Determination of dioxins and dibenzofurans using GC/MS
Applications | 2011 | Agilent TechnologiesInstrumentation
Dioxins and dibenzofurans are persistent environmental pollutants with recognized toxicity and bioaccumulation potential. Accurate quantification of these congeners is critical for environmental monitoring, regulatory compliance, and risk assessment.
This application note presents a gas chromatography/mass spectrometry (GC/MS) method for reliable separation and identification of 36 dioxin and dibenzofuran congeners. Utilizing Agilent FactorFour VF-5ms columns, the study aims to achieve baseline resolution and robust detection employing high-resolution magnetic sector MS.
The method uses an Agilent FactorFour VF-5ms fused silica capillary column (60 m × 0.25 mm, 0.25 µm film). The temperature program begins at 160 °C, ramps at 30 °C/min to 220 °C, then at 5 °C/min to 230 °C and 310 °C with holds. Helium is the carrier gas at constant pressure (290 kPa). Samples are introduced splitless at 260 °C. Detection is performed by high-resolution magnetic sector mass spectrometry with an interface temperature of 260 °C.
The optimized GC temperature program provided clear separation of all target congeners, including tetra- through octa-chlorinated dibenzo-p-dioxins and dibenzofurans. High-resolution MS enabled unequivocal peak identification and quantitation. Chromatograms demonstrate baseline resolution for challenging isomer pairs, ensuring reliable analysis in complex matrices.
This method offers high sensitivity and selectivity for environmental and industrial laboratories conducting QA/QC and regulatory monitoring of dioxins and furans. The robust column performance and reproducible MS detection support long-term operation with minimal maintenance and high throughput.
Advancements in ultrahigh resolution MS and multidimensional GC promise further improvements in congener separation and detection limits. Emerging applications include real-time monitoring, non-target screening, and integration with automated sample preparation workflows for broader environmental and food safety assessment.
The described GC/MS method using Agilent FactorFour VF-5ms columns and high-resolution sector MS delivers comprehensive separation and accurate quantitation of dioxins and dibenzofurans. It fulfills stringent analytical requirements, facilitating reliable environmental and industrial monitoring.
GC/MSD, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Dioxins and dibenzofurans are persistent environmental pollutants with recognized toxicity and bioaccumulation potential. Accurate quantification of these congeners is critical for environmental monitoring, regulatory compliance, and risk assessment.
Objectives and Study Overview
This application note presents a gas chromatography/mass spectrometry (GC/MS) method for reliable separation and identification of 36 dioxin and dibenzofuran congeners. Utilizing Agilent FactorFour VF-5ms columns, the study aims to achieve baseline resolution and robust detection employing high-resolution magnetic sector MS.
Methodology and Instrumentation
The method uses an Agilent FactorFour VF-5ms fused silica capillary column (60 m × 0.25 mm, 0.25 µm film). The temperature program begins at 160 °C, ramps at 30 °C/min to 220 °C, then at 5 °C/min to 230 °C and 310 °C with holds. Helium is the carrier gas at constant pressure (290 kPa). Samples are introduced splitless at 260 °C. Detection is performed by high-resolution magnetic sector mass spectrometry with an interface temperature of 260 °C.
Main Results and Discussion
The optimized GC temperature program provided clear separation of all target congeners, including tetra- through octa-chlorinated dibenzo-p-dioxins and dibenzofurans. High-resolution MS enabled unequivocal peak identification and quantitation. Chromatograms demonstrate baseline resolution for challenging isomer pairs, ensuring reliable analysis in complex matrices.
Benefits and Practical Application
This method offers high sensitivity and selectivity for environmental and industrial laboratories conducting QA/QC and regulatory monitoring of dioxins and furans. The robust column performance and reproducible MS detection support long-term operation with minimal maintenance and high throughput.
Future Trends and Potential Applications
Advancements in ultrahigh resolution MS and multidimensional GC promise further improvements in congener separation and detection limits. Emerging applications include real-time monitoring, non-target screening, and integration with automated sample preparation workflows for broader environmental and food safety assessment.
