SPE Troubleshooting - Extra Peak

We, 28.1.2026
| Original article from: Phenomenex
An SPE technical tip focused on identifying the sources of interferences and extra peaks observed after sample extraction.
<p><strong>Phenomenex: </strong><span>SPE Troubleshooting - Extra Peak</span></p>

Phenomenex: SPE Troubleshooting - Extra Peak

Welcome to our new SPE technical tip!

What is the cause when you encounter interferences post-extraction regardless of the analytical or detection method? Possible sources of the mystery peak(s) are outlined below.

Possible Cause/Source   /   Suggested Remedy

  • Contaminated reagents: Analyze the purity of reagents and additives or try using these from a different supplier.
  • Sample pre-treatment: Compare chromatograms of pre-treated and non-pre-treated sample.
  • Mismatched injection solvent: Analytes that are dissolved in a strong injection solvent may retain poorly and thus appear as extra peaks. Using the initial mobile phase as the injection solvent is recommended.
  • Incorrect SPE method: The SPE sorbent might not have been properly conditioned and equilibrated. Typically, the strongest solvent used in the method should be used to condition the sorbent. Analyze the conditioning, loading and washing fractions to check if the same interferences appear.
  • SPE tube leaching: Before conditioning, clean the tube with elution solvent. If reagents or the nature of the sample are too harsh for the polypropylene SPE tube, a glass tube or PTFE tube could be used instead.
  • Degradation/side reaction: Check the reactivity and stability of all components in your sample. Degradation can happen if compounds are sensitive to factors such as heat, light, acids, bases, etc. Sample components may be reacting with each other or with reagents/additives.

We hope this tip was useful. See you soon for another tip!

Phenomenex
LinkedIn Logo
 

Related content

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
 

Related content

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
 

Related content

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
 

Related content

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Alcohol Determination of Sanitizer Gel Using Nexis GC-2060

Applications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

News from LabRulezGCMS Library - Week 41, 2024
Article | Application

News from LabRulezGCMS Library - Week 41, 2024

This week we bring to you applications by Agilent Technologies, Thermo Fisher Scientific, and Shimadzu!
LabRulez
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike