Detector-Switching Analysis Using a Capillary Switching Device

Applications |  | ShimadzuInstrumentation
GC, HeadSpace
Industries
Environmental
Manufacturer
Shimadzu

Summary

Importance of the Topic


The development of a capillary switching device enables precise and simultaneous analyses by multiple detectors within a single gas chromatography (GC) run. This approach preserves analytical sensitivity and accuracy by directing the full sample flow to the selected detector at defined times, overcoming limitations of conventional detector splitting techniques.

Objectives and Study Overview


This application note demonstrates three key uses of detector-switching analysis with a capillary switching device:
  • Simultaneous detection of pesticide residues using electron capture detector (ECD) and flame photometric detector (FPD).
  • Removal of solvent interference (dichloromethane) prior to detection.
  • Elimination of headspace air in volatile organic compound (VOC) analysis to enhance ECD stability.

Methodology and Instrumentation


All experiments were performed on a Shimadzu GC-2010 Plus system equipped with a capillary switching device and dedicated control software. Key analytical parameters included:
  • Columns: Rtx-5MS and Rtx-1 for pesticide and solvent elimination studies; DB-624 coupled to TurboMatrix HS40 for headspace VOC analysis.
  • Carrier gas: Helium at constant pressure (150–234 kPa).
  • Detectors: ECD (300 °C, 1 nA) and FPD (300 °C, H₂/Air), or headspace-ECD (250 °C, N₂ make-up).
  • Switching pressure: 90 kPa; restrictors of 0.5 m length with 0.15–0.18 mm I.D.
  • Injection modes: Splitless for pesticide/solvent studies and split (1:4 or 1:15) for headspace and solvent elimination.

Main Results and Discussion


1. Pesticide Analysis (ECD/FPD Switching): Chromatograms showed clear separation and detection of multiple organochlorine and organophosphorus pesticides. The device switched flow between FPD and ECD at preprogrammed times, achieving high sensitivity for sulfur- or phosphorus-containing analytes and halogenated compounds without cross-detector dilution.

2. Solvent Elimination: By diverting dichloromethane to waste prior to detectors, the system prevented ECD baseline disturbances and FPD source degradation. Comparative chromatograms confirmed the effective removal of the solvent peak and improved detector lifespan.

3. Air Elimination in Headspace ECD: Prior to analyte introduction, headspace air was vented, minimizing oxygen-induced sensitivity fluctuations. The stability of ECD response for VOCs was significantly enhanced, as illustrated by reproducible peak areas for trihalomethanes and chlorinated solvents.

Benefits and Practical Applications


The capillary switching approach offers:
  • Enhanced analytical sensitivity by delivering the entire sample to each detector.
  • Improved selectivity through time-programmed detector assignments.
  • Reduction of chemical interferences (solvents, air) to maintain detector integrity.
  • Streamlined workflows via a single injection for multi-detector data acquisition.

Future Trends and Potential Applications


Advancements may include integration with mass spectrometry for comprehensive profiling, automated switching protocols for complex sample matrices, and expansion into environmental, food safety, and petrochemical analyses. Enhanced software interfaces and miniaturized switching modules could further broaden applications in high-throughput and field-deployable GC systems.

Conclusion


Detector-switching using a capillary device significantly improves GC multi-detector performance by combining sensitivity, selectivity, and interference elimination in a single run. This versatile technique is practical for pesticide monitoring, solvent cleanup, and headspace VOC analysis, with promising prospects for wider analytical use.

Reference


Shimadzu Application News LAAN-A-GC-E033A

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Gas Sample Using Single-Oven MDGC
LAAN-A-GC-E032 Application News Gas Chro mato g r a p h y No.G265 Analysis of Gas Sample Using Single-Oven MDGC MDGC. A switching program can easily be created with special software that can be downloaded from the Shimadzu website free…
Key words
tcd, tcdswitching, switchingfid, fidrouted, routedmdgc, mdgccolumn, columninj, injgas, gasdevice, devicemethane, methanedetector, detectorcut, cutmato, matoinorganic, inorganicinjection
Screening for Hazardous Chemicals in Homeland Security and Environmental Samples Using a GC/MS/ECD/FPD with a 731 Compound DRS Database
Screening for Hazardous Chemicals in Homeland Security and Environmental Samples Using a GC/MS/ECD/FPD with a 731 Compound DRS Database Application Note Homeland Security, Environmental Authors Bruce Quimby Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 UASA Abstract…
Key words
methyl, methylfpd, fpdamdis, amdissim, simester, esterphthalate, phthalatemsd, msdsplit, splitscan, scanecd, ecddfpd, dfpdsplitter, splitterhazardous, hazardousparathion, parathionepc
Shimadzu Gas Chromatograph Nexis GC-2030
Shimadzu Gas Chromatograph Nexis GC-2030
2025|Shimadzu|Brochures and specifications
C184-E043E Gas Chromatograph Nexis GC-2030 GC-2030 The Next Industry Standard In today's environment, results are needed regardless of whether the analyst is working in the laboratory, the office, or from home. With the flexibility to accommodate a variety of applications,…
Key words
analysis, analysisgas, gasfunction, functionlabsolutions, labsolutionschromatograph, chromatographanaly, analydetector, detectorcarrier, carriersulfur, sulfurbeer, beerstatus, statussystem, systemselector, selectormind, mindhigh
Agilent 8890 Gas Chromatograph - Data Sheet
Agilent 8890 Gas Chromatograph - Data Sheet
2021|Agilent Technologies|Brochures and specifications
Data Sheet Agilent 8890 Gas Chromatograph Chromatographic performance* • Retention time repeatability <0.008 % or <0.0008 minutes • Area repeatability <0.5 % RSD The Agilent 8890 GC is a state-of-the-art gas chromatograph that provides superior performance for all applications. Key…
Key words
epc, epcpressure, pressurecontrol, controldetector, detectortemperature, temperaturecapillary, capillaryelectronic, electronicflow, flowauxiliary, auxiliarymounted, mountedseptum, septuminlets, inletscompensated, compensatedpurged, purgedcolumn
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike