Highly Reproducible Detailed cis/trans FAMEs Analysis Ensured by New Optimized Rt-2560 Column Manufacturing and Application-Specific QC Test
Applications | 2017 | RestekInstrumentation
Accurate separation and quantification of cis and trans fatty acid methyl esters (FAMEs) in food and oils is essential for nutritional labeling, regulatory compliance and product quality control. Detailed cis/trans profiling supports enforcement of dietary trans fat limits, informs health risk assessments and guides product formulation across the food industry.
This work describes enhancements in manufacturing of Restek’s Rt-2560 gas chromatography column and development of an application-specific quality control (QC) test. The goals were to minimize column-to-column variability, preserve high selectivity for cis/trans isomers and ensure robust performance for standard methods, including AOAC 996.06 and AOCS Ce 1j-07.
Columns were produced under an optimized coating process and subjected to a new QC test using a representative FAME test mix. Chromatographic evaluation included retention index, retention factor, resolution and plate count metrics. Key parameters:
Columns produced under the revised coating procedure demonstrated highly consistent retention times and low bleed, even at 250 °C. Comparative chromatograms of lot-to-lot samples showed negligible shifts in elution order. The QC test confirmed equivalent separation of key cis/trans pairs (e.g., C18:1-t9 vs. C18:1-c9) and baseline resolution of critical FAMEs such as C20:1-c11 and C18:3-c9,c12,c15. Retention indices measured against homologous saturated FAMEs ensured phase selectivity was maintained without artifacts from n-alkane bracketing.
Advances in column manufacturing and QC may extend to other stationary phases for lipid profiling, including simultaneous analysis of triacylglycerols and FAMEs. Integration with two-dimensional GC (GC×GC) and high-resolution MS will further enhance isomer differentiation. Automated QC algorithms and real-time performance monitoring may streamline column qualification in high-throughput environments.
The optimized manufacturing process and targeted QC test guarantee that Rt-2560 columns deliver consistent, high-resolution separation of cis/trans FAMEs. These improvements facilitate compliance with regulatory methods, improve data reliability and reduce operational variability in food and oil analysis.
GC, GC columns, Consumables
IndustriesFood & Agriculture
ManufacturerAgilent Technologies, Restek
Summary
Significance of the Topic
Accurate separation and quantification of cis and trans fatty acid methyl esters (FAMEs) in food and oils is essential for nutritional labeling, regulatory compliance and product quality control. Detailed cis/trans profiling supports enforcement of dietary trans fat limits, informs health risk assessments and guides product formulation across the food industry.
Objectives and Study Overview
This work describes enhancements in manufacturing of Restek’s Rt-2560 gas chromatography column and development of an application-specific quality control (QC) test. The goals were to minimize column-to-column variability, preserve high selectivity for cis/trans isomers and ensure robust performance for standard methods, including AOAC 996.06 and AOCS Ce 1j-07.
Methodology and Instrumentation
Columns were produced under an optimized coating process and subjected to a new QC test using a representative FAME test mix. Chromatographic evaluation included retention index, retention factor, resolution and plate count metrics. Key parameters:
- Column: Rt-2560, 100 m × 0.25 mm × 0.20 µm polysiloxane stationary phase
- Instrument: Agilent 7890B GC with FID at 285 °C
- Carrier gas: Helium, constant flow 1.3 mL/min (or hydrogen at 2.5 mL/min in QC test)
- Injection: 1 µL split (20:1 or 200:1) with precision liner
- Oven program: Initial hold at 100 °C, ramp to 250 °C at 3 °C/min, final hold
Main Results and Discussion
Columns produced under the revised coating procedure demonstrated highly consistent retention times and low bleed, even at 250 °C. Comparative chromatograms of lot-to-lot samples showed negligible shifts in elution order. The QC test confirmed equivalent separation of key cis/trans pairs (e.g., C18:1-t9 vs. C18:1-c9) and baseline resolution of critical FAMEs such as C20:1-c11 and C18:3-c9,c12,c15. Retention indices measured against homologous saturated FAMEs ensured phase selectivity was maintained without artifacts from n-alkane bracketing.
Benefits and Practical Applications of the Method
- High reproducibility reduces revalidation effort in regulated laboratories.
- Consistent selectivity enables reliable trans fat quantification.
- Low bleed and thermal stability support extended column life.
- Compatibility with established AOAC and AOCS methods ensures broad adoption.
Future Trends and Potential Applications
Advances in column manufacturing and QC may extend to other stationary phases for lipid profiling, including simultaneous analysis of triacylglycerols and FAMEs. Integration with two-dimensional GC (GC×GC) and high-resolution MS will further enhance isomer differentiation. Automated QC algorithms and real-time performance monitoring may streamline column qualification in high-throughput environments.
Conclusion
The optimized manufacturing process and targeted QC test guarantee that Rt-2560 columns deliver consistent, high-resolution separation of cis/trans FAMEs. These improvements facilitate compliance with regulatory methods, improve data reliability and reduce operational variability in food and oil analysis.
References
- AOAC 996.06: Standard Method for Trans Fat Analysis by GC
- AOCS Ce 1j-07: Official Method for cis/trans Fatty Acid Determination
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
High-Resolution GC Analyses of Fatty Acid Methyl Esters (FAMEs)
2018|Agilent Technologies|Applications
High-Resolution GC Analyses of Fatty Acid Methyl Esters (FAMEs) Fatty acid methyl esters (FAMEs) analysis is an important tool both for characterizing fats and oils and for determining the total fat content in foods. Fats can be extracted from a…
Key words
fame, famemethyl, methylnomenclature, nomenclaturemouth, mouthfames, famesoven, ovenmin, mincarrier, carriertrans, transbig, bigpeaks, peaksfat, fatstructural, structuralinsert, insertcrossbond
Unsaturated FAMES QC Test Mix on Rt-2560
|Agilent Technologies|Applications
Unsaturated FAMES QC Test Mix on Rt-2560 Peaks tR (min) 1. Methyl octadecanoate 11.68 2. Methyl elaidate 12.64 3. Methyl oleate 13.09 4. Methyl nonadecanoate 13.84 5. Methyl eicosanoate 16.66 6. Methyl eicosenoate 18.85 7. Methyl linolenate 19.20 8.…
Key words
nomenclature, nomenclaturemake, makeunsaturated, unsaturatedgas, gasfames, fameshexane, hexanestructural, structuralmin, minmix, mixtest, testoven, ovenpeaks, peakscarrier, carrierdetector, detectorinstrument
Rt-2560 Columns Ensure Accurate, Reliable AOAC 996.06 and AOCS Ce 1j-07 FAMEs Analysis
2017|Agilent Technologies|Applications
Rt-2560 Columns Ensure Accurate, Reliable AOAC 996.06 and AOCS Ce 1j-07 FAMEs Analysis By Rebecca Stevens and Kristi Sellers Restek’s Rt-2560 GC column was originally developed for detailed analysis of FAMEs in foods and food products. Its highly polar biscyanopropyl…
Key words
nomenclature, nomenclaturestructural, structuralmin, mindichloromethane, dichloromethanekristi, kristisellers, sellerstimetime, timetimehexane, hexanepeaks, peaksstevens, stevensrebecca, rebeccaoven, ovencarrier, carrierrestek, restekdetector
Food Industry FAMEs on Rt-2560 by AOCS Method Ce-1h-05
2019|Agilent Technologies|Applications
Food Industry FAMEs on Rt-2560 by AOCS Method Ce-1h-05 11 3 2 5 1 7 4 6 89 10 15 17 13 12 14 20 16 18 19 10 22 23 21 24 20 26 25 30 29 27 28…
Key words