in high-throughput laboratories.
  • Integration with machine learning to predict retention indices of novel or uncharacterized compounds.

Conclusion


This technical note demonstrates that InertCap WAX provides reliable, reproducible retention indices for a broad set of organic solvents under both programmed and isothermal conditions. The strong linear correlation with alkane carbon number confirms its suitability for qualitative GC analysis and library building in diverse analytical settings.

Used Instrumentation


  • GC System with FID detector
  • InertCap WAX capillary column (60 m × 0.25 mm × 0.25 µm)
  • Helium carrier gas at 160 kPa
  • Split injection (150 mL/min) and FID at 240 °C

Reference


GL Sciences Inc. GC Technical Note GT065, “Analytical and Retention Index of 61 Components of Organic Solvents – Using InertCap WAX.”

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis and Retention Index of 61 Components of Organic Solvents using Nitrogen Carrier Gas - Using InertCap Pure-WAX
GC Technical Note Analysis and Retention Index of 61 Components of Organic Solvents using Nitrogen Carrier Gas - Using InertCap Pure-WAX GT079 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes…
Key words
retention, retentionindex, indexcellosolve, cellosolvetime, timealkanes, alkanesstraight, straightisoamyl, isoamylalcohol, alcoholacetate, acetatechain, chaincomponent, componentt'r, t'rcomponents, componentsprecede, precedesciences
Analysis and Retention Indicators for 61 Organic Solvents - Using InertCap 25
GC Technical Note Analysis and Retention Indicators for 61 Organic Solvents Using InertCap 25 GT074 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes and is used to study constituents based…
Key words
retention, retentionindex, indexcellosolve, cellosolvetime, timealkanes, alkanesstraight, straightisoamyl, isoamylalcohol, alcoholacetate, acetatechain, chaint'r, t'rcomponent, componentcomponents, componentsprecede, precedesciences
Analytical ad Retention Index of 61 Components of Organic Solvents - Using InertCap 5
GC Technical Note Analytical ad Retention Index of 61 Components of Organic Solvents - Using InertCap 5 GT063 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes and is used to…
Key words
retention, retentionindex, indexcellosolve, cellosolvetime, timealkanes, alkanesstraight, straightisoamyl, isoamylalcohol, alcoholcomponent, componentacetate, acetatechain, chaint'r, t'rprecede, precedesciences, sciencesisopentyl
Analytical and Retention Index of 61 Components of Organic Solvents - Using InertCap 1
GC Technical Note Analytical and Retention Index of 61 Components of Organic Solvents - Using InertCap 1 GT053 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes and is used to…