Conclusion
The described GC/MS method using Agilent FactorFour VF-5ms columns and high-resolution sector MS delivers comprehensive separation and accurate quantitation of dioxins and dibenzofurans. It fulfills stringent analytical requirements, facilitating reliable environmental and industrial monitoring.
References
- Agilent Technologies, Inc. Application Note A02459, October 31, 2011.
- Fishman V.N., Martin G.D., Lamparski L.L., The Dow Chemical Company, Midland, MI.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Tetra- Through Octa-Chlorinated Dioxins and Furans Analysis in Water by Isotope Dilution GC/MS/MS
2019|Agilent Technologies|Applications
Application Note Environmental, Food Testing and Agriculture Tetra- Through Octa-Chlorinated Dioxins and Furans Analysis in Water by Isotope Dilution GC/MS/MS Authors Hui Lin, Jessica Betz, Jim Dunn, Greg Martin, and David Somerville The DOW Chemical Company Midland, MI, USA Diana…
Key words
cdds, cddscdfs, cdfspecdf, pecdftcdd, tcddtcdf, tcdfdioxin, dioxinhxcdf, hxcdfpecdd, pecddhpcdf, hpcdfhpcdd, hpcddhxcdd, hxcdddioxins, dioxinsabundance, abundancerrt, rrtocdf
An Alternate Testing Protocol for EPA 1613B using Agilent Triple Quadrupole GC/MS
2021|Agilent Technologies|Applications
Application Note An Alternate Testing Protocol for EPA 1613B using Agilent Triple Quadrupole GC/MS Determination of 2,3,7,8-substituted tetra- through octa-chlorinated dibenzo-p-dioxins and dibenzofurans Authors Coreen Hamilton and Xinhui Xie, SGS AXYS Analytical Services Ltd. Tarun Anumol, Anastasia Andrianova, and Dale…
Key words
dioxins, dioxinsfurans, furanssgs, sgsaxys, axyspcdfs, pcdfsocdd, ocddpcdd, pcddhrms, hrmspcdf, pcdfocta, octapcdds, pcddsocdf, ocdfmrm, mrmrrf, rrfwere
Determination of Ultratrace Polychlorinated Dibenzo-p-Dioxins and Dibenzofurans Using GC/MS/MS
2019|Agilent Technologies|Applications
Application Note Environmental Determination of Ultratrace Polychlorinated Dibenzo-p-Dioxins and Dibenzofurans Using GC/MS/MS Authors Xu Chi, Gao Lirong, and Zheng Minghui State Key Laboratory of Environmental Chemistry and Ecotoxicology, Research Center for Eco‑Environmental Sciences, Chinese Academy of Sciences, Beijing, China Wang…
Key words
pcdd, pcdddioxins, dioxinsteq, teqocdf, ocdfocdd, ocddfish, fishultratrace, ultratraceash, ashfly, flywere, werecongeners, congenerspolychlorinated, polychlorinatedhrms, hrmsdibenzo, dibenzotissue
Analysis of Dioxins in Environmental Samples using GC/MS - Consumable Workflow Ordering Guide
2021|Agilent Technologies|Guides
Analysis of Dioxins in Environmental Samples using GC/MS Consumable Workflow Ordering Guide 9 8 O 1 2 PCDDs 7 Clx 6 O 3 4 Cly Analysis of Dioxins in Environmental Samples using GC/MS Determination of 2,3,7,8-substituted tetra- through octa-chlorinated dibenzo-p-dioxins…
Key words
dioxins, dioxinsmylist, mylistfurans, furanspolychlorinated, polychlorinatedpcdds, pcddsdibenzofurans, dibenzofuransdibenzo, dibenzototal, totalhrms, hrmsaxys, axyssgs, sgsenvironmental, environmentalhrgc, hrgcapproved, approvedhepta