Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Applications | 2014 | ShimadzuInstrumentation
GC/MSD, HeadSpace, GC/SQ
Industries
Environmental
Manufacturer
Shimadzu

Summary

Significance of the Topic


Volatile Organic Compounds (VOCs) are common environmental contaminants found in gasoline, solvents, and industrial effluents. Due to their toxicity and potential carcinogenicity, regulatory bodies such as the U.S. EPA enforce strict limits for VOCs in drinking water. The headspace trap GC-MS approach offers a streamlined alternative to traditional purge-and-trap methods, enabling laboratories to ensure water safety and regulatory compliance.

Objectives and Study Overview


This study demonstrates a robust analytical workflow for quantifying 76 target VOCs in finished drinking water using Shimadzu HS-20 headspace trap sampler coupled to GCMS-QP2010Ultra. It evaluates method performance against EPA Method 524.3 criteria, focusing on sensitivity, precision, accuracy, and compliance with calibration and quality-control requirements.

Methodology and Instrumentation


Sample Preparation and Headspace Trapping:
  • Water samples (10 mL) sealed in vials with acid preservatives.
  • Dynamic headspace trap mode: incubation at 60 °C, trap cooled to 5 °C, equilibrated 30 min, and desorbed at 250 °C.
Gas Chromatography Conditions:
  • Column: Rtx-VMS 60 m × 0.25 mm, 1.4 μm film.
  • Carrier gas: He at 31.3 mL/min (linear velocity control).
  • Oven program: 45 °C (4.5 min); 12 °C/min to 100 °C; 25 °C/min to 240 °C.
Mass Spectrometry Conditions:
  • Ionization: 70 eV EI.
  • Source/Interface: 200 °C.
  • Scan range: m/z 35–300; solvent cut 1.5 min.
Calibration and Standards:
  • Mass spectrometer tuned with 4-bromofluorobenzene (BFB) per EPA specifications.
  • Primary dilution standards (2–160 μg/mL) prepared in microvials.
  • Final calibration standards ranged 0.5–40 μg/L; internal and surrogate standards spiked at each level.
  • Continuing Calibration Checks (CCC) at 5 μg/L performed at batch start, every tenth sample, and end.

Results and Discussion


Calibration Performance:
  • All target analytes achieved R² > 0.995 (some up to 1.000).
  • Recoveries across seven calibration levels ranged 80–120% for most compounds.
Detection Limits and Sensitivity:
  • Method Detection Limits (MDLs) for VOCs were below the lowest calibration point (0.5 μg/L), except t-butyl alcohol at 0.88 μg/L.
Precision and Accuracy:
  • Seven replicates at 5 μg/L showed RSD < 20% and recoveries within 80–120%. Example: chlorodifluoromethane recovery 90% with 2.9% RSD.
Quality Control:
  • CCC recoveries for all analytes remained within ±30% of true value.
  • Internal standards peak areas stable within ±30% for mid-range and ±50% for lowest level.
  • Surrogate standards showed recoveries between 70% and 130% across 10 field samples.
Chromatographic Separation:
  • All 76 analytes were resolved except m-/p-xylene, which co-eluted and were integrated together.

Benefits and Practical Applications


This headspace trap GC-MS method provides:
  • Reduced sample handling compared to purge-trap, minimizing analyte loss.
  • High throughput with automated incubation and trapping.
  • Comprehensive VOC coverage meeting EPA 524.3 regulatory requirements.
  • Robust performance in routine environmental monitoring and QA/QC labs.

Future Trends and Opportunities


Advances may include:
  • Integration with high-resolution MS for enhanced selectivity and lower detection limits.
  • Miniaturized or portable headspace-GC-MS systems for field screening.
  • Automated data processing with AI-driven identification and quantification.
  • Expanded methods for emerging contaminants and other environmental matrices.

Conclusion


Shimadzu HS-20 headspace trap coupled with GCMS-QP2010Ultra offers a reliable, compliant alternative to traditional purge-and-trap techniques for VOC analysis in drinking water. The method meets EPA 524.3 criteria for sensitivity, accuracy, precision, and quality control, supporting robust environmental monitoring and regulatory compliance.

References


1. B. Prakash, A. D. Zaffiro, M. Zimmerman, D. J. Munch, B. V. Pepich. Method 524.3: Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry. EPA Document 815-B-09, Version 1.0, June 2009.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Static Headspace GC-MS
Application News AD-0086 HS-20 GCMS-QP2010 Ultra Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Static Headspace GC-MS  Introduction Determination of Volatile Organic Compounds (VOCs) has been well established by the Environmental Protection Agency (EPA) to…
Key words
ether, etherccc, cccsurrogate, surrogatestandard, standardbutylbenzene, butylbenzenepass, passheadspace, headspacemethyl, methylinternal, internalbutyl, butylvolatile, volatileethyl, ethylchloride, chlorideamyl, amylcompound
Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste
No. SSI-GCMS-1503 Gas Chromatograph Mass Spectrometer No. GCMS-1503 Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste ■ Introduction Environmental contamination has been at the forefront of government policy and…
Key words
pass, passistd, istdacetate, acetatebfb, bfbtert, tertbutylbenzene, butylbenzenerestek, restekccv, ccvrsd, rsdmethyl, methylsurr, surrbutyl, butylvinyl, vinylethyl, ethylamyl
US EPA Method 524.3 with the Tekmar Lumin P&T Concentrator and Agilent 7890B GC and 5977A MSD System
Application Note US EPA Method 524.3 with the Tekmar Lumin P&T Concentrator and Agilent 7890B GC and 5977A MSD System Author Roger Bardsley Applications Chemist, Teledyne Tekmar Abstract US EPA Method 524.3 is used primarily by environmental labs for the…
Key words
temp, temppurge, purgebake, bakelumin, luminwater, waterether, etherdesorb, desorbtekmar, tekmarvariable, variabletime, timemcs, mcsteledyne, teledyneamount, amountbutylbenzene, butylbenzenelpir
Analysis of volatile organic compounds in drinking water according to U.S. EPA Method 524.4
Application note | 003028 Environmental Analysis of volatile organic compounds in drinking water according to U.S. EPA Method 524.4 Goal Authors Adam Ladak , Terry Jeffers , and Demonstration of an analytical method that meets the requirements outlined in U.S.…
Key words
lin, linaquatek, aquatekavgcalfact, avgcalfactether, etherlva, lvabutylbenzene, butylbenzenemethyl, methylbutyl, butylbake, bakeprecision, precisionethyl, ethylamyl, amyllumin, luminpurge, purgechloride
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike