Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency10 m Capillary GC Column with Optimization in Three Carrier Gases
Applications | 2016 | Thermo Fisher ScientificInstrumentation
Accurate and efficient separation of fatty acid methyl esters (FAMEs) is critical for food quality control, nutritional analysis and regulatory compliance. Traditional methods using long GC columns limit sample throughput and raise analytical costs. Implementing a short, high-efficiency 10 m TRACE TR-FAME column addresses these challenges by delivering rapid, high-resolution separations suitable for routine laboratory workflows.
Using helium, all 37 FAMEs eluted in 11.9 min with maintained resolution comparable to a 100 m column but with 5× faster analysis. Hydrogen further reduced run time to 9.5 min, with nearly all peak pairs remaining baseline resolved (resolution > 1.5 for most). Nitrogen delivered the highest chromatographic efficiency at the cost of longer analysis (23.7 min), improving resolution of critical pairs (25–26 and 28–29) by up to 22% relative to hydrogen. Plate counts for these critical peaks increased by an average of 39% versus hydrogen. Sample throughput calculations show up to 100 samples/day with hydrogen, 80/day with helium and 48/day with nitrogen, compared to 24/day on a standard 100 m column.
Ongoing developments in short, high-efficiency GC columns will further accelerate fatty acid profiling in food, biofuel and clinical research. Hydrogen is emerging as an attractive helium alternative amid supply constraints, while novel stationary phases may enhance selectivity for specialized isomeric analyses. Integration with multidimensional GC and high-resolution mass spectrometry promises deeper insight into complex lipid matrices.
A 10 m TRACE TR-FAME column offers a robust, high-throughput solution for profiling 37 FAMEs. Carrier gas choice enables method tuning: hydrogen for fastest analysis, nitrogen for maximum resolution and helium for balanced performance. This approach streamlines fatty acid analysis, lowers costs and supports demanding QA/QC environments.
GC, GC columns, Consumables
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Accurate and efficient separation of fatty acid methyl esters (FAMEs) is critical for food quality control, nutritional analysis and regulatory compliance. Traditional methods using long GC columns limit sample throughput and raise analytical costs. Implementing a short, high-efficiency 10 m TRACE TR-FAME column addresses these challenges by delivering rapid, high-resolution separations suitable for routine laboratory workflows.
Objectives and Study Overview
- Demonstrate separation of 37 common FAMEs on a 10 m × 0.10 mm × 0.20 µm TRACE TR-FAME column.
- Compare performance and sample throughput using three carrier gases: helium, hydrogen and nitrogen.
- Evaluate improvements in analysis time, resolution of critical pairs and chromatographic efficiency versus a conventional 100 m column.
Methodology and Instrumentation
- Sample preparation: direct injection of a standard mixture of 37 FAMEs after extraction, saponification and methyl derivatization.
- Column: Thermo Scientific TRACE TR-FAME, 10 m × 0.10 mm × 0.20 µm (70% cyanopropyl polysilphenylene-siloxane).
- GC system: Thermo Scientific TRACE 1310 with TriPlus RSH autosampler and Instant Connect FID detector; data collected in Chromeleon 7.2 SR3.
- Carrier gases: experiments performed under constant-flow split mode with helium, hydrogen or nitrogen; optimized oven programs applied in each case.
Main Results and Discussion
Using helium, all 37 FAMEs eluted in 11.9 min with maintained resolution comparable to a 100 m column but with 5× faster analysis. Hydrogen further reduced run time to 9.5 min, with nearly all peak pairs remaining baseline resolved (resolution > 1.5 for most). Nitrogen delivered the highest chromatographic efficiency at the cost of longer analysis (23.7 min), improving resolution of critical pairs (25–26 and 28–29) by up to 22% relative to hydrogen. Plate counts for these critical peaks increased by an average of 39% versus hydrogen. Sample throughput calculations show up to 100 samples/day with hydrogen, 80/day with helium and 48/day with nitrogen, compared to 24/day on a standard 100 m column.
Benefits and Practical Applications
- Up to 400% increase in laboratory throughput compared to a 100 m column.
- Flexibility to balance analysis speed and resolution by selecting an appropriate carrier gas.
- Reduced operating costs through shorter analysis times and lower gas consumption.
- High thermal stability and low bleed phase compatible with FID and MS detection.
- Improved quantitation confidence for closely eluting FAME pairs.
Future Trends and Potential Applications
Ongoing developments in short, high-efficiency GC columns will further accelerate fatty acid profiling in food, biofuel and clinical research. Hydrogen is emerging as an attractive helium alternative amid supply constraints, while novel stationary phases may enhance selectivity for specialized isomeric analyses. Integration with multidimensional GC and high-resolution mass spectrometry promises deeper insight into complex lipid matrices.
Conclusion
A 10 m TRACE TR-FAME column offers a robust, high-throughput solution for profiling 37 FAMEs. Carrier gas choice enables method tuning: hydrogen for fastest analysis, nitrogen for maximum resolution and helium for balanced performance. This approach streamlines fatty acid analysis, lowers costs and supports demanding QA/QC environments.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Modifying AOAC Method 996.06 for FAME Analysis in Foods: Faster Throughput Using Hydrogen Carrier Gas
2020|Shimadzu|Applications
No. SSI-GC-2006 Gas Chromatography No. GC-2006 Modifying AOAC Method 996.06 for FAME Analysis in Foods: Faster Throughput Using Hydrogen Carrier Gas ■ Abstract As the cost of helium continues to rise, more labs are converting to hydrogen carrier gas for…
Key words
methyl, methyleicosatrienoate, eicosatrienoatehydrogen, hydrogenaoac, aoaclinolenate, linolenatemethod, methodvelocity, velocitylinear, lineargas, gasarchidate, archidatearchidonate, archidonatecarrier, carrierdocosadienoate, docosadienoateheptadecenoate, heptadecenoatepentadecenoate
Equivalent GC systems performance for regulatory method compliance and validation
2021|Thermo Fisher Scientific|Technical notes
WHITE PAPER 74062 Equivalent GC systems performance for regulatory method compliance and validation Author: Giulia Riccardino, David Lee, and Cristian Cojocariu, Thermo Fisher Scientific, Runcorn, UK Keywords: Gas Chromatography, TRACE 1300 Series, TRACE 1600 Series, method transfer, method portability, system…
Key words
methyl, methylinstant, instantequivalency, equivalencyfid, fidgro, grochromatographic, chromatographicprocedure, procedureconnect, connectsuitability, suitabilityresidual, residualwisconsin, wisconsinmethod, methodssl, sslrequirements, requirementsheadspace
A GC-FID Method for the Comparison of Acid- and Base-Catalyzed Derivatization of Fatty Acids to FAMEs in Three Edible Oils
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words TR-FAME, fatty acid methyl esters (FAMEs), BF3-methanol, derivatization, cis- and trans-fatty acid Abstract This application note demonstrates the analysis of 37 fatty acid methyl esters (FAMEs) separated by a highly…
Key words
fatty, fattyscientific, scientificresponse, responsereacti, reactiesterified, esterifiedfames, famesretention, retentionesterification, esterificationfat, fatbase, basemin, minfame, fameacid, acidfisher, fishergrade
Biodiesel quality assessment: determination of esters and linolenic acid methyl ester content in biodiesel (B100) by GC-FID, according to EN 14103
2023|Thermo Fisher Scientific|Applications
Application note | 001899 Industrial Biodiesel quality assessment: determination of esters and linolenic acid methyl ester content in biodiesel (B100) by GC-FID, according to EN 14103 Authors Goal Giulia Riccardino1, Daniela Cavagnino1, and The aim of this study is to…
Key words
triplus, triplusmethyl, methyllinolenic, linolenicrsh, rshesters, esterssmart, smartbiodiesel, biodieselmin, minstation, stationester, esteristd, istdacid, acidtcf, tcfautosampler, autosamplerrepeatability