SPE Troubleshooting - Extra Peak

We, 28.1.2026
| Original article from: Phenomenex
An SPE technical tip focused on identifying the sources of interferences and extra peaks observed after sample extraction.
<p><strong>Phenomenex: </strong><span>SPE Troubleshooting - Extra Peak</span></p>

Phenomenex: SPE Troubleshooting - Extra Peak

Welcome to our new SPE technical tip!

What is the cause when you encounter interferences post-extraction regardless of the analytical or detection method? Possible sources of the mystery peak(s) are outlined below.

Possible Cause/Source   /   Suggested Remedy

  • Contaminated reagents: Analyze the purity of reagents and additives or try using these from a different supplier.
  • Sample pre-treatment: Compare chromatograms of pre-treated and non-pre-treated sample.
  • Mismatched injection solvent: Analytes that are dissolved in a strong injection solvent may retain poorly and thus appear as extra peaks. Using the initial mobile phase as the injection solvent is recommended.
  • Incorrect SPE method: The SPE sorbent might not have been properly conditioned and equilibrated. Typically, the strongest solvent used in the method should be used to condition the sorbent. Analyze the conditioning, loading and washing fractions to check if the same interferences appear.
  • SPE tube leaching: Before conditioning, clean the tube with elution solvent. If reagents or the nature of the sample are too harsh for the polypropylene SPE tube, a glass tube or PTFE tube could be used instead.
  • Degradation/side reaction: Check the reactivity and stability of all components in your sample. Degradation can happen if compounds are sensitive to factors such as heat, light, acids, bases, etc. Sample components may be reacting with each other or with reagents/additives.

We hope this tip was useful. See you soon for another tip!

Phenomenex
LinkedIn Logo
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters
Scientific article | Water

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters

In the article in the Journal of Chromatography A the researchers presented a simple on-site SPE extraction, followed by GC-MS for analyzing rotenone-containing piscicides in aquatic environments.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters
Scientific article | Water

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters

In the article in the Journal of Chromatography A the researchers presented a simple on-site SPE extraction, followed by GC-MS for analyzing rotenone-containing piscicides in aquatic environments.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters
Scientific article | Water

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters

In the article in the Journal of Chromatography A the researchers presented a simple on-site SPE extraction, followed by GC-MS for analyzing rotenone-containing piscicides in aquatic environments.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters
Scientific article | Water

Simple on-site extraction and GC-MS analysis of rotenone and degradation products for monitoring invasive fish eradication treatments in fresh and brackish waters

In the article in the Journal of Chromatography A the researchers presented a simple on-site SPE extraction, followed by GC-MS for analyzing rotenone-containing piscicides in aquatic environments.
LabRulez
tag
share
more
News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike