Analytical and Retention Index of 61 Components of Organic Solvents - Using InertCap 1

Applications |  | GL SciencesInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
GL Sciences

Summary

Analytical and Retention Index of 61 Organic Solvent Components Using InertCap 1



Significance of the Topic


Accurate determination of retention indices is essential for reliable identification of volatile organic compounds in complex mixtures. Retention indices based on straight-chain alkanes provide a reproducible parameter that correlates with carbon number, supporting robust qualitative analysis across laboratories and instruments.

Study Objectives


This technical report aimed to establish retention indices for 61 common organic solvents using an InertCap 1 capillary column. Both isothermal and temperature-programmed gas chromatography were evaluated to compare index values and assess method suitability for qualitative screening.

Methodology and Instrumentation


Gas chromatography with flame ionization detection (GC-FID) was employed. The InertCap 1 column (0.25 mm I.D., 60 m length, 0.25 µm film) operated under two modes:
  • Temperature programming: 40 °C initial, ramp at 5 °C/min to 220 °C, helium carrier at 160 kPa, split injection at 240 °C, FID at 240 °C.
  • Isothermal runs: constant temperatures of 40, 80, 120, and 160 °C.

Main Results and Discussion


Under temperature programming, retention indices ranged from 357 for methanol to 1047 for m-cresol, showing a linear increase with carbon number. Isothermal analysis produced consistent indices across four temperature settings, with minor shifts reflecting temperature sensitivity. The linear relationship between log retention ratios and carbon count was confirmed in both modes, validating the general retention index equation and demonstrating the column’s inertness and reproducibility.

Practical Benefits and Applications


Establishing a reliable retention index library enhances qualitative screening in environmental monitoring, pharmaceutical quality control, and petrochemical analysis. Key advantages include:
  • Improved compound identification by matching observed indices to reference values.
  • Method transferability across laboratories using InertCap 1 columns.
  • Support for regulatory compliance through standardized retention parameters.

Future Trends and Applications


Integration of retention index libraries with mass spectrometry and advanced data-processing software will further accelerate compound identification. Machine learning approaches may predict retention indices for novel analytes, while ultrafast GC methods could expand throughput without sacrificing index reliability.

Conclusion


This study presents a comprehensive retention index dataset for 61 organic solvents on an InertCap 1 column under both isothermal and temperature-programmed conditions. The results confirm the method’s linearity, reproducibility, and practical utility for qualitative GC analysis.

Reference


  • GL Sciences Inc. GC Technical Note: Analytical and Retention Index of 61 Components of Organic Solvents Using InertCap 1.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis and Retention Index of 61 Components of Organic Solvents using Nitrogen Carrier Gas - Using InertCap Pure-WAX
GC Technical Note Analysis and Retention Index of 61 Components of Organic Solvents using Nitrogen Carrier Gas - Using InertCap Pure-WAX GT079 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes…
Key words
retention, retentionindex, indexcellosolve, cellosolvetime, timealkanes, alkanesstraight, straightisoamyl, isoamylalcohol, alcoholacetate, acetatechain, chaincomponent, componentt'r, t'rcomponents, componentsprecede, precedesciences
Analytical ad Retention Index of 61 Components of Organic Solvents - Using InertCap 5
GC Technical Note Analytical ad Retention Index of 61 Components of Organic Solvents - Using InertCap 5 GT063 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes and is used to…
Key words
retention, retentionindex, indexcellosolve, cellosolvetime, timealkanes, alkanesstraight, straightisoamyl, isoamylalcohol, alcoholcomponent, componentacetate, acetatechain, chaint'r, t'rprecede, precedesciences, sciencesisopentyl
Analysis and Retention Indicators for 61 Organic Solvents - Using InertCap 25
GC Technical Note Analysis and Retention Indicators for 61 Organic Solvents Using InertCap 25 GT074 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes and is used to study constituents based…
Key words
retention, retentionindex, indexcellosolve, cellosolvetime, timealkanes, alkanesstraight, straightisoamyl, isoamylalcohol, alcoholacetate, acetatechain, chaint'r, t'rcomponent, componentcomponents, componentsprecede, precedesciences
Analytical and Retention Index of 61 Components of Organic Solvents - Using InertCap WAX
GC Technical Note Analytical and Retention Index of 61 Components of Organic Solvents - Using InertCap WAX GT065 GL Sciences Inc. The retention index is a relatively representative index of the retention ratio of straight-chain alkanes and is used to…
Key words
retention, retentionindex, indexalkanes, alkanescellosolve, cellosolvetime, timestraight, straightchain, chaint'r, t'ralcohol, alcoholcomponents, componentsprecede, precedesciences, sciencesinterest, interestmibk, mibkcomponent
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